Oxygen concentrators performance with nitrous oxide at 50:50 volume

Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods...

Full description

Saved in:
Bibliographic Details
Published in:Revista brasileira de anestesiologia 2014-06, Vol.64 (3), p.164-168
Main Authors: Jorge Ronaldo Moll, Joaquim Edson Vieira, Judymara Lauzi Gozzani, Lígia Andrade Silva Telles Mathias
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page 168
container_issue 3
container_start_page 164
container_title Revista brasileira de anestesiologia
container_volume 64
creator Jorge Ronaldo Moll
Joaquim Edson Vieira
Judymara Lauzi Gozzani
Lígia Andrade Silva Telles Mathias
description Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods: Adult patients undergoing low risk surgery were randomly allocated into two groups, receiving a fresh gas flow of oxygen from concentrators (O293) or of oxygen from concentrators and nitrous oxide (O293N2O). The fraction of inspired oxygen and the percentage of oxygen from fresh gas flow were measured every 10 min. The ratio of FiO2/oxygen concentration delivered was compared at various time intervals and between the groups. Results: Thirty patients were studied in each group. There was no difference in oxygen from concentrators over time for both groups, but there was a significant improvement in the FiO2 (p < 0.001) for O293 group while a significant decline (p < 0.001) for O293N2O. The FiO2/oxygen ratio varied in both groups, reaching a plateau in the O293 group. Pulse oximetry did not fall below 98.5% in either group. Conclusion: The FiO2 in the mixture of O293 and nitrous oxide fell during the observation period although oxygen saturation was higher than 98.5% throughout the study. Concentrators can be considered a stable source of oxygen for use during short anesthetic procedures, either pure or in association with nitrous oxide at 50:50 volume.
doi_str_mv 10.1016/j.bjane.2013.06.011
format article
fullrecord <record><control><sourceid>doaj</sourceid><recordid>TN_cdi_doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><doaj_id>oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3</doaj_id><sourcerecordid>oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3</sourcerecordid><originalsourceid>FETCH-doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca33</originalsourceid><addsrcrecordid>eNqtzL1OwzAUQGEPRWr5eQIWv0Dd6zoJadcKBBMLA5t1Y9-0jhK7unahfXtQxSMwHekbjhCPGpQG3awG1Q0YSa1BGwWNAq1nYqFbaJYbePqci9ucB4B1U2tYiN37-bKnKF2KjmJhLImzPBL3iSf8NfkdykHGUDidskzn4ElikTVsa5BfaTxNdC9uehwzPfz1Try9PH_sXpc-4WCPHCbki00Y7BUS7y1yCW4kCwZbjR4cVX21qaDreu-c8W2FbefQmP98_QDRQFvp</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Oxygen concentrators performance with nitrous oxide at 50:50 volume</title><source>BACON - Elsevier - GLOBAL_SCIENCEDIRECT-OPENACCESS</source><source>SciELO</source><creator>Jorge Ronaldo Moll ; Joaquim Edson Vieira ; Judymara Lauzi Gozzani ; Lígia Andrade Silva Telles Mathias</creator><creatorcontrib>Jorge Ronaldo Moll ; Joaquim Edson Vieira ; Judymara Lauzi Gozzani ; Lígia Andrade Silva Telles Mathias</creatorcontrib><description>Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods: Adult patients undergoing low risk surgery were randomly allocated into two groups, receiving a fresh gas flow of oxygen from concentrators (O293) or of oxygen from concentrators and nitrous oxide (O293N2O). The fraction of inspired oxygen and the percentage of oxygen from fresh gas flow were measured every 10 min. The ratio of FiO2/oxygen concentration delivered was compared at various time intervals and between the groups. Results: Thirty patients were studied in each group. There was no difference in oxygen from concentrators over time for both groups, but there was a significant improvement in the FiO2 (p &lt; 0.001) for O293 group while a significant decline (p &lt; 0.001) for O293N2O. The FiO2/oxygen ratio varied in both groups, reaching a plateau in the O293 group. Pulse oximetry did not fall below 98.5% in either group. Conclusion: The FiO2 in the mixture of O293 and nitrous oxide fell during the observation period although oxygen saturation was higher than 98.5% throughout the study. Concentrators can be considered a stable source of oxygen for use during short anesthetic procedures, either pure or in association with nitrous oxide at 50:50 volume.