Calibration of a Field Spectroradiometer

The processes used to calibrate a field spectroradiometer are described. The spectro-radiometer uses four CVF's and four detectors with two of each in two separate instruments. Wavelength calibration procedure was to illuminate the aperture of the spectroradiometer with energy containing spectr...

Full description

Saved in:
Bibliographic Details
Published in:Optical Engineering 1973-08, Vol.12 (4), p.124124-124124
Main Authors: Leamer, Ross W, La Rocca, Anthony J
Format: Article
Language:English
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The processes used to calibrate a field spectroradiometer are described. The spectro-radiometer uses four CVF's and four detectors with two of each in two separate instruments. Wavelength calibration procedure was to illuminate the aperture of the spectroradiometer with energy containing spectral lines at known wavelengths, and to note the linearity of the wavelength signal over the range of the spectral lines. Spectral radiance calibration consisted of comparing analog signals from the detectors to calibrated external references under known conditions.