Oxygen concentrators performance with nitrous oxide at 50:50 volume
Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods...
Saved in:
Published in: | Revista brasileira de anestesiologia 2014-06, Vol.64 (3), p.164-168 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | |
---|---|
cites | |
container_end_page | 168 |
container_issue | 3 |
container_start_page | 164 |
container_title | Revista brasileira de anestesiologia |
container_volume | 64 |
creator | Jorge Ronaldo Moll Joaquim Edson Vieira Judymara Lauzi Gozzani Lígia Andrade Silva Telles Mathias |
description | Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods: Adult patients undergoing low risk surgery were randomly allocated into two groups, receiving a fresh gas flow of oxygen from concentrators (O293) or of oxygen from concentrators and nitrous oxide (O293N2O). The fraction of inspired oxygen and the percentage of oxygen from fresh gas flow were measured every 10 min. The ratio of FiO2/oxygen concentration delivered was compared at various time intervals and between the groups. Results: Thirty patients were studied in each group. There was no difference in oxygen from concentrators over time for both groups, but there was a significant improvement in the FiO2 (p < 0.001) for O293 group while a significant decline (p < 0.001) for O293N2O. The FiO2/oxygen ratio varied in both groups, reaching a plateau in the O293 group. Pulse oximetry did not fall below 98.5% in either group. Conclusion: The FiO2 in the mixture of O293 and nitrous oxide fell during the observation period although oxygen saturation was higher than 98.5% throughout the study. Concentrators can be considered a stable source of oxygen for use during short anesthetic procedures, either pure or in association with nitrous oxide at 50:50 volume. |
doi_str_mv | 10.1016/j.bjane.2013.06.011 |
format | article |
fullrecord | <record><control><sourceid>doaj</sourceid><recordid>TN_cdi_doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><doaj_id>oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3</doaj_id><sourcerecordid>oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3</sourcerecordid><originalsourceid>FETCH-doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca33</originalsourceid><addsrcrecordid>eNqtzL1OwzAUQGEPRWr5eQIWv0Dd6zoJadcKBBMLA5t1Y9-0jhK7unahfXtQxSMwHekbjhCPGpQG3awG1Q0YSa1BGwWNAq1nYqFbaJYbePqci9ucB4B1U2tYiN37-bKnKF2KjmJhLImzPBL3iSf8NfkdykHGUDidskzn4ElikTVsa5BfaTxNdC9uehwzPfz1Try9PH_sXpc-4WCPHCbki00Y7BUS7y1yCW4kCwZbjR4cVX21qaDreu-c8W2FbefQmP98_QDRQFvp</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Oxygen concentrators performance with nitrous oxide at 50:50 volume</title><source>BACON - Elsevier - GLOBAL_SCIENCEDIRECT-OPENACCESS</source><source>SciELO</source><creator>Jorge Ronaldo Moll ; Joaquim Edson Vieira ; Judymara Lauzi Gozzani ; Lígia Andrade Silva Telles Mathias</creator><creatorcontrib>Jorge Ronaldo Moll ; Joaquim Edson Vieira ; Judymara Lauzi Gozzani ; Lígia Andrade Silva Telles Mathias</creatorcontrib><description>Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods: Adult patients undergoing low risk surgery were randomly allocated into two groups, receiving a fresh gas flow of oxygen from concentrators (O293) or of oxygen from concentrators and nitrous oxide (O293N2O). The fraction of inspired oxygen and the percentage of oxygen from fresh gas flow were measured every 10 min. The ratio of FiO2/oxygen concentration delivered was compared at various time intervals and between the groups. Results: Thirty patients were studied in each group. There was no difference in oxygen from concentrators over time for both groups, but there was a significant improvement in the FiO2 (p < 0.001) for O293 group while a significant decline (p < 0.001) for O293N2O. The FiO2/oxygen ratio varied in both groups, reaching a plateau in the O293 group. Pulse oximetry did not fall below 98.5% in either group. Conclusion: The FiO2 in the mixture of O293 and nitrous oxide fell during the observation period although oxygen saturation was higher than 98.5% throughout the study. Concentrators can be considered a stable source of oxygen for use during short anesthetic procedures, either pure or in association with nitrous oxide at 50:50 volume.</description><identifier>ISSN: 1806-907X</identifier><identifier>DOI: 10.1016/j.bjane.2013.06.011</identifier><language>eng</language><publisher>Sociedade Brasileira de Anestesiologia</publisher><subject>Anestesia ; Equipos y suministros ; Inhalación ; Oxigenoterapia ; Países en desarrollo ; Óxido nitroso</subject><ispartof>Revista brasileira de anestesiologia, 2014-06, Vol.64 (3), p.164-168</ispartof><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27922,27923</link.rule.ids></links><search><creatorcontrib>Jorge Ronaldo Moll</creatorcontrib><creatorcontrib>Joaquim Edson Vieira</creatorcontrib><creatorcontrib>Judymara Lauzi Gozzani</creatorcontrib><creatorcontrib>Lígia Andrade Silva Telles Mathias</creatorcontrib><title>Oxygen concentrators performance with nitrous oxide at 50:50 volume</title><title>Revista brasileira de anestesiologia</title><description>Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods: Adult patients undergoing low risk surgery were randomly allocated into two groups, receiving a fresh gas flow of oxygen from concentrators (O293) or of oxygen from concentrators and nitrous oxide (O293N2O). The fraction of inspired oxygen and the percentage of oxygen from fresh gas flow were measured every 10 min. The ratio of FiO2/oxygen concentration delivered was compared at various time intervals and between the groups. Results: Thirty patients were studied in each group. There was no difference in oxygen from concentrators over time for both groups, but there was a significant improvement in the FiO2 (p < 0.001) for O293 group while a significant decline (p < 0.001) for O293N2O. The FiO2/oxygen ratio varied in both groups, reaching a plateau in the O293 group. Pulse oximetry did not fall below 98.5% in either group. Conclusion: The FiO2 in the mixture of O293 and nitrous oxide fell during the observation period although oxygen saturation was higher than 98.5% throughout the study. Concentrators can be considered a stable source of oxygen for use during short anesthetic procedures, either pure or in association with nitrous oxide at 50:50 volume.