</description><identifier>ISSN: 1806-907X</identifier><identifier>DOI: 10.1016/j.bjane.2013.06.011</identifier><language>eng</language><publisher>Sociedade Brasileira de Anestesiologia</publisher><subject>Anestesia ; Equipos y suministros ; Inhalación ; Oxigenoterapia ; Países en desarrollo ; Óxido nitroso</subject><ispartof>Revista brasileira de anestesiologia, 2014-06, Vol.64 (3), p.164-168</ispartof><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27922,27923</link.rule.ids></links><search><creatorcontrib>Jorge Ronaldo Moll</creatorcontrib><creatorcontrib>Joaquim Edson Vieira</creatorcontrib><creatorcontrib>Judymara Lauzi Gozzani</creatorcontrib><creatorcontrib>Lígia Andrade Silva Telles Mathias</creatorcontrib><title>Oxygen concentrators performance with nitrous oxide at 50:50 volume</title><title>Revista brasileira de anestesiologia</title><description>Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods: Adult patients undergoing low risk surgery were randomly allocated into two groups, receiving a fresh gas flow of oxygen from concentrators (O293) or of oxygen from concentrators and nitrous oxide (O293N2O). The fraction of inspired oxygen and the percentage of oxygen from fresh gas flow were measured every 10 min. The ratio of FiO2/oxygen concentration delivered was compared at various time intervals and between the groups. Results: Thirty patients were studied in each group. There was no difference in oxygen from concentrators over time for both groups, but there was a significant improvement in the FiO2 (p &lt; 0.001) for O293 group while a significant decline (p &lt; 0.001) for O293N2O. The FiO2/oxygen ratio varied in both groups, reaching a plateau in the O293 group. Pulse oximetry did not fall below 98.5% in either group. Conclusion: The FiO2 in the mixture of O293 and nitrous oxide fell during the observation period although oxygen saturation was higher than 98.5% throughout the study. Concentrators can be considered a stable source of oxygen for use during short anesthetic procedures, either pure or in association with nitrous oxide at 50:50 volume.</description><subject>Anestesia</subject><subject>Equipos y suministros</subject><subject>Inhalación</subject><subject>Oxigenoterapia</subject><subject>Países en desarrollo</subject><subject>Óxido nitroso</subject><issn>1806-907X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2014</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNqtzL1OwzAUQGEPRWr5eQIWv0Dd6zoJadcKBBMLA5t1Y9-0jhK7unahfXtQxSMwHekbjhCPGpQG3awG1Q0YSa1BGwWNAq1nYqFbaJYbePqci9ucB4B1U2tYiN37-bKnKF2KjmJhLImzPBL3iSf8NfkdykHGUDidskzn4ElikTVsa5BfaTxNdC9uehwzPfz1Try9PH_sXpc-4WCPHCbki00Y7BUS7y1yCW4kCwZbjR4cVX21qaDreu-c8W2FbefQmP98_QDRQFvp</recordid><startdate>20140601</startdate><enddate>20140601</enddate><creator>Jorge Ronaldo Moll</creator><creator>Joaquim Edson Vieira</creator><creator>Judymara Lauzi Gozzani</creator><creator>Lígia Andrade Silva Telles Mathias</creator><general>Sociedade Brasileira de Anestesiologia</general><scope>DOA</scope></search><sort><creationdate>20140601</creationdate><title>Oxygen concentrators performance with nitrous oxide at 50:50 volume</title><author>Jorge Ronaldo Moll ; Joaquim Edson Vieira ; Judymara Lauzi Gozzani ; Lígia Andrade Silva Telles Mathias</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca33</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2014</creationdate><topic>Anestesia</topic><topic>Equipos