Performance characterization of siemens primus linear accelerator under small monitor unit and small segments for the implementation of step-and-shoot intensitymodulated radiotherapy

Implementation of step-and-shoot intensity-modulated radiotherapy (IMRT) needs careful understanding of the accelerator start-up characteristic to ensure accurate and precise delivery of radiation dose to patient. The dosimetric characteristic of a Siemens Primus linear accelerator (LA) which delive...

Full description

Saved in:
Bibliographic Details
Published in:Journal of medical physics 2006-10, Vol.31 (4), p.269-274
Main Authors: Reena, P, Dayananda, S, Pai, Rajeshri, Jamema, S, Gupta, Tejpal, Deepak, D, Rajeev, S
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c355t-b7b3ca3f15d5ee9747dc4b78c06c4c9873992556a7ab68f2e1103ad4f81b33843
cites
container_end_page 274
container_issue 4
container_start_page 269
container_title Journal of medical physics
container_volume 31
creator Reena, P
Dayananda, S
Pai, Rajeshri
Jamema, S
Gupta, Tejpal
Deepak, D
Rajeev, S
description Implementation of step-and-shoot intensity-modulated radiotherapy (IMRT) needs careful understanding of the accelerator start-up characteristic to ensure accurate and precise delivery of radiation dose to patient. The dosimetric characteristic of a Siemens Primus linear accelerator (LA) which delivers 6 and 18 MV x-rays at the dose rate of 300 and 500 monitor unit (MU) per minutes (min) respectively was studied under the condition of small MU ranging from 1 to 100. Dose monitor linearity was studied at different dose calibration parameter (D1_C0) by measuring ionization at 10 cm depth in a solid water phantom using a 0.6 cc ionization chamber. Monitor unit stability was studied from different intensity modulated (IM) groups comprising various combinations of MU per field and number of fields. Stability of beam flatness and symmetry was investigated under normal and IMRT mode for 20x20 cm2 field under small MU using a 2D Profiler kept isocentrically at 5 cm depth. Inter segment response was investigated form 1 to 10 MU by measuring the dose per MU from various IM groups, each consisting of four segments with inter-segment separation of 2 cm. In the range 1-4 MU, the dose linearity error was more than 5% (max -32% at 1 MU) for 6 MV x-rays at factory calibrated D1_C0 value of 6000. The dose linearity error was reduced to -10.95% at 1 MU, within -3% for 2 and 3 MU and ±1% for MU ≥4 when the D1_C0 was subsequently tuned at 4500. For 18 MV x-rays, the dose linearity error at factory calibrated D1_C0 value of 4400 was within ±1% for MU ≥ 3 with maximum of -13.5 observed at 1 MU. For both the beam energies and MU/field ≥ 4, the stability of monitor unit tested for different IM groups was within ±1% of the dose from the normal treatment field. This variation increases to -2.6% for 6 MV and -2.7% for 18 MV x-rays for 2 MU/field. No significant variation was observed in the stability of beam profile measured from normal and IMRT mode. The beam flatness was within 3% for 6 MV x-rays and more than 3% (Max 3.5%) for 18 MV x-rays at lesser irradiation time ≤ 3 MU. The beam stability improves with the increase in irradiation time. Both the beam energies show very good symmetry (≤ 2%) at all irradiation time. For all the three segment sizes studied, the nonlinearity was observed at smaller MU/segment in both the energies. When the MU/segment is ≥ 4, all segment size shows fairly linear relation with dose/MU. The smaller segment size shows larger nonlinearity at smaller MU/
doi_str_mv 10.4103/0971-6203.29197
format article
