Development of a system for monitoring the shape, position, and intensity of the extracted relativistic ion beam at the Nuclotron-M accelerator complex at JINR
Test samples of detectors and electronics for them constructed for the purpose of monitoring the “intense” relativistic ion beams extracted from the accelerator of the Nuclotron-M accelerator complex in real time are described. The system was tested in a series of acceleration runs with deuteron bea...
Saved in:
Published in: | Physics of particles and nuclei letters 2014-11, Vol.11 (6), p.788-796 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Test samples of detectors and electronics for them constructed for the purpose of monitoring the “intense” relativistic ion beams extracted from the accelerator of the Nuclotron-M accelerator complex in real time are described. The system was tested in a series of acceleration runs with deuteron beams with an intensity of up to 10
10
1/s and beams of carbon nuclei. The system allows one to perform multiple measurements of the two-dimensional distribution of the beam intensity in the plane perpendicular to it and the beam position in this plane during the beam dump and measure the two-dimensional distribution of the target irradiation dose after each beam dump. |
---|---|