Loading…
Determination of Glucose in Human Serum Using Background Subtraction of Cyclic Voltammetry of a Low Oxidation Peak at NiO Based Electrode
In this study, the background subtraction of cyclic voltammetry of a low oxidation peak at NiO based electrode is employed for the determination of glucose in human serum in 0.1 M NaOH medium with 2 linear ranges of 10 to 100 μ M and 100 to 2000 μ M with a detection limit of 2 μ M and the sensitivit...
Saved in:
Published in: | Journal of the Electrochemical Society 2022-02, Vol.169 (2), p.27503 |
---|---|
Main Authors: | , , , , , , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In this study, the background subtraction of cyclic voltammetry of a low oxidation peak at NiO based electrode is employed for the determination of glucose in human serum in 0.1 M NaOH medium with 2 linear ranges of 10 to 100
μ
M and 100 to 2000
μ
M with a detection limit of 2
μ
M and the sensitivities of 2942.2 and 754.2
μ
Acm
−2
mM
−1
, respectively. As a result, determining the glucose concentration in the real sample shows that this type of NiO-based sensor can be used in practice and approach to accurately determining the glucose concentration in human serum. |
---|---|
ISSN: | 0013-4651 1945-7111 |
DOI: | 10.1149/1945-7111/ac4e5a |