Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data

Sensor errors resulting in elevated values of N concentration [N ] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of...

Full description

Saved in:
Bibliographic Details
Published in:Journal of applied physiology (1985) 2023-04, Vol.134 (4), p.879-886
Main Authors: Prisk, G Kim, Rutting, Sandra, Bozier, Jack, Thamrin, Cindy, Robinson, Paul D, Thompson, Bruce R
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673
cites cdi_FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673
container_end_page 886
container_issue 4
container_start_page 879
container_title Journal of applied physiology (1985)
container_volume 134
creator Prisk, G Kim
Rutting, Sandra
Bozier, Jack
Thamrin, Cindy
Robinson, Paul D
Thompson, Bruce R
description Sensor errors resulting in elevated values of N concentration [N ] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of conductive and acinar ventilatory heterogeneity ( and ) although the mechanism by which this change in values occurs remains unclear. Exploring these effects also provides a timely opportunity to examine the appropriateness of the algorithm used to calculate these indices. Using a two-compartment model with differing specific ventilation (SV) such that the lower SV unit empties late, noise-free MBW were simulated both corrected and uncorrected for the recent sensor error. was calculated using regression of normalized phase III slope (Sn ) against lung turnover (TO) from a TO range starting at 1.5 and ending at an upper turnover (TO ) between 4 and 8 (default 6) over a range of simulated values. The principal effect of the sensor error was that as the MBW proceeded the phase III slope of successive breaths was normalized by an increasingly overestimated [N ], resulting in Sn values that fell precipitously at high TO, greatly reducing . Reanalysis of previously published data and of simulated data showed a large proportional bias in , whereas was only minimally affected. In adult subject data, reducing TO below 5.5 was associated with a large drop of up to ∼60% in calculated from data corrected for sensor error. Raising the upper TO limit elevated by ∼20% but with a large concomitant increase in variability. In contrast to , was relatively unaffected by changes in TO with changes of
doi_str_mv 10.1152/japplphysiol.00423.2022
format article
fullrecord <record><control><sourceid>pubmed_cross</sourceid><recordid>TN_cdi_crossref_primary_10_1152_japplphysiol_00423_2022</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>36825644</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673</originalsourceid><addsrcrecordid>eNpN0EtOwzAQgGELgWgpXAF8gRQ_4iRdVhUvCYlFYR059rhxlcaR7RTlBhybNgXEahYz3yx-hO4omVMq2P1Wdl3T1UOwrpkTkjI-Z4SxMzQ9bFlCM0LP0bTIBUlyUeQTdBXClhCapoJeognPCiayNJ2ir2UIEIJtNzjWgJVsVN_IaF2LncHKtbpX0e4By1ZjqWwrPd5DG-3hyPkB1xDBuw20YOOAbautgoDXoxzNelTYeLfDu76JtmsgqTzIWONPGWrXR6xllNfowsgmwM3PnKGPx4f31XPy-vb0slq-JooykSa6yPUCWEG55gvKM0MrIUUBvDIcNOOmEgyoYSaT3ChhWEEKVWmW0tQIleV8hvLTX-VdCB5M2Xm7k34oKSmPbcv_bcuxbXlse5C3J9n11Q70n_uNyb8BOyZ9BQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data</title><source>American Physiological Society:Jisc Collections:American Physiological Society Journals ‘Read Publish &amp; Join’ Agreement:2023-2024 (Reading list)</source><source>American Physiological Society Free</source><creator>Prisk, G Kim ; Rutting, Sandra ; Bozier, Jack ; Thamrin, Cindy ; Robinson, Paul D ; Thompson, Bruce R</creator><creatorcontrib>Prisk, G Kim ; Rutting, Sandra ; Bozier, Jack ; Thamrin, Cindy ; Robinson, Paul D ; Thompson, Bruce R</creatorcontrib><description>Sensor errors resulting in elevated values of N concentration [N ] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of conductive and acinar ventilatory heterogeneity ( and ) although the mechanism by which this change in values occurs remains unclear. Exploring these effects also provides a timely opportunity to examine the appropriateness of the algorithm used to calculate these indices. Using