Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study
Respiration-gated radiotherapy (RGRT) can decrease treatment toxicity by allowing for smaller treatment volumes for mobile tumors. RGRT is commonly performed using external surrogates of tumor motion. We describe the use of time-integrated electronic portal imaging (TI-EPI) to verify the position of...
Saved in:
Published in: | Radiation oncology (London, England) England), 2007-08, Vol.2 (1), p.32-32, Article 32 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-b653t-e1b0434488126d81488c4301cb3ce530696887d458ca72e3b51192d88e639dbc3 |
---|---|
cites | cdi_FETCH-LOGICAL-b653t-e1b0434488126d81488c4301cb3ce530696887d458ca72e3b51192d88e639dbc3 |
container_end_page | 32 |
container_issue | 1 |
container_start_page | 32 |
container_title | Radiation oncology (London, England) |
container_volume | 2 |
creator | van Sörnsen de Koste, John R Cuijpers, Johan P de Geest, Frank G M Lagerwaard, Frank J Slotman, Ben J Senan, Suresh |
description | Respiration-gated radiotherapy (RGRT) can decrease treatment toxicity by allowing for smaller treatment volumes for mobile tumors. RGRT is commonly performed using external surrogates of tumor motion. We describe the use of time-integrated electronic portal imaging (TI-EPI) to verify the position of internal structures during RGRT delivery
TI-EPI portals were generated by continuously collecting exit dose data (aSi500 EPID, Portal vision, Varian Medical Systems) when a respiratory motion phantom was irradiated during expiration, inspiration and free breathing phases. RGRT was delivered using the Varian RPM system, and grey value profile plots over a fixed trajectory were used to study object positions. Time-related positional information was derived by subtracting grey values from TI-EPI portals sharing the pixel matrix. TI-EPI portals were also collected in 2 patients undergoing RPM-triggered RGRT for a lung and hepatic tumor (with fiducial markers), and corresponding planning 4-dimensional CT (4DCT) scans were analyzed for motion amplitude.
Integral grey values of phantom TI-EPI portals correlated well with mean object position in all respiratory phases. Cranio-caudal motion of internal structures ranged from 17.5-20.0 mm on planning 4DCT scans. TI-EPI of bronchial images reproduced with a mean value of 5.3 mm (1 SD 3.0 mm) located cranial to planned position. Mean hepatic fiducial markers reproduced with 3.2 mm (SD 2.2 mm) caudal to planned position. After bony alignment to exclude set-up errors, mean displacement in the two structures was 2.8 mm and 1.4 mm, respectively, and corresponding reproducibility in anatomy improved to 1.6 mm (1 SD).
TI-EPI appears to be a promising method for verifying delivery of RGRT. The RPM system was a good indirect surrogate of internal anatomy, but use of TI-EPI allowed for a direct link between anatomy and breathing patterns. |
doi_str_mv | 10.1186/1748-717x-2-32 |
format | article |
fullrecord | <record><control><sourceid>gale_doaj_</sourceid><recordid>TN_cdi_doaj_primary_oai_doaj_org_article_05d7a7dd731c459e92c7f11410e8fb21</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A171172036</galeid><doaj_id>oai_doaj_org_article_05d7a7dd731c459e92c7f11410e8fb21</doaj_id><sourcerecordid>A171172036</sourcerecordid><originalsourceid>FETCH-LOGICAL-b653t-e1b0434488126d81488c4301cb3ce530696887d458ca72e3b51192d88e639dbc3</originalsourceid><addsrcrecordid>eNqNkk1v1DAQhiMEoqVw5YgicU7x2I7tcEBala9KlbgA4mY59iTrKokjx4vYf4_TLdAVRUI-2Jr39eMZzxTFcyDnAEq8AslVJUH-qGjF6IPi9Ffg28M755PiybJcE8JrRprHxQlIKUgjyGkxfsXou72f-pK_LXuT0JXROB_SFqOZ9-VuWbXkR6z8lLCPNxYc0KYYJm_LOcRkhtKPps_O16Up562ZUhhLM7nSDj6bsr6knds_LR51Zljw2e1-Vnx5_-7zxcfq6tOHy4vNVdWKmqUKoSWcca4UUOEU5IPljIBtmcVcgmiEUtLxWlkjKbK2BmioUwoFa1xr2VlxeeC6YK71HHNyca-D8fomEGKvTUzeDqhJ7aSRzkkGltcNNtTKDoADQdW1FDLrzYE179oRncUpRTMcQY-VyW91H75rSmRdU5oBmwOg9eEfgGPFhlGvrdNr6zTVbGW8PDB6k3P2Uxey045-sXoDEkBSwkR2nd_jysvh6G2YsPM5vhGSMq4Elf974Z4XbAzLErH7XQYQvQ7k34m_uPt7f-y3E8h-AmuY2_k</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study</title><source>PubMed Central</source><creator>van Sörnsen de Koste, John R ; Cuijpers, Johan P ; de Geest, Frank G M ; Lagerwaard, Frank J ; Slotman, Ben J ; Senan, Suresh</creator><creatorcontrib>van Sörnsen de Koste, John R ; Cuijpers, Johan P ; de Geest, Frank G M ; Lagerwaard, Frank J ; Slotman, Ben J ; Senan, Suresh</creatorcontrib><description>Respiration-gated radiotherapy (RGRT) can decrease treatment toxicity by allowing for smaller treatment volumes for mobile tumors. RGRT is commonly performed using external surrogates of tumor motion. We describe the use of time-integrated electronic portal imaging (TI-EPI) to verify the position of internal structures during RGRT delivery
