Loading…

Field conjugate acoustic lenses for ultrasound hyperthermia

Multipoint foci have been synthesized by applying the pseudoinverse field conjugation method to a single ultrasonic transducer coupled to a polystyrene lens. The lens design is based on phased array calculations are then fabricated on a computer-controlled milling machine. The measured beam patterns...

Full description

Saved in:
Bibliographic Details
Published in:IEEE transactions on ultrasonics, ferroelectrics, and frequency control ferroelectrics, and frequency control, 1993-09, Vol.40 (5), p.592-602
Main Authors: Lalonde, R.J., Worthington, A., Hunt, J.W.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Multipoint foci have been synthesized by applying the pseudoinverse field conjugation method to a single ultrasonic transducer coupled to a polystyrene lens. The lens design is based on phased array calculations are then fabricated on a computer-controlled milling machine. The measured beam patterns from the lenses agree closely with the beam patterns predicted by theory for the equivalent phased arrays. Temperature distributions from thermal modeling and those measured in tissue equivalent phantoms show that the lens system is capable of generating strongly localized, controlled temperature fields for hyperthermia.< >
ISSN:0885-3010
1525-8955
DOI:10.1109/58.238113