fullrecord <record><control><sourceid>gale_doaj_</sourceid><recordid>TN_cdi_doaj_primary_oai_doaj_org_article_297389fa804a4c70bb0ec18eb3f80139</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A156027468</galeid><doaj_id>oai_doaj_org_article_297389fa804a4c70bb0ec18eb3f80139</doaj_id><sourcerecordid>A156027468</sourcerecordid><originalsourceid>FETCH-LOGICAL-c355t-b7b3ca3f15d5ee9747dc4b78c06c4c9873992556a7ab68f2e1103ad4f81b33843</originalsourceid><addsrcrecordid>eNptUrtuHCEUnSKR4jip0_IDs4aBGaC0rDwsWUqKpEZ34LKLNQMrYIvNh-X7wngtu7EokI7OA45O131hdCcY5TdUS9ZPA-W7QTMt33VXL8iH7mMpj5SKUQtx1f37hdmnvEK0SOwBMtiKOfyFGlIkyZMScMVYyDGH9VTIEiJCJmAtLpihpkxO0WEmZYVlIWuK4YKFSiC6Z7jgvpnUQloUqQckYT0um299zal47JuiL4eUKgmxttRQz2typwUqOpLBhdTEGY7nT917D0vBz8_3dffn29ffdz_6h5_f7-9uH3rLx7H2s5y5Be7Z6EZELYV0VsxSWTpZYbWSXOthHCeQME_KD8hae-CEV2zmXAl-3d1ffF2CR7N1APlsEgTzBKS8N5BrsAuaQUuutAdFBQgr6TxTtEzhzL2ijOvm1V-89tDoIfpUW9t7jO1HS4roQ4Nv2TjRQYpJNf7uDX47Dtdg3xTcXAQ2p1Iy-pcHM2q2XZhtBWZbgXnaBf8PtRayFQ</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Performance characterization of siemens primus linear accelerator under small monitor unit and small segments for the implementation of step-and-shoot intensitymodulated radiotherapy</title><source>Medknow Open Access Medical Journals</source><source>IngentaConnect Journals</source><source>PubMed Central</source><creator>Reena, P ; Dayananda, S ; Pai, Rajeshri ; Jamema, S ; Gupta, Tejpal ; Deepak, D ; Rajeev, S</creator><creatorcontrib>Reena, P ; Dayananda, S ; Pai, Rajeshri ; Jamema, S ; Gupta, Tejpal ; Deepak, D ; Rajeev, S</creatorcontrib><description>Implementation of step-and-shoot intensity-modulated radiotherapy (IMRT) needs careful understanding of the accelerator start-up characteristic to ensure accurate and precise delivery of radiation dose to patient. The dosimetric characteristic of a Siemens Primus linear accelerator (LA) which delivers 6 and 18 MV x-rays at the dose rate of 300 and 500 monitor unit (MU) per minutes (min) respectively was studied under the condition of small MU ranging from 1 to 100. Dose monitor linearity was studied at different dose calibration parameter (D1_C0) by measuring ionization at 10 cm depth in a solid water phantom using a 0.6 cc ionization chamber. Monitor unit stability was studied from different intensity modulated (IM) groups comprising various combinations of MU per field and number of fields. Stability of beam flatness and symmetry was investigated under normal and IMRT mode for 20x20 cm2 field under small MU using a 2D Profiler kept isocentrically at 5 cm depth. Inter segment response was investigated form 1 to 10 MU by measuring the dose per MU from various IM groups, each consisting of four segments with inter-segment separation of 2 cm. In the range 1-4 MU, the dose linearity error was more than 5% (max -32% at 1 MU) for 6 MV x-rays at factory calibrated D1_C0 value of 6000. The dose linearity error was reduced to -10.95% at 1 MU, within -3% for 2 and 3 MU and ±1% for MU ≥4 when the D1_C0 was subsequently tuned at 4500. For 18 MV x-rays, the dose linearity error at factory calibrated D1_C0 value of 4400 was within ±1% for MU ≥ 3 with maximum of -13.5 observed at 1 MU. For both the beam energies and MU/field ≥ 4, the stability of monitor unit tested for different IM groups was within ±1% of the dose from the normal treatment field. This variation increases to -2.6% for 6 MV and -2.7% for 18 MV x-rays for 2 MU/field. No significant variation was observed in the stability of beam profile measured from normal and IMRT mode. The beam flatness was within 3% for 6 MV x-rays and more than 3% (Max 3.5%) for 18 MV x-rays