Comments on "Digital Filters for Real-Time ECG Signal Processing Using Microprocessors"

In an earlier paper1 an adaptive filtering algorithm was proposed for the elimination of 60 Hz interference in the ECG. It is shown here that this algorithm is not truly adaptive, but is approximately equivalent to a fixed 60 Hz notch filter.

Saved in:
Bibliographic Details
Published in:IEEE transactions on biomedical engineering 1987-12, Vol.BME-34 (12), p.962-963
Main Author: Glover, John R.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In an earlier paper1 an adaptive filtering algorithm was proposed for the elimination of 60 Hz interference in the ECG. It is shown here that this algorithm is not truly adaptive, but is approximately equivalent to a fixed 60 Hz notch filter.