Exact frequency-domain reconstruction for thermoacoustic tomography. II. Cylindrical geometry
For pt. I see ibid., vol. 21, no. 7, p. 823-8 (2002). Microwave-induced thermoacoustic tomography (TAT) in a cylindrical configuration is developed to image biological tissue. Thermoacoustic signals are acquired by scanning a flat ultrasonic transducer. Using a new expansion of a spherical wave in c...
Saved in:
Published in: | IEEE transactions on medical imaging 2002-07, Vol.21 (7), p.829-833 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | For pt. I see ibid., vol. 21, no. 7, p. 823-8 (2002). Microwave-induced thermoacoustic tomography (TAT) in a cylindrical configuration is developed to image biological tissue. Thermoacoustic signals are acquired by scanning a flat ultrasonic transducer. Using a new expansion of a spherical wave in cylindrical coordinates, we apply the Fourier and Hankel transforms to TAT and obtain an exact frequency-domain reconstruction method. The effect of discrete spatial sampling on image quality is analyzed. An aliasing-proof reconstruction method is proposed. Numerical and experimental results are included. |
---|---|
ISSN: | 0278-0062 1558-254X |
DOI: | 10.1109/TMI.2002.801171 |