at lesser irradiation time ≤ 3 MU. The beam stability improves with the increase in irradiation time. Both the beam energies show very good symmetry (≤ 2%) at all irradiation time. For all the three segment sizes studied, the nonlinearity was observed at smaller MU/segment in both the energies. When the MU/segment is ≥ 4, all segment size shows fairly linear relation with dose/MU. The smaller segment size shows larger nonlinearity at smaller MU/segment and become more linear at larger MU/segment. Based on our study, we conclude that the Primus LA from Siemens installed at our hospital is ideally suited for step-and-shoot IMRT preferably for radiation ON time ³4MU per segment.</description><identifier>ISSN: 0971-6203</identifier><identifier>DOI: 10.4103/0971-6203.29197</identifier><language>eng</language><publisher>Medknow Publications and Media Pvt. Ltd</publisher><subject>Performance characterization ; small monitor unit ; small segment ; step-and-shoot intensity-modulated radiotherapy</subject><ispartof>Journal of medical physics, 2006-10, Vol.31 (4), p.269-274</ispartof><rights>COPYRIGHT 2006 Medknow Publications and Media Pvt. Ltd.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c355t-b7b3ca3f15d5ee9747dc4b78c06c4c9873992556a7ab68f2e1103ad4f81b33843</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids></links><search><creatorcontrib>Reena, P</creatorcontrib><creatorcontrib>Dayananda, S</creatorcontrib><creatorcontrib>Pai, Rajeshri</creatorcontrib><creatorcontrib>Jamema, S</creatorcontrib><creatorcontrib>Gupta, Tejpal</creatorcontrib><creatorcontrib>Deepak, D</creatorcontrib><creatorcontrib>Rajeev, S</creatorcontrib><title>Performance characterization of siemens primus linear accelerator under small monitor unit and small segments for the implementation of step-and-shoot intensitymodulated radiotherapy</title><title>Journal of medical physics</title><description>Implementation of step-and-shoot intensity-modulated radiotherapy (IMRT) needs careful understanding of the accelerator start-up characteristic to ensure accurate and precise delivery of radiation dose to patient. The dosimetric characteristic of a Siemens Primus linear accelerator (LA) which delivers 6 and 18 MV x-rays at the dose rate of 300 and 500 monitor unit (MU) per minutes (min) respectively was studied under the condition of small MU ranging from 1 to 100. Dose monitor linearity was studied at different dose calibration parameter (D1_C0) by measuring ionization at 10 cm depth in a solid water phantom using a 0.6 cc ionization chamber. Monitor unit stability was studied from different intensity modulated (IM) groups comprising various combinations of MU per field and number of fields. Stability of beam flatness and symmetry was investigated under normal and IMRT mode for 20x20 cm2 field under small MU using a 2D Profiler kept isocentrically at 5 cm depth. Inter segment response was investigated form 1 to 10 MU by measuring the dose per MU from various IM groups, each consisting of four segments with inter-segment separation of 2 cm. In the range 1-4 MU, the dose linearity error was more than 5% (max -32% at 1 MU) for 6 MV x-rays at factory calibrated D1_C0 value of 6000. The dose linearity error was reduced to -10.95% at 1 MU, within -3% for 2 and 3 MU and ±1% for MU ≥4 when the D1_C0 was subsequently tuned at 4500. For 18 MV x-rays, the dose linearity error at factory calibrated D1_C0 value of 4400 was within ±1% for MU ≥ 3 with maximum of -13.5 observed at 1 MU. For both the beam energies and MU/field ≥ 4, the stability of monitor unit tested for different IM groups was within ±1% of