Loading…
Leakage Radiations in a Medical Electron Accelerator Facility: X-Rays and Capture Gamma-Rays
Absorbed dose rates of photons are measured in the maze of a medical electron accelerator facility. Radiations in the maze include high energy recoiled electrons other than low energy reflected X-rays. An ionization chamber for low energy X-rays is used for measurement, and is put in a build-up cap...
Saved in:
Published in: | RADIOISOTOPES 1990/03/15, Vol.39(3), pp.83-90 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | eng ; jpn |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Absorbed dose rates of photons are measured in the maze of a medical electron accelerator facility. Radiations in the maze include high energy recoiled electrons other than low energy reflected X-rays. An ionization chamber for low energy X-rays is used for measurement, and is put in a build-up cap thick enough to obstruct the recoiled electrons of which contribution to the chamber output comes sometimes equal to that of X-rays. Dose rates of capture gamma-rays in the maze are estimated applying the calculation of neutron dose equivalent rates reported in the preceding paper. It is proved that capture gamma-rays are dominant in the dose rates measured in the maze at 14 MV X-ray generation. The estimated absorbed dose rates of total photons including capture gamma-rays and reflected X-rays agree with the measured ones in the range from +10% to -27%. |
---|---|
ISSN: | 0033-8303 1884-4111 |
DOI: | 10.3769/radioisotopes.39.3_83 |