Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data
Sensor errors resulting in elevated values of N concentration [N ] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of...
Saved in:
Published in: | Journal of applied physiology (1985) 2023-04, Vol.134 (4), p.879-886 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673 |
---|---|
cites | cdi_FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673 |
container_end_page | 886 |
container_issue | 4 |
container_start_page | 879 |
container_title | Journal of applied physiology (1985) |
container_volume | 134 |
creator | Prisk, G Kim Rutting, Sandra Bozier, Jack Thamrin, Cindy Robinson, Paul D Thompson, Bruce R |
description | Sensor errors resulting in elevated values of N
concentration [N
] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of conductive and acinar ventilatory heterogeneity (
and
) although the mechanism by which this change in values occurs remains unclear. Exploring these effects also provides a timely opportunity to examine the appropriateness of the algorithm used to calculate these indices. Using a two-compartment model with differing specific ventilation (SV) such that the lower SV unit empties late, noise-free MBW were simulated both corrected and uncorrected for the recent sensor error.
was calculated using regression of normalized phase III slope (Sn
) against lung turnover (TO) from a TO range starting at 1.5 and ending at an upper turnover (TO
) between 4 and 8 (default 6) over a range of simulated values. The principal effect of the sensor error was that as the MBW proceeded the phase III slope of successive breaths was normalized by an increasingly overestimated [N
], resulting in Sn
values that fell precipitously at high TO, greatly reducing
. Reanalysis of previously published data and of simulated data showed a large proportional bias in
, whereas
was only minimally affected. In adult subject data, reducing TO
below 5.5 was associated with a large drop of up to ∼60% in
calculated from data corrected for sensor error. Raising the upper TO limit elevated
by ∼20% but with a large concomitant increase in variability. In contrast to
,
was relatively unaffected by changes in TO
with changes of |
doi_str_mv | 10.1152/japplphysiol.00423.2022 |
format | article |
fullrecord | <record><control><sourceid>pubmed_cross</sourceid><recordid>TN_cdi_crossref_primary_10_1152_japplphysiol_00423_2022</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>36825644</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673</originalsourceid><addsrcrecordid>eNpN0EtOwzAQgGELgWgpXAF8gRQ_4iRdVhUvCYlFYR059rhxlcaR7RTlBhybNgXEahYz3yx-hO4omVMq2P1Wdl3T1UOwrpkTkjI-Z4SxMzQ9bFlCM0LP0bTIBUlyUeQTdBXClhCapoJeognPCiayNJ2ir2UIEIJtNzjWgJVsVN_IaF2LncHKtbpX0e4By1ZjqWwrPd5DG-3hyPkB1xDBuw20YOOAbautgoDXoxzNelTYeLfDu76JtmsgqTzIWONPGWrXR6xllNfowsgmwM3PnKGPx4f31XPy-vb0slq-JooykSa6yPUCWEG55gvKM0MrIUUBvDIcNOOmEgyoYSaT3ChhWEEKVWmW0tQIleV8hvLTX-VdCB5M2Xm7k34oKSmPbcv_bcuxbXlse5C3J9n11Q70n_uNyb8BOyZ9BQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data</title><source>American Physiological Society:Jisc Collections:American Physiological Society Journals ‘Read Publish & Join’ Agreement:2023-2024 (Reading list)</source><source>American Physiological Society Free</source><creator>Prisk, G Kim ; Rutting, Sandra ; Bozier, Jack ; Thamrin, Cindy ; Robinson, Paul D ; Thompson, Bruce R</creator><creatorcontrib>Prisk, G Kim ; Rutting, Sandra ; Bozier, Jack ; Thamrin, Cindy ; Robinson, Paul D ; Thompson, Bruce R</creatorcontrib><description>Sensor errors resulting in elevated values of N
concentration [N
] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of conductive and acinar ventilatory heterogeneity (
and
) although the mechanism by which this change in values occurs remains unclear. Exploring these effects also provides a timely opportunity to examine the appropriateness of the algorithm used to calculate these indices. Using a two-compartment model with differing specific ventilation (SV) such that the lower SV unit empties late, noise-free MBW were simulated both corrected and uncorrected for the recent sensor error.
