Loading…

RNA extraction optimization from dengue-infected clinical residual serum using laboratory prepared reagent

In RNA isolation protocol, two essential factors decide the quality of RNA. One is an extra care of avoiding effect of omnipresent ribonucleases using potent inhibitors such as guanidine isothiocyanate (GITC). The other is the extraction reagent’s chemical combination and concentration, which is dec...

Full description

Saved in:
Bibliographic Details
Published in:Environmental and experimental biology 2022-01, Vol.19 (4), p.219-230
Main Authors: Ramachandra, Sowrabha, Randhawa, Aayushi, Kelmani, Chandrakanth, Shankar, Sathyanarayan Muthur, Doddamane, Manjulakumari
Format: Article
Language:English
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In RNA isolation protocol, two essential factors decide the quality of RNA. One is an extra care of avoiding effect of omnipresent ribonucleases using potent inhibitors such as guanidine isothiocyanate (GITC). The other is the extraction reagent’s chemical combination and concentration, which is decided based on the source sample of RNA. In virus-infected clinical samples, the extraction efficiency relies on the aetiologic agent’s titre value, which is dependent on the day of sample collection from the onset of the disease. The quality of RNA is crucial for accurate diagnosis and other downstream applications. Dengue, a flaviviral infection, displays a high titre and becomes desirable for molecular diagnosis. Here we optimized viral RNA isolation from dengue-infected NS1 (nonstructural protein) positive residual clinical serum samples based on the original GITC-phenol-chloroform extraction method. A concentrated RNA extracting reagent called REC reagent was defined and used along with the glycogen carrier. This laboratory-prepared reagent and glycogen as co-precipitant aided in better isolation of RNA. Glycogen also contributed to the visibility of the pellet, which helped to process the washing step efficiently. This methodology is cost-effective in a research laboratory setup against a commercial kit, especially when a vast number of specimens are involved. The manual methods give scope for modifications, thus bringing down the cost. The challenges connected to the usage of residual samples and the methodology is discussed in detail.
ISSN:2255-9582
2255-9582
DOI:10.22364/eeb.19.21