Loading…

An algorithm for beat-to-beat heart rate detection from the BCG based on the continuous spline wavelet transform

•An algorithm for beat-to-beat heart rate detection from the BCG is proposed.•BCG is obtained from seated subjects.•Algorithm is based on the continuous wavelet transform with splines.•Algorithm detects J-wave peak of the BCG for heart rate calculation.•Good agreement between heart rates from BCG an...

Full description

Saved in:
Bibliographic Details
Published in:Biomedical signal processing and control 2016-05, Vol.27, p.96-102
Main Authors: Alvarado-Serrano, Carlos, Luna-Lozano, Pablo Samuel, Pallàs-Areny, Ramon
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:•An algorithm for beat-to-beat heart rate detection from the BCG is proposed.•BCG is obtained from seated subjects.•Algorithm is based on the continuous wavelet transform with splines.•Algorithm detects J-wave peak of the BCG for heart rate calculation.•Good agreement between heart rates from BCG and standard ECG was obtained. This work describes an algorithm intended to detect the beat-to-beat heart rate from the ballistocardiogram (BCG) obtained from seated subjects. The algorithm is based on the continuous wavelet transform with splines, which enables the selection of an optimum scale for reducing noise and mechanical interferences. The first step of the algorithm is a learning phase in which the first four heartbeats in the BCG are detected to define initial thresholds, search windows and interval limits. The learned parameters serve to identify the next heartbeat and are readapted after each heartbeat detected to follow the heart rate and signal-amplitude changes. To evaluate the agreement between results from the algorithm and the heart rate obtained from the ECG, a Bland–Altman plot has been used to compare them for seven seated subjects. The mean error obtained was −0.03beats/min and the 95% confidence interval (±2 SD) was ±2.7beats/min, which is within the accuracy limits recommended by the Association for the Advancement of Medical Instrumentation (AAMI) standard for heart rate meters.
ISSN:1746-8094
DOI:10.1016/j.bspc.2016.02.002