TI-EPI portals were generated by continuously collecting exit dose data (aSi500 EPID, Portal vision, Varian Medical Systems) when a respiratory motion phantom was irradiated during expiration, inspiration and free breathing phases. RGRT was delivered using the Varian RPM system, and grey value profile plots over a fixed trajectory were used to study object positions. Time-related positional information was derived by subtracting grey values from TI-EPI portals sharing the pixel matrix. TI-EPI portals were also collected in 2 patients undergoing RPM-triggered RGRT for a lung and hepatic tumor (with fiducial markers), and corresponding planning 4-dimensional CT (4DCT) scans were analyzed for motion amplitude.
Integral grey values of phantom TI-EPI portals correlated well with mean object position in all respiratory phases. Cranio-caudal motion of internal structures ranged from 17.5-20.0 mm on planning 4DCT scans. TI-EPI of bronchial images reproduced with a mean value of 5.3 mm (1 SD 3.0 mm) located cranial to planned position. Mean hepatic fiducial markers reproduced with 3.2 mm (SD 2.2 mm) caudal to planned position. After bony alignment to exclude set-up errors, mean displacement in the two structures was 2.8 mm and 1.4 mm, respectively, and corresponding reproducibility in anatomy improved to 1.6 mm (1 SD).
TI-EPI appears to be a promising method for verifying delivery of RGRT. The RPM system was a good indirect surrogate of internal anatomy, but use of TI-EPI allowed for a direct link between anatomy and breathing patterns.</description><identifier>ISSN: 1748-717X</identifier><identifier>EISSN: 1748-717X</identifier><identifier>DOI: 10.1186/1748-717x-2-32</identifier><identifier>PMID: 17760960</identifier><language>eng</language><publisher>England: BioMed Central Ltd</publisher><subject>Electronics ; Humans ; Lung Neoplasms - pathology ; Lung Neoplasms - radiotherapy ; Medical imaging equipment ; Methods ; Movement ; Neoplasm Staging ; Phantoms, Imaging ; Radiographic Image Enhancement - methods ; Radiotherapy ; Radiotherapy - methods ; Radiotherapy, Computer-Assisted - methods ; Reproducibility of Results ; Respiratory Physiological Phenomena</subject><ispartof>Radiation oncology (London, England), 2007-08, Vol.2 (1), p.32-32, Article 32</ispartof><rights>COPYRIGHT 2007 BioMed Central Ltd.</rights><rights>Copyright © 2007 van Sörnsen de Koste et al; licensee BioMed Central Ltd. 2007 van Sörnsen de Koste et al; licensee BioMed Central Ltd.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-b653t-e1b0434488126d81488c4301cb3ce530696887d458ca72e3b51192d88e639dbc3</citedby><cites>FETCH-LOGICAL-b653t-e1b0434488126d81488c4301cb3ce530696887d458ca72e3b51192d88e639dbc3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC2075522/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC2075522/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,885,27924,27925,53791,53793</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/17760960$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>van Sörnsen de Koste, John R</creatorcontrib><creatorcontrib>Cuijpers, Johan P</creatorcontrib><creatorcontrib>de Geest, Frank G M</creatorcontrib><creatorcontrib>Lagerwaard, Frank J</creatorcontrib><creatorcontrib>Slotman, Ben J</creatorcontrib><creatorcontrib>Senan, Suresh</creatorcontrib><title>Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study</title><title>Radiation oncology (London, England)</title><addtitle>Radiat Oncol</addtitle><description>Respiration-gated radiotherapy (RGRT) can decrease treatment toxicity by allowing for smaller treatment volumes for mobile tumors. RGRT is commonly performed using external surrogates of tumor motion. We describe the use of time-integrated electronic portal imaging (TI-EPI) to verify the position of internal structures during RGRT delivery
TI-EPI portals were generated by continuously collecting exit dose data (aSi500 EPID, Portal vision, Varian Medical Systems) when a respiratory motion phantom was irradiated during expiration, inspiration and free breathing phases. RGRT was delivered using the Varian RPM system, and grey value profile plots over a fixed trajectory were used to study object positions. Time-related positional information was derived by subtracting grey values from TI-EPI portals sharing the pixel matrix. TI-EPI portals were also collected in 2 patients undergoing RPM-triggered RGRT for a lung and hepatic tumor (with fiducial markers), and corresponding planning 4-dimensional CT (4DCT) scans were analyzed for motion amplitude.
Integral grey values of phantom TI-EPI portals correlated well with mean object position in all respiratory phases. Cranio-caudal motion of internal structures ranged from 17.5-20.0 mm on planning 4DCT scans. TI-EPI of bronchial images reproduced with a mean value of 5.3 mm (1 SD 3.0 mm) located cranial to planned position. Mean hepatic fiducial markers reproduced with 3.2 mm (SD 2.2 mm) caudal to planned position. After bony alignment to exclude set-up errors, mean displacement in the two structures was 2.8 mm and 1.4 mm, respectively, and corresponding reproducibility in anatomy improved to 1.6 mm (1 SD).
TI-EPI appears to be a promising method for verifying delivery of RGRT. The RPM system was a good indirect surrogate of internal anatomy, but use of TI-EPI allowed for a direct link between anatomy and breathing patterns.</description><subject>Electronics</subject><subject>Humans</subject><subject>Lung Neoplasms - pathology</subject><subject>Lung Neoplasms - radiotherapy</subject><subject>Medical imaging equipment</subject><subject>Methods</subject><subject>Movement</subject><subject>Neoplasm Staging</subject><subject>Phantoms, Imaging</subject><subject>Radiographic Image Enhancement - methods</subject><subject>Radiotherapy</subject><subject>Radiotherapy - methods</subject><subject>Radiotherapy, Computer-Assisted - methods</subject><subject>Reproducibility of Results</subject><subject>Respiratory Physiological Phenomena</subject><issn>1748-717X</issn><issn>1748-717X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2007</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNqNkk1v1DAQhiMEoqVw5YgicU7x2I7tcEBala9KlbgA4mY59iTrKokjx4vYf4_TLdAVRUI-2Jr39eMZzxTFcyDnAEq8AslVJUH-qGjF6IPi9Ffg28M755PiybJcE8JrRprHxQlIKUgjyGkxfsXou72f-pK_LXuT0JXROB_SFqOZ9-VuWbXkR6z8lLCPNxYc0KYYJm_LOcRkhtKPps_O16Up562ZUhhLM7nSDj6bsr6knds_LR51Zljw2e1-Vnx5_-7zxcfq6tOHy4vNVdWKmqUKoSWcca4UUOEU5IPljIBtmcVcgmiEUtLxWlkjKbK2BmioUwoFa1xr2VlxeeC6YK71HHNyca-D8fomEGKvTUzeDqhJ7aSRzkkGltcNNtTKDoADQdW1FDLrzYE179oRncUpRTMcQY-VyW91H75rSmRdU5oBmwOg9eEfgGPFhlGvrdNr6zTVbGW8PDB6k3P2Uxey045-sXoDEkBSwkR2nd_jysvh6G2YsPM5vhGSMq4Elf974Z4XbAzLErH7XQYQvQ7k34m_uPt7f-y3E8h-AmuY2_k</recordid><startdate>20070830</startdate><enddate>20070830</enddate><creator>van Sörnsen de Koste, John R</creator><creator>Cuijpers, Johan P</creator><creator>de Geest, Frank G M</creator><creator>Lagerwaard, Frank J</creator><creator>Slotman, Ben J</creator><creator>Senan, Suresh</creator><general>BioMed Central Ltd</general><general>BioMed Central</general><general>BMC</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>5PM</scope><scope>DOA</scope></search><sort><creationdate>20070830</creationdate><title>Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study</title><author>van Sörnsen de Koste, John R ; Cuijpers, Johan P ; de Geest, Frank G M ; Lagerwaard, Frank J ; Slotman, Ben J ; Senan, Suresh</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-b653t-e1b0434488126d81488c4301cb3ce530696887d458ca72e3b51192d88e639dbc3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2007</creationdate><topic>Electronics</topic><topic>Humans</topic><topic>Lung Neoplasms - pathology</topic><topic>Lung Neoplasms - radiotherapy</topic><topic>Medical imaging equipment</topic><topic>Methods</topic><topic>Movement</topic><topic>Neoplasm Staging</topic><topic>Phantoms, Imaging</topic><topic>Radiographic Image Enhancement - methods</topic><topic>Radiotherapy</topic><topic>Radiotherapy - methods</topic><topic>Radiotherapy, Computer-Assisted - methods</topic><topic>Reproducibility of Results</topic><topic>Respiratory Physiological Phenomena</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>van Sörnsen de Koste, John R</creatorcontrib><creatorcontrib>Cuijpers, Johan P</creatorcontrib><creatorcontrib>de Geest, Frank G M</creatorcontrib><creatorcontrib>Lagerwaard, Frank J</creatorcontrib><creatorcontrib>Slotman, Ben J</creatorcontrib><creatorcontrib>Senan, Suresh</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>PubMed Central (Full Participant titles)</collection><collection>DOAJ Directory of Open Access Journals</collection><jtitle>Radiation oncology (London, England)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>van Sörnsen de Koste, John R</au><au>Cuijpers, Johan P</au><au>de Geest, Frank G M</au><au>Lagerwaard, Frank J</au><au>Slotman, Ben J</au><au>Senan, Suresh</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study</atitle><jtitle>Radiation oncology (London, England)</jtitle><addtitle>Radiat Oncol</addtitle><date>2007-08-30</date><risdate>2007</risdate><volume>2</volume><issue>1</issue><spage>32</spage><epage>32</epage><pages>32-32</pages><artnum>32</artnum><issn>1748-717X</issn><eissn>1748-717X</eissn><abstract>Respiration-gated radiotherapy (RGRT) can decrease treatment toxicity by allowing for smaller treatment volumes for mobile tumors. RGRT is commonly performed using external surrogates of tumor motion. We describe the use of time-integrated electronic portal imaging (TI-EPI) to verify the position of internal structures during RGRT delivery
TI-EPI portals were generated by continuously collecting exit dose data (aSi500 EPID, Portal vision, Varian Medical Systems) when a respiratory motion phantom was irradiated during expiration, inspiration and free breathing phases. RGRT was delivered using the Varian RPM system, and grey value profile plots over a fixed trajectory were used to study object positions. Time-related positional information was derived by subtracting grey values from TI-EPI portals sharing the pixel matrix. TI-EPI portals were also collected in 2 patients undergoing RPM-triggered RGRT for a lung and hepatic tumor (with fiducial markers), and corresponding planning 4-dimensional CT (4DCT) scans were analyzed for motion amplitude.
Integral grey values of phantom TI-EPI portals correlated well with mean object position in all respiratory phases. Cranio-caudal motion of internal structures ranged from 17.5-20.0 mm on planning 4DCT scans. TI-EPI of bronchial images reproduced with a mean value of 5.3 mm (1 SD 3.0 mm) located cranial to planned position. Mean hepatic fiducial markers reproduced with 3.2 mm (SD 2.2 mm) caudal to planned position. After bony alignment to exclude set-up errors, mean displacement in the two structures was 2.8 mm and 1.4 mm, respectively, and corresponding reproducibility in anatomy improved to 1.6 mm (1 SD).
TI-EPI appears to be a promising method for verifying delivery of RGRT. The RPM system was a good indirect surrogate of internal anatomy, but use of TI-EPI allowed for a direct link between anatomy and breathing patterns.</abstract><cop>England</cop><pub>BioMed Central Ltd</pub><pmid>17760960</pmid><doi>10.1186/1748-717x-2-32</doi><tpages>1</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1748-717X |
ispartof | Radiation oncology (London, England), 2007-08, Vol.2 (1), p.32-32, Article 32 |
issn | 1748-717X 1748-717X |
language | eng |
recordid | cdi_doaj_primary_oai_doaj_org_article_05d7a7dd731c459e92c7f11410e8fb21 |
source | PubMed Central |
subjects | Electronics Humans Lung Neoplasms - pathology Lung Neoplasms - radiotherapy Medical imaging equipment Methods Movement Neoplasm Staging Phantoms, Imaging Radiographic Image Enhancement - methods Radiotherapy Radiotherapy - methods Radiotherapy, Computer-Assisted - methods Reproducibility of Results Respiratory Physiological Phenomena |
title | Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-26T17%3A48%3A38IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_doaj_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Verifying%204D%20gated%20radiotherapy%20using%20time-integrated%20electronic%20portal%20imaging:%20a%20phantom%20and%20clinical%20study&rft.jtitle=Radiation%20oncology%20(London,%20England)&rft.au=van%20S%C3%B6rnsen%20de%20Koste,%20John%20R&rft.date=2007-08-30&rft.volume=2&rft.issue=1&rft.spage=32&rft.epage=32&rft.pages=32-32&rft.artnum=32&rft.issn=1748-717X&rft.eissn=1748-717X&rft_id=info:doi/10.1186/1748-717x-2-32&rft_dat=%3Cgale_doaj_%3EA171172036%3C/gale_doaj_%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-b653t-e1b0434488126d81488c4301cb3ce530696887d458ca72e3b51192d88e639dbc3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/17760960&rft_galeid=A171172036&rfr_iscdi=true |