the dose from the normal treatment field. This variation increases to -2.6% for 6 MV and -2.7% for 18 MV x-rays for 2 MU/field. No significant variation was observed in the stability of beam profile measured from normal and IMRT mode. The beam flatness was within 3% for 6 MV x-rays and more than 3% (Max 3.5%) for 18 MV x-rays at lesser irradiation time ≤ 3 MU. The beam stability improves with the increase in irradiation time. Both the beam energies show very good symmetry (≤ 2%) at all irradiation time. For all the three segment sizes studied, the nonlinearity was observed at smaller MU/segment in both the energies. When the MU/segment is ≥ 4, all segment size shows fairly linear relation with dose/MU. The smaller segment size shows larger nonlinearity at smaller MU/segment and become more linear at larger MU/segment. Based on our study, we conclude that the Primus LA from Siemens installed at our hospital is ideally suited for step-and-shoot IMRT preferably for radiation ON time ³4MU per segment.</description><subject>Performance characterization</subject><subject>small monitor unit</subject><subject>small segment</subject><subject>step-and-shoot intensity-modulated radiotherapy</subject><issn>0971-6203</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2006</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNptUrtuHCEUnSKR4jip0_IDs4aBGaC0rDwsWUqKpEZ34LKLNQMrYIvNh-X7wngtu7EokI7OA45O131hdCcY5TdUS9ZPA-W7QTMt33VXL8iH7mMpj5SKUQtx1f37hdmnvEK0SOwBMtiKOfyFGlIkyZMScMVYyDGH9VTIEiJCJmAtLpihpkxO0WEmZYVlIWuK4YKFSiC6Z7jgvpnUQloUqQckYT0um299zal47JuiL4eUKgmxttRQz2typwUqOpLBhdTEGY7nT917D0vBz8_3dffn29ffdz_6h5_f7-9uH3rLx7H2s5y5Be7Z6EZELYV0VsxSWTpZYbWSXOthHCeQME_KD8hae-CEV2zmXAl-3d1ffF2CR7N1APlsEgTzBKS8N5BrsAuaQUuutAdFBQgr6TxTtEzhzL2ijOvm1V-89tDoIfpUW9t7jO1HS4roQ4Nv2TjRQYpJNf7uDX47Dtdg3xTcXAQ2p1Iy-pcHM2q2XZhtBWZbgXnaBf8PtRayFQ</recordid><startdate>20061001</startdate><enddate>20061001</enddate><creator>Reena, P</creator><creator>Dayananda, S</creator><creator>Pai, Rajeshri</creator><creator>Jamema, S</creator><creator>Gupta, Tejpal</creator><creator>Deepak, D</creator><creator>Rajeev, S</creator><general>Medknow Publications and Media Pvt. Ltd</general><general>Wolters Kluwer Medknow Publications</general><scope>AAYXX</scope><scope>CITATION</scope><scope>DOA</scope></search><sort><creationdate>20061001</creationdate><title>Performance characterization of siemens primus linear accelerator under small monitor unit and small segments for the implementation of step-and-shoot intensitymodulated radiotherapy</title><author>Reena, P ; Dayananda, S ; Pai, Rajeshri ; Jamema, S ; Gupta, Tejpal ; Deepak, D ; Rajeev, S</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c355t-b7b3ca3f15d5ee9747dc4b78c06c4c9873992556a7ab68f2e1103ad4f81b33843</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2006</creationdate><topic>Performance characterization</topic><topic>small monitor unit</topic><topic>small segment</topic><topic>step-and-shoot intensity-modulated radiotherapy</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Reena, P</creatorcontrib><creatorcontrib>Dayananda, S</creatorcontrib><creatorcontrib>Pai, Rajeshri</creatorcontrib><creatorcontrib>Jamema, S</creatorcontrib><creatorcontrib>Gupta, Tejpal</creatorcontrib><creatorcontrib>Deepak, D</creatorcontrib><creatorcontrib>Rajeev, S</creatorcontrib><collection>CrossRef</collection><collection>Directory of Open Access Journals</collection><jtitle>Journal of medical physics</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Reena, P</au><au>Dayananda, S</au><au>Pai, Rajeshri</au><au>Jamema, S</au><au>Gupta, Tejpal</au><au>Deepak, D</au><au>Rajeev, S</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Performance