was calculated using regression of normalized phase III slope (Sn
) against lung turnover (TO) from a TO range starting at 1.5 and ending at an upper turnover (TO
) between 4 and 8 (default 6) over a range of simulated values. The principal effect of the sensor error was that as the MBW proceeded the phase III slope of successive breaths was normalized by an increasingly overestimated [N
], resulting in Sn
values that fell precipitously at high TO, greatly reducing
. Reanalysis of previously published data and of simulated data showed a large proportional bias in
, whereas
was only minimally affected. In adult subject data, reducing TO
below 5.5 was associated with a large drop of up to ∼60% in
calculated from data corrected for sensor error. Raising the upper TO limit elevated
by ∼20% but with a large concomitant increase in variability. In contrast to
,
was relatively unaffected by changes in TO
with changes of <3%. This work serves to emphasize that the upper limit of TO of 6 represents an appropriate upper limit for the calculation of
.
Sensor errors that elevated values of N
concentration in commercial multiple-breath washout (MBW) devices resulted in errors in calculated parameters including
and
. We examined the mechanism of the change in values produced by these errors and explored the appropriateness of the calculation of
and
. This work serves to emphasize that the current algorithm in use is appropriate for the calculation of
and
.</description><identifier>ISSN: 8750-7587</identifier><identifier>EISSN: 1522-1601</identifier><identifier>DOI: 10.1152/japplphysiol.00423.2022</identifier><identifier>PMID: 36825644</identifier><language>eng</language><publisher>United States</publisher><subject>Adult ; Breath Tests - methods ; Functional Residual Capacity ; Humans ; Lung ; Respiration ; Respiratory Function Tests - methods</subject><ispartof>Journal of applied physiology (1985), 2023-04, Vol.134 (4), p.879-886</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673</citedby><cites>FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673</cites><orcidid>0000-0002-3365-9659 ; 0000-0003-3284-3558 ; 0000-0002-6344-4720 ; 0000-0001-7397-105X ; 0000-0002-4827-1374</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,778,782,27907,27908</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/36825644$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Prisk, G Kim</creatorcontrib><creatorcontrib>Rutting, Sandra</creatorcontrib><creatorcontrib>Bozier, Jack</creatorcontrib><creatorcontrib>Thamrin, Cindy</creatorcontrib><creatorcontrib>Robinson, Paul D</creatorcontrib><creatorcontrib>Thompson, Bruce R</creatorcontrib><title>Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data</title><title>Journal of applied physiology (1985)</title><addtitle>J Appl Physiol (1985)</addtitle><description>Sensor errors resulting in elevated values of N
concentration [N
] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of conductive and acinar ventilatory heterogeneity (
and
) although the mechanism by which this change in values occurs remains unclear. Exploring these effects also provides a timely opportunity to examine the appropriateness of the algorithm used to calculate these indices. Using a two-compartment model with differing specific ventilation (SV) such that the lower SV unit empties late, noise-free MBW were simulated both corrected and uncorrected for the recent sensor error.
was calculated using regression of normalized phase III slope (Sn
) against lung turnover (TO) from a TO range starting at 1.5 and ending at an upper turnover (TO
) between 4 and 8 (default 6) over a range of simulated values. The principal effect of the sensor error was that as the MBW proceeded the phase III slope of successive breaths was normalized by an increasingly overestimated [N
], resulting in Sn
values that fell precipitously at high TO, greatly reducing
. Reanalysis of previously published data and of simulated data showed a large proportional bias in
, whereas
was only minimally affected. In adult subject data, reducing TO
below 5.5 was associated with a large drop of up to ∼60% in
calculated from data corrected for sensor error. Raising the upper TO limit elevated
by ∼20% but with a large concomitant increase in variability. In contrast to
,
was relatively unaffected by changes in TO
with changes of <3%. This work serves to emphasize that the upper limit of TO of 6 represents an appropriate upper limit for the calculation of
.
