Loading…

Corroded Nitinol Wires in Explanted Aortic Endografts: An Important Mechanism of Failure?

Purpose: To analyze surface alterations and fractures observed in the nitinol stent wires of explanted endovascular grafts used for treatment of abdominal aortic aneurysm. Methods: Twenty-one explanted Stentor devices and 1 Cragg stent were received from investigators in Germany and France. After ma...

Full description

Saved in:
Bibliographic Details
Published in:Journal of endovascular therapy 2001-06, Vol.8 (3), p.248-253
Main Authors: Heintz, Carsten, Riepe, Gunnar, Birken, Lars, Kaiser, Edelgard, Chakfé, Nabil, Morlock, Michael, Delling, Günter, Imig, Herbert
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Purpose: To analyze surface alterations and fractures observed in the nitinol stent wires of explanted endovascular grafts used for treatment of abdominal aortic aneurysm. Methods: Twenty-one explanted Stentor devices and 1 Cragg stent were received from investigators in Germany and France. After macroscopy and photography, the explants were cleaned and the polyester coating removed. The frame was examined completely by stereomicroscopy, and irregularities were assayed by scanning electron microscopy and energy dispersive x-ray analysis (EDAX). The observed alterations were classified according to stereomicroscopic and electron microscopic morphology. Results: The mean implantation interval for the endografts was 29.1 ± 13.2 months (range 5–46). All examined explants, even those retrieved after only a few months in situ, showed pitlike surface damage 10–25 µm in diameter. Larger, irregularly shaped surface alterations were observed in ∼70% of the explants. Older explants (age >32 months) presented vast regions of decay, with bending of the wire and stress cracks in some areas. EDAX examination revealed decreased nickel concentration in the corroded regions. Conclusions: Corrosion of the nitinol wire in endovascular grafts is confirmed. Presumably, the observed pitting and irregularly shaped corrosion defects are the precursors of material failure. They weaken the thin wire, which leads to stress cracks and eventually fracture of the stent wire under circulatory pulsation. Cell-induced electrochemical corrosion and active cellular destruction of surfaces are well-known mechanisms that must be investigated for their possible roles in the corrosion of stent metals.
ISSN:1526-6028
1545-1550
DOI:10.1177/152660280100800303