Stripping-voltammetric signals on carbon electrodes modified with mercury

Conditions for the formation of a stable and reproducible thin-film mercury microdrop cover on electrodes from carbon glass ceramics, carbon glass, and graphite are proposed. The influence of various factors on the parameters of signals from cadmium (II), lead (II), and copper (II) is studied by str...

Full description

Saved in:
Bibliographic Details
Published in:Journal of analytical chemistry (New York, N.Y.) N.Y.), 2009-11, Vol.64 (11), p.1141-1147
Main Authors: Kamenev, A. I., Viter, I. P., Mishukova, E. V.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Conditions for the formation of a stable and reproducible thin-film mercury microdrop cover on electrodes from carbon glass ceramics, carbon glass, and graphite are proposed. The influence of various factors on the parameters of signals from cadmium (II), lead (II), and copper (II) is studied by stripping voltammetry with linear, differential-pulse, sinusoidal, and square-wave potential sweep. The parameters of the stripping voltammetric experiment are optimized. The value of RSD in determining the specified modeling trace components did not exceed 15%. Approaches to the multicomponent stripping voltammetric analysis of waters are proposed.
ISSN:1061-9348