Sensor errors that elevated values of N
concentration in commercial multiple-breath washout (MBW) devices resulted in errors in calculated parameters including
and
. We examined the mechanism of the change in values produced by these errors and explored the appropriateness of the calculation of
and
. This work serves to emphasize that the current algorithm in use is appropriate for the calculation of
and
.</description><subject>Adult</subject><subject>Breath Tests - methods</subject><subject>Functional Residual Capacity</subject><subject>Humans</subject><subject>Lung</subject><subject>Respiration</subject><subject>Respiratory Function Tests - methods</subject><issn>8750-7587</issn><issn>1522-1601</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><recordid>eNpN0EtOwzAQgGELgWgpXAF8gRQ_4iRdVhUvCYlFYR059rhxlcaR7RTlBhybNgXEahYz3yx-hO4omVMq2P1Wdl3T1UOwrpkTkjI-Z4SxMzQ9bFlCM0LP0bTIBUlyUeQTdBXClhCapoJeognPCiayNJ2ir2UIEIJtNzjWgJVsVN_IaF2LncHKtbpX0e4By1ZjqWwrPd5DG-3hyPkB1xDBuw20YOOAbautgoDXoxzNelTYeLfDu76JtmsgqTzIWONPGWrXR6xllNfowsgmwM3PnKGPx4f31XPy-vb0slq-JooykSa6yPUCWEG55gvKM0MrIUUBvDIcNOOmEgyoYSaT3ChhWEEKVWmW0tQIleV8hvLTX-VdCB5M2Xm7k34oKSmPbcv_bcuxbXlse5C3J9n11Q70n_uNyb8BOyZ9BQ</recordid><startdate>20230401</startdate><enddate>20230401</enddate><creator>Prisk, G Kim</creator><creator>Rutting, Sandra</creator><creator>Bozier, Jack</creator><creator>Thamrin, Cindy</creator><creator>Robinson, Paul D</creator><creator>Thompson, Bruce R</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><orcidid>https://orcid.org/0000-0002-3365-9659</orcidid><orcidid>https://orcid.org/0000-0003-3284-3558</orcidid><orcidid>https://orcid.org/0000-0002-6344-4720</orcidid><orcidid>https://orcid.org/0000-0001-7397-105X</orcidid><orcidid>https://orcid.org/0000-0002-4827-1374</orcidid></search><sort><creationdate>20230401</creationdate><title>Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data</title><author>Prisk, G Kim ; Rutting, Sandra ; Bozier, Jack ; Thamrin, Cindy ; Robinson, Paul D ; Thompson, Bruce R</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><topic>Adult</topic><topic>Breath Tests - methods</topic><topic>Functional Residual Capacity</topic><topic>Humans</topic><topic>Lung</topic><topic>Respiration</topic><topic>Respiratory Function Tests - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Prisk, G Kim</creatorcontrib><creatorcontrib>Rutting, Sandra</creatorcontrib><creatorcontrib>Bozier, Jack</creatorcontrib><creatorcontrib>Thamrin, Cindy</creatorcontrib><creatorcontrib>Robinson, Paul D</creatorcontrib><creatorcontrib>Thompson, Bruce R</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><jtitle>Journal of applied physiology (1985)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Prisk, G Kim</au><au>Rutting, Sandra</au><au>Bozier, Jack</au><au>Thamrin, Cindy</au><au>Robinson, Paul D</au><au>Thompson, Bruce R</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data</atitle><jtitle>Journal of applied physiology (1985)</jtitle><addtitle>J Appl Physiol (1985)</addtitle><date>2023-04-01</date><risdate>2023</risdate><volume>134</volume><issue>4</issue><spage>879</spage><epage>886</epage><pages>879-886</pages><issn>8750-7587</issn><eissn>1522-1601</eissn><abstract>Sensor errors resulting in elevated values of N
concentration [N
] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of conductive and acinar ventilatory heterogeneity (
and
) although the mechanism by which this change in values occurs remains unclear. Exploring these effects also provides a timely opportunity to examine the appropriateness of the algorithm used to calculate these indices. Using a two-compartment model with differing specific ventilation (SV) such that the lower SV unit empties late, noise-free MBW were simulated both corrected and uncorrected for the recent sensor error.
