Fenestrated aortic endograft branching with Gore VBX poses failure risk from delayed-onset branch kinking

Abstract The Gore VBX stent graft (Gore Medical, Flagstaff, Ariz) provides a new option for branching of fenestrated aortic endografts. However, its modular stent structure has raised concerns about potential kinking at the interspace between stent rings if lateralizing force exists between the fene...

Full description

Saved in:
Bibliographic Details
Published in:Journal of vascular surgery cases and innovative techniques 2019-03, Vol.5 (1), p.18-21
Main Authors: Yi, Jeniann A., MD, MS, Kuwayama, David P., MD, MPA
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Abstract The Gore VBX stent graft (Gore Medical, Flagstaff, Ariz) provides a new option for branching of fenestrated aortic endografts. However, its modular stent structure has raised concerns about potential kinking at the interspace between stent rings if lateralizing force exists between the fenestration and target vessel orifice. We present a case of near-occlusion of a VBX celiac branch due to narrowing of this interspace identified at postoperative month 3. Although the Gore VBX offers several potential advantages as a branch endoprosthesis, its design poses the risk of unpredictable, delayed-onset kinking and raises concern for its use as a fenestrated endograft branch.