a two-compartment model with differing specific ventilation (SV) such that the lower SV unit empties late, noise-free MBW were simulated both corrected and uncorrected for the recent sensor error. was calculated using regression of normalized phase III slope (Sn ) against lung turnover (TO) from a TO range starting at 1.5 and ending at an upper turnover (TO ) between 4 and 8 (default 6) over a range of simulated values. The principal effect of the sensor error was that as the MBW proceeded the phase III slope of successive breaths was normalized by an increasingly overestimated [N ], resulting in Sn values that fell precipitously at high TO, greatly reducing . Reanalysis of previously published data and of simulated data showed a large proportional bias in , whereas was only minimally affected. In adult subject data, reducing TO below 5.5 was associated with a large drop of up to ∼60% in calculated from data corrected for sensor error. Raising the upper TO limit elevated by ∼20% but with a large concomitant increase in variability. In contrast to , was relatively unaffected by changes in TO with changes of &lt;3%. This work serves to emphasize that the upper limit of TO of 6 represents an appropriate upper limit for the calculation of . Sensor errors that elevated values of N concentration in commercial multiple-breath washout (MBW) devices resulted in errors in calculated parameters including and . We examined the mechanism of the change in values produced by these errors and explored the appropriateness of the calculation of and . This work serves to emphasize that the current algorithm in use is appropriate for the calculation of and .</description><identifier>ISSN: 8750-7587</identifier><identifier>EISSN: 1522-1601</identifier><identifier>DOI: 10.1152/japplphysiol.00423.2022</identifier><identifier>PMID: 36825644</identifier><language>eng</language><publisher>United States</publisher><subject>Adult ; Breath Tests - methods ; Functional Residual Capacity ; Humans ; Lung ; Respiration ; Respiratory Function Tests - methods</subject><ispartof>Journal of applied physiology (1985), 2023-04, Vol.134 (4), p.879-886</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673</citedby><cites>FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673</cites><orcidid>0000-0002-3365-9659 ; 0000-0003-3284-3558 ; 0000-0002-6344-4720 ; 0000-0001-7397-105X ; 0000-0002-4827-1374</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,778,782,27907,27908</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/36825644$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Prisk, G Kim</creatorcontrib><creatorcontrib>Rutting, Sandra</creatorcontrib><creatorcontrib>Bozier, Jack</creatorcontrib><creatorcontrib>Thamrin, Cindy</creatorcontrib><creatorcontrib>Robinson, Paul D</creatorcontrib><creatorcontrib>Thompson, Bruce R</creatorcontrib><title>Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data</title><title>Journal of applied physiology (1985)</title><addtitle>J Appl Physiol (1985)</addtitle><description>Sensor errors resulting in elevated values of N concentration [N ] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of conductive and acinar ventilatory heterogeneity ( and ) although the mechanism by which this change in values occurs remains unclear. Exploring these effects also provides a timely opportunity to examine the appropriateness of the algorithm used to calculate these indices. Using a two-compartment model with differing specific ventilation (SV) such that the lower SV unit empties late, noise-free MBW were simulated both corrected and uncorrected for the recent sensor error. was calculated using regression of normalized phase III slope (Sn ) against lung turnover (TO) from a TO range starting at 1.5 and ending at an upper turnover (TO ) between 4 and 8 (default 6) over a range of simulated values. The principal effect of the sensor error was that as the MBW proceeded the phase III slope of successive breaths was normalized by an increasingly overestimated [N ], resulting