was calculated using regression of normalized phase III slope (Sn
) against lung turnover (TO) from a TO range starting at 1.5 and ending at an upper turnover (TO
) between 4 and 8 (default 6) over a range of simulated values. The principal effect of the sensor error was that as the MBW proceeded the phase III slope of successive breaths was normalized by an increasingly overestimated [N
], resulting in Sn
values that fell precipitously at high TO, greatly reducing
. Reanalysis of previously published data and of simulated data showed a large proportional bias in
, whereas
was only minimally affected. In adult subject data, reducing TO
below 5.5 was associated with a large drop of up to ∼60% in
calculated from data corrected for sensor error. Raising the upper TO limit elevated
by ∼20% but with a large concomitant increase in variability. In contrast to
,
was relatively unaffected by changes in TO
with changes of <3%. This work serves to emphasize that the upper limit of TO of 6 represents an appropriate upper limit for the calculation of
.
Sensor errors that elevated values of N
concentration in commercial multiple-breath washout (MBW) devices resulted in errors in calculated parameters including
and
. We examined the mechanism of the change in values produced by these errors and explored the appropriateness of the calculation of
and
. This work serves to emphasize that the current algorithm in use is appropriate for the calculation of
and
.</abstract><cop>United States</cop><pmid>36825644</pmid><doi>10.1152/japplphysiol.00423.2022</doi><tpages>8</tpages><orcidid>https://orcid.org/0000-0002-3365-9659</orcidid><orcidid>https://orcid.org/0000-0003-3284-3558</orcidid><orcidid>https://orcid.org/0000-0002-6344-4720</orcidid><orcidid>https://orcid.org/0000-0001-7397-105X</orcidid><orcidid>https://orcid.org/0000-0002-4827-1374</orcidid></addata></record> |
fulltext | fulltext |
identifier | ISSN: 8750-7587 |
ispartof | Journal of applied physiology (1985), 2023-04, Vol.134 (4), p.879-886 |
issn | 8750-7587 1522-1601 |
language | eng |
recordid | cdi_crossref_primary_10_1152_japplphysiol_00423_2022 |
source | American Physiological Society:Jisc Collections:American Physiological Society Journals ‘Read Publish & Join’ Agreement:2023-2024 (Reading list); American Physiological Society Free |
subjects | Adult Breath Tests - methods Functional Residual Capacity Humans Lung Respiration Respiratory Function Tests - methods |
title | Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-16T22%3A18%3A51IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-pubmed_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Assessing%20the%20calculation%20of%20conductive%20and%20acinar%20ventilatory%20heterogeneity%20indices%20S%20cond%20and%20S%20acin%20from%20multiple-breath%20washout%20data&rft.jtitle=Journal%20of%20applied%20physiology%20(1985)&rft.au=Prisk,%20G%20Kim&rft.date=2023-04-01&rft.volume=134&rft.issue=4&rft.spage=879&rft.epage=886&rft.pages=879-886&rft.issn=8750-7587&rft.eissn=1522-1601&rft_id=info:doi/10.1152/japplphysiol.00423.2022&rft_dat=%3Cpubmed_cross%3E36825644%3C/pubmed_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/36825644&rfr_iscdi=true |