Loading…

Influence of Critical Flow in the Differentially Pumped Chamber AQUASEM

The article deals with the effect of the critical flow, which is examined at pumping of the differential pumping chamber of the Environmental Scanning Electron Microscope (ESEM). The article compares the state of the flow in the area of pumped chamber at pressure ratio on the apertures when the crit...

Full description

Saved in:
Bibliographic Details
Published in:Advances in Military Technology 2017-12, Vol.12 (2), p.301-310
Main Authors: Hlavatá, P., Maxa, J., Bílek, M., Vyroubal, P., Bayer, R.
Format: Article
Language:English
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The article deals with the effect of the critical flow, which is examined at pumping of the differential pumping chamber of the Environmental Scanning Electron Microscope (ESEM). The article compares the state of the flow in the area of pumped chamber at pressure ratio on the apertures when the critical flow occurs and its advantages in design of the chambers versus the flow without the clogging the nozzle. Problematics were solved using finite volume method by Ansys Fluent system.
ISSN:1802-2308
2533-4123
DOI:10.3849/aimt.01201