Loading…
Comments on "Digital Filters for Real-Time ECG Signal Processing Using Microprocessors"
In an earlier paper1 an adaptive filtering algorithm was proposed for the elimination of 60 Hz interference in the ECG. It is shown here that this algorithm is not truly adaptive, but is approximately equivalent to a fixed 60 Hz notch filter.
Saved in:
Published in: | IEEE transactions on biomedical engineering 1987-12, Vol.BME-34 (12), p.962-963 |
---|---|
Main Author: | |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In an earlier paper1 an adaptive filtering algorithm was proposed for the elimination of 60 Hz interference in the ECG. It is shown here that this algorithm is not truly adaptive, but is approximately equivalent to a fixed 60 Hz notch filter. |
---|---|
ISSN: | 0018-9294 1558-2531 |
DOI: | 10.1109/TBME.1987.325936 |