Electron Linear Accelerator Structures and Design for Radiation Therapy Machines
Requirements for straight- and bent-beam medical accelerator configurations are compared. Technical progress of straight-beam accelerators over the past 12 years is reviewed and the trade-off between optimized shunt impedance and maximum accelerating gradient in structure design is examined. Relativ...
Saved in:
Published in: | IEEE transactions on nuclear science 1981-01, Vol.28 (2), p.1884-1888 |
---|---|
Main Author: | |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Requirements for straight- and bent-beam medical accelerator configurations are compared. Technical progress of straight-beam accelerators over the past 12 years is reviewed and the trade-off between optimized shunt impedance and maximum accelerating gradient in structure design is examined. Relative advantages of side-coupled and interlaced structures are delineated in the light of much higher permissible internal electric fields, recently demonstrated. A novel concept for a single-section standing-wave structure with multiple photon mode capability is presented. |
---|---|