in Sn values that fell precipitously at high TO, greatly reducing . Reanalysis of previously published data and of simulated data showed a large proportional bias in , whereas was only minimally affected. In adult subject data, reducing TO below 5.5 was associated with a large drop of up to ∼60% in calculated from data corrected for sensor error. Raising the upper TO limit elevated by ∼20% but with a large concomitant increase in variability. In contrast to , was relatively unaffected by changes in TO with changes of &lt;3%. This work serves to emphasize that the upper limit of TO of 6 represents an appropriate upper limit for the calculation of . Sensor errors that elevated values of N concentration in commercial multiple-breath washout (MBW) devices resulted in errors in calculated parameters including and . We examined the mechanism of the change in values produced by these errors and explored the appropriateness of the calculation of and . This work serves to emphasize that the current algorithm in use is appropriate for the calculation of and .</description><subject>Adult</subject><subject>Breath Tests - methods</subject><subject>Functional Residual Capacity</subject><subject>Humans</subject><subject>Lung</subject><subject>Respiration</subject><subject>Respiratory Function Tests - methods</subject><issn>8750-7587</issn><issn>1522-1601</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><recordid>eNpN0EtOwzAQgGELgWgpXAF8gRQ_4iRdVhUvCYlFYR059rhxlcaR7RTlBhybNgXEahYz3yx-hO4omVMq2P1Wdl3T1UOwrpkTkjI-Z4SxMzQ9bFlCM0LP0bTIBUlyUeQTdBXClhCapoJeognPCiayNJ2ir2UIEIJtNzjWgJVsVN_IaF2LncHKtbpX0e4By1ZjqWwrPd5DG-3hyPkB1xDBuw20YOOAbautgoDXoxzNelTYeLfDu76JtmsgqTzIWONPGWrXR6xllNfowsgmwM3PnKGPx4f31XPy-vb0slq-JooykSa6yPUCWEG55gvKM0MrIUUBvDIcNOOmEgyoYSaT3ChhWEEKVWmW0tQIleV8hvLTX-VdCB5M2Xm7k34oKSmPbcv_bcuxbXlse5C3J9n11Q70n_uNyb8BOyZ9BQ</recordid><startdate>20230401</startdate><enddate>20230401</enddate><creator>Prisk, G Kim</creator><creator>Rutting, Sandra</creator><creator>Bozier, Jack</creator><creator>Thamrin, Cindy</creator><creator>Robinson, Paul D</creator><creator>Thompson, Bruce R</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><orcidid>https://orcid.org/0000-0002-3365-9659</orcidid><orcidid>https://orcid.org/0000-0003-3284-3558</orcidid><orcidid>https://orcid.org/0000-0002-6344-4720</orcidid><orcidid>https://orcid.org/0000-0001-7397-105X</orcidid><orcidid>https://orcid.org/0000-0002-4827-1374</orcidid></search><sort><creationdate>20230401</creationdate><title>Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data</title><author>Prisk, G Kim ; Rutting, Sandra ; Bozier, Jack ; Thamrin, Cindy ; Robinson, Paul D ; Thompson, Bruce R</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><topic>Adult</topic><topic>Breath Tests - methods</topic><topic>Functional Residual Capacity</topic><topic>Humans</topic><topic>Lung</topic><topic>Respiration</topic><topic>Respiratory Function Tests - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Prisk, G Kim</creatorcontrib><creatorcontrib>Rutting, Sandra</creatorcontrib><creatorcontrib>Bozier, Jack</creatorcontrib><creatorcontrib>Thamrin, Cindy</creatorcontrib><creatorcontrib>Robinson, Paul D</creatorcontrib><creatorcontrib>Thompson, Bruce R</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><jtitle>Journal of applied physiology (1985)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Prisk, G Kim</au><au>Rutting, Sandra</au><au>Bozier, Jack</au><au>Thamrin, Cindy</au><au>Robinson, Paul D</au><au>Thompson, Bruce R</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data</atitle><jtitle>Journal of applied physiology (1985)</jtitle><addtitle>J Appl Physiol (1985)</addtitle><date>2023-04-01</date><risdate>2023</risdate><volume>134</volume><issue>4</issue><spage>879</spage><epage>886</epage><pages>879-886</pages><issn>8750-7587</issn><eissn>1522-1601</eissn><abstract>Sensor errors resulting in elevated values of N concentration [N ] in commercial multiple-breath washout (MBW) devices have been shown to prolong the washout and