y suministros</topic><topic>Inhalación</topic><topic>Oxigenoterapia</topic><topic>Países en desarrollo</topic><topic>Óxido nitroso</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Jorge Ronaldo Moll</creatorcontrib><creatorcontrib>Joaquim Edson Vieira</creatorcontrib><creatorcontrib>Judymara Lauzi Gozzani</creatorcontrib><creatorcontrib>Lígia Andrade Silva Telles Mathias</creatorcontrib><collection>DOAJ Directory of Open Access Journals</collection><jtitle>Revista brasileira de anestesiologia</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Jorge Ronaldo Moll</au><au>Joaquim Edson Vieira</au><au>Judymara Lauzi Gozzani</au><au>Lígia Andrade Silva Telles Mathias</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Oxygen concentrators performance with nitrous oxide at 50:50 volume</atitle><jtitle>Revista brasileira de anestesiologia</jtitle><date>2014-06-01</date><risdate>2014</risdate><volume>64</volume><issue>3</issue><spage>164</spage><epage>168</epage><pages>164-168</pages><issn>1806-907X</issn><abstract>Background and objectives: Few investigations have addressed the safety of oxygen from concentrators for use in anesthesia in association with nitrous oxide. This study evaluated the percent of oxygen from a concentrator in association with nitrous oxide in a semi-closed rebreathing circuit. Methods: Adult patients undergoing low risk surgery were randomly allocated into two groups, receiving a fresh gas flow of oxygen from concentrators (O293) or of oxygen from concentrators and nitrous oxide (O293N2O). The fraction of inspired oxygen and the percentage of oxygen from fresh gas flow were measured every 10 min. The ratio of FiO2/oxygen concentration delivered was compared at various time intervals and between the groups. Results: Thirty patients were studied in each group. There was no difference in oxygen from concentrators over time for both groups, but there was a significant improvement in the FiO2 (p &lt; 0.001) for O293 group while a significant decline (p &lt; 0.001) for O293N2O. The FiO2/oxygen ratio varied in both groups, reaching a plateau in the O293 group. Pulse oximetry did not fall below 98.5% in either group. Conclusion: The FiO2 in the mixture of O293 and nitrous oxide fell during the observation period although oxygen saturation was higher than 98.5% throughout the study. Concentrators can be considered a stable source of oxygen for use during short anesthetic procedures, either pure or in association with nitrous oxide at 50:50 volume.</abstract><pub>Sociedade Brasileira de Anestesiologia</pub><doi>10.1016/j.bjane.2013.06.011</doi><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 1806-907X
ispartof Revista brasileira de anestesiologia, 2014-06, Vol.64 (3), p.164-168
issn 1806-907X
language eng
recordid cdi_doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3
source BACON - Elsevier - GLOBAL_SCIENCEDIRECT-OPENACCESS; SciELO
subjects Anestesia
Equipos y suministros
Inhalación
Oxigenoterapia
Países en desarrollo
Óxido nitroso
title Oxygen concentrators performance with nitrous oxide at 50:50 volume
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-09T17%3A09%3A28IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-doaj&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Oxygen%20concentrators%20performance%20with%20nitrous%20oxide%20at%2050:50%20volume&rft.jtitle=Revista%20brasileira%20de%20anestesiologia&rft.au=Jorge%20Ronaldo%20Moll&rft.date=2014-06-01&rft.volume=64&rft.issue=3&rft.spage=164&rft.epage=168&rft.pages=164-168&rft.issn=1806-907X&rft_id=info:doi/10.1016/j.bjane.2013.06.011&rft_dat=%3Cdoaj%3Eoai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca3%3C/doaj%3E%3Cgrp_id%3Ecdi_FETCH-doaj_primary_oai_doaj_org_article_03a81ad0ce4f4940bbfdcc3d84a8bca33%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true