Loading…

Simultaneous measurement of OH radical, H 2 O concentration, and temperature in a premixed CH 4 /air flame using TDLAS with an improved analysis method

The OH radical concentration was measured by applying tunable diode laser absorption spectroscopy, which is an in situ optical method. An optical absorption region (P7.5ff transition at 1502.7 nm) of the OH radical was selected in the near-infrared range to measure the OH radicals quantitatively in...

Full description

Saved in:
Bibliographic Details
Published in:Optics express 2022-08, Vol.30 (18), p.32031
Main Authors: So, Sunghyun, Park, Jiyeon, Yoo, Miyeon, Hwang, Jungho, Kim, Daehae, Lee, Changyeop
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The OH radical concentration was measured by applying tunable diode laser absorption spectroscopy, which is an in situ optical method. An optical absorption region (P7.5ff transition at 1502.7 nm) of the OH radical was selected in the near-infrared range to measure the OH radicals quantitatively in premixed CH 4 /air flames. An improved direct absorption spectroscopy (DAS) method based on wavelength division multiplexing was proposed to extract the H 2 O absorption signal that interfered with the OH light absorption signal, and the integral intensity of OH* chemiluminescence was compared to the measured OH radical concentration based on the improved DAS method.
ISSN:1094-4087
1094-4087
DOI:10.1364/OE.466138