result in erroneously high functional residual capacity (FRC) and lung clearance index (LCI) values. The errors also affect the indices of conductive and acinar ventilatory heterogeneity ( and ) although the mechanism by which this change in values occurs remains unclear. Exploring these effects also provides a timely opportunity to examine the appropriateness of the algorithm used to calculate these indices. Using a two-compartment model with differing specific ventilation (SV) such that the lower SV unit empties late, noise-free MBW were simulated both corrected and uncorrected for the recent sensor error. was calculated using regression of normalized phase III slope (Sn ) against lung turnover (TO) from a TO range starting at 1.5 and ending at an upper turnover (TO ) between 4 and 8 (default 6) over a range of simulated values. The principal effect of the sensor error was that as the MBW proceeded the phase III slope of successive breaths was normalized by an increasingly overestimated [N ], resulting in Sn values that fell precipitously at high TO, greatly reducing . Reanalysis of previously published data and of simulated data showed a large proportional bias in , whereas was only minimally affected. In adult subject data, reducing TO below 5.5 was associated with a large drop of up to ∼60% in calculated from data corrected for sensor error. Raising the upper TO limit elevated by ∼20% but with a large concomitant increase in variability. In contrast to , was relatively unaffected by changes in TO with changes of &lt;3%. This work serves to emphasize that the upper limit of TO of 6 represents an appropriate upper limit for the calculation of . Sensor errors that elevated values of N concentration in commercial multiple-breath washout (MBW) devices resulted in errors in calculated parameters including and . We examined the mechanism of the change in values produced by these errors and explored the appropriateness of the calculation of and . This work serves to emphasize that the current algorithm in use is appropriate for the calculation of and .</abstract><cop>United States</cop><pmid>36825644</pmid><doi>10.1152/japplphysiol.00423.2022</doi><tpages>8</tpages><orcidid>https://orcid.org/0000-0002-3365-9659</orcidid><orcidid>https://orcid.org/0000-0003-3284-3558</orcidid><orcidid>https://orcid.org/0000-0002-6344-4720</orcidid><orcidid>https://orcid.org/0000-0001-7397-105X</orcidid><orcidid>https://orcid.org/0000-0002-4827-1374</orcidid></addata></record>
fulltext fulltext
identifier ISSN: 8750-7587
ispartof Journal of applied physiology (1985), 2023-04, Vol.134 (4), p.879-886
issn 8750-7587
1522-1601
language eng
recordid cdi_crossref_primary_10_1152_japplphysiol_00423_2022
source American Physiological Society:Jisc Collections:American Physiological Society Journals ‘Read Publish & Join’ Agreement:2023-2024 (Reading list); American Physiological Society Free
subjects Adult
Breath Tests - methods
Functional Residual Capacity
Humans
Lung
Respiration
Respiratory Function Tests - methods
title Assessing the calculation of conductive and acinar ventilatory heterogeneity indices S cond and S acin from multiple-breath washout data
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-16T22%3A18%3A51IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-pubmed_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Assessing%20the%20calculation%20of%20conductive%20and%20acinar%20ventilatory%20heterogeneity%20indices%20S%20cond%20and%20S%20acin%20from%20multiple-breath%20washout%20data&rft.jtitle=Journal%20of%20applied%20physiology%20(1985)&rft.au=Prisk,%20G%20Kim&rft.date=2023-04-01&rft.volume=134&rft.issue=4&rft.spage=879&rft.epage=886&rft.pages=879-886&rft.issn=8750-7587&rft.eissn=1522-1601&rft_id=info:doi/10.1152/japplphysiol.00423.2022&rft_dat=%3Cpubmed_cross%3E36825644%3C/pubmed_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c1254-d87d9e2813d39136f1b5a58e3bf3ed23fb52e1f2f6a3fc5f2808cbd2414f5c673%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/36825644&rfr_iscdi=true