</description><subject>Anestesia</subject><subject>Equipos y suministros</subject><subject>Inhalación</subject><subject>Oxigenoterapia</subject><subject>Países en desarrollo</subject><subject>Óxido nitroso</subject><issn>1806-907X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2014</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNqtzL1OwzAUQGEPRWr5eQIWv0Dd6zoJadcKBBMLA5t1Y9-0jhK7unahfXtQxSMwHekbjhCPGpQG3awG1Q0YSa1BGwWNAq1nYqFbaJYbePqci9ucB4B1U2tYiN37-bKnKF2KjmJhLImzPBL3iSf8NfkdykHGUDidskzn4ElikTVsa5BfaTxNdC9uehwzPfz1Try9PH_sXpc-4WCPHCbki00Y7BUS7y1yCW4kCwZbjR4cVX21qaDreu-c8W2FbefQmP98_QDRQFvp</recordid><startdate>20140601</startdate><enddate>20140601</enddate><creator>Jorge Ronaldo Moll</creator><creator>Joaquim Edson Vieira</creator><creator>Judymara Lauzi Gozzani</creator><creator>Lígia Andrade Silva Telles Mathias</creator><general>Sociedade Brasileira de Anestesiologia</general><scope>DOA</scope></search><sort><creationdate>20140601</creationdate><title>Oxygen concentrators performance with nitrous oxide at 50:50 volume</title><author>Jorge Ronaldo Moll ; Joaquim Edson Vieira ; Judymara Lauzi Gozzani ; Lígia Andrade Silva Telles Mathias</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca33</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2014</creationdate><topic>Anestesia</topic><topic>Equipos y suministros</topic><topic>Inhalación</topic><topic>Oxigenoterapia</topic><topic>Países en desarrollo</topic><topic>Óxido nitroso</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Jorge Ronaldo Moll</creatorcontrib><creatorcontrib>Joaquim Edson Vieira</creatorcontrib><creatorcontrib>Judymara Lauzi Gozzani</creatorcontrib><creatorcontrib>Lígia Andrade Silva Telles Mathias</creatorcontrib><collection>DOAJ Directory of Open Access Journals</collection><jtitle>Revista brasileira de anestesiologia</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Jorge Ronaldo Moll</au><au>Joaquim Edson Vieira</au><au>Judymara Lauzi Gozzani</au><au>Lígia Andrade Silva Telles Mathias</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Oxygen concentrators performance with nitrous oxide at 50:50 volume</atitle><jtitle>Revista brasileira de anestesiologia</jtitle><date>2014-06-01</date><risdate>2014</risdate><volume>64</volume><issue>3</issue><spage>164</spage><epage>168</epage><pages>164-168</pages><issn>1806-907X</issn><abstract>Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods: Adult patients undergoing low risk surgery were randomly allocated into two groups, receiving a fresh gas flow of oxygen from concentrators (O293) or of oxygen from concentrators and nitrous oxide (O293N2O). The fraction of inspired oxygen and the percentage of oxygen from fresh gas flow were measured every 10 min. The ratio of FiO2/oxygen concentration delivered was compared at various time intervals and between the groups. Results: Thirty patients were studied in each group. There was no difference in oxygen from concentrators over time for both groups, but there was a significant improvement in the FiO2 (p < 0.001) for O293 group while a significant decline (p < 0.001) for O293N2O. The FiO2/oxygen ratio varied in both groups, reaching a plateau in the O293 group. Pulse oximetry did not fall below 98.5% in either group. Conclusion: The FiO2 in the mixture of O293 and nitrous oxide fell during the observation period although oxygen saturation was higher than 98.5% throughout the study. Concentrators can be considered a stable source of oxygen for use during short anesthetic procedures, either pure or in association with nitrous oxide at 50:50 volume.</abstract><pub>Sociedade Brasileira de Anestesiologia</pub><doi>10.1016/j.bjane.2013.06.011</doi><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1806-907X |
ispartof | Revista brasileira de anestesiologia, 2014-06, Vol.64 (3), p.164-168 |
issn | 1806-907X |
language | eng |
recordid | cdi_doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3 |
source | BACON - Elsevier - GLOBAL_SCIENCEDIRECT-OPENACCESS; SciELO |
subjects | Anestesia Equipos y suministros Inhalación Oxigenoterapia Países en desarrollo Óxido nitroso |
title | Oxygen concentrators performance with nitrous oxide at 50:50 volume |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-09T17%3A09%3A28IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-doaj&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Oxygen%20concentrators%20performance%20with%20nitrous%20oxide%20at%2050:50%20volume&rft.jtitle=Revista%20brasileira%20de%20anestesiologia&rft.au=Jorge%20Ronaldo%20Moll&rft.date=2014-06-01&rft.volume=64&rft.issue=3&rft.spage=164&rft.epage=168&rft.pages=164-168&rft.issn=1806-907X&rft_id=info:doi/10.1016/j.bjane.2013.06.011&rft_dat=%3Cdoaj%3Eoai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3%3C/doaj%3E%3Cgrp_id%3Ecdi_FETCH-doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca33%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true |