characterization of siemens primus linear accelerator under small monitor unit and small segments for the implementation of step-and-shoot intensitymodulated radiotherapy</atitle><jtitle>Journal of medical physics</jtitle><date>2006-10-01</date><risdate>2006</risdate><volume>31</volume><issue>4</issue><spage>269</spage><epage>274</epage><pages>269-274</pages><issn>0971-6203</issn><abstract>Implementation of step-and-shoot intensity-modulated radiotherapy (IMRT) needs careful understanding of the accelerator start-up characteristic to ensure accurate and precise delivery of radiation dose to patient. The dosimetric characteristic of a Siemens Primus linear accelerator (LA) which delivers 6 and 18 MV x-rays at the dose rate of 300 and 500 monitor unit (MU) per minutes (min) respectively was studied under the condition of small MU ranging from 1 to 100. Dose monitor linearity was studied at different dose calibration parameter (D1_C0) by measuring ionization at 10 cm depth in a solid water phantom using a 0.6 cc ionization chamber. Monitor unit stability was studied from different intensity modulated (IM) groups comprising various combinations of MU per field and number of fields. Stability of beam flatness and symmetry was investigated under normal and IMRT mode for 20x20 cm2 field under small MU using a 2D Profiler kept isocentrically at 5 cm depth. Inter segment response was investigated form 1 to 10 MU by measuring the dose per MU from various IM groups, each consisting of four segments with inter-segment separation of 2 cm. In the range 1-4 MU, the dose linearity error was more than 5% (max -32% at 1 MU) for 6 MV x-rays at factory calibrated D1_C0 value of 6000. The dose linearity error was reduced to -10.95% at 1 MU, within -3% for 2 and 3 MU and ±1% for MU ≥4 when the D1_C0 was subsequently tuned at 4500. For 18 MV x-rays, the dose linearity error at factory calibrated D1_C0 value of 4400 was within ±1% for MU ≥ 3 with maximum of -13.5 observed at 1 MU. For both the beam energies and MU/field ≥ 4, the stability of monitor unit tested for different IM groups was within ±1% of the dose from the normal treatment field. This variation increases to -2.6% for 6 MV and -2.7% for 18 MV x-rays for 2 MU/field. No significant variation was observed in the stability of beam profile measured from normal and IMRT mode. The beam flatness was within 3% for 6 MV x-rays and more than 3% (Max 3.5%) for 18 MV x-rays at lesser irradiation time ≤ 3 MU. The beam stability improves with the increase in irradiation time. Both the beam energies show very good symmetry (≤ 2%) at all irradiation time. For all the three segment sizes studied, the nonlinearity was observed at smaller MU/segment in both the energies. When the MU/segment is ≥ 4, all segment size shows fairly linear relation with dose/MU. The smaller segment size shows larger nonlinearity at smaller MU/segment and become more linear at larger MU/segment. Based on our study, we conclude that the Primus LA from Siemens installed at our hospital is ideally suited for step-and-shoot IMRT preferably for radiation ON time ³4MU per segment.</abstract><pub>Medknow Publications and Media Pvt. Ltd</pub><doi>10.4103/0971-6203.29197</doi><tpages>6</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0971-6203
ispartof Journal of medical physics, 2006-10, Vol.31 (4), p.269-274
issn 0971-6203
language eng
recordid cdi_doaj_primary_oai_doaj_org_article_297389fa804a4c70bb0ec18eb3f80139
source Medknow Open Access Medical Journals; IngentaConnect Journals; PubMