Loading…

NEMA NU 1-2018 performance characterization and Monte Carlo model validation of the Cubresa Spark SiPM-based preclinical SPECT scanner

Background The Cubresa Spark is a novel benchtop silicon-photomultiplier (SiPM)-based preclinical SPECT system. SiPMs in SPECT significantly improve resolution and reduce detector size compared to preclinical cameras with photomultiplier tubes requiring highly magnifying collimators. The NEMA NU 1 S...

Full description

Saved in:
Bibliographic Details
Published in:EJNMMI physics 2023-06, Vol.10 (1), p.35-35, Article 35
Main Authors: Strugari, Matthew E., DeBay, Drew R., Beyea, Steven D., Brewer, Kimberly D.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Background The Cubresa Spark is a novel benchtop silicon-photomultiplier (SiPM)-based preclinical SPECT system. SiPMs in SPECT significantly improve resolution and reduce detector size compared to preclinical cameras with photomultiplier tubes requiring highly magnifying collimators. The NEMA NU 1 Standard for Performance Measurements of Gamma Cameras provides methods that can be readily applied or extended to characterize preclinical cameras with minor modifications. The primary objective of this study is to characterize the Spark according to the NEMA NU 1-2018 standard to gain insight into its nuclear medicine imaging capabilities. The secondary objective is to validate a GATE Monte Carlo simulation model of the Spark for use in preclinical SPECT studies. Methods NEMA NU 1-2018 guidelines were applied to characterize the Spark’s intrinsic, system, and tomographic performance with single- and multi-pinhole collimators. Phantoms were fabricated according to NEMA specifications with deviations involving high-resolution modifications. GATE was utilized to model the detector head with the single-pinhole collimator, and NEMA measurements were employed to tune and validate the model. Single-pinhole and multi-pinhole SPECT data were reconstructed with the Software for Tomographic Image Reconstruction and HiSPECT, respectively. Results The limiting intrinsic resolution was measured as 0.85 mm owing to a high-resolution SiPM array combined with a 3 mm-thick scintillation crystal. The average limiting tomographic resolution was 1.37 mm and 1.19 mm for the single- and multi-pinhole collimators, respectively, which have magnification factors near unity at the center of rotation. The maximum observed count rate was 15,400 cps, and planar sensitivities of 34 cps/MBq and 150 cps/MBq were measured at the center of rotation for the single- and multi-pinhole collimators, respectively. All simulated tests agreed well with measurement, where the most considerable deviations were below 7%. Conclusions NEMA NU 1-2018 standards determined that a SiPM detector mitigates the need for highly magnifying pinhole collimators while preserving detailed information in projection images. Measured and simulated NEMA results were highly comparable with differences on the order of a few percent, confirming simulation accuracy and validating the GATE model. Of the collimators initially provided with the Spark, the multi-pinhole collimator offers high resolution and sensitivity for organ-spe
ISSN:2197-7364
2197-7364
DOI:10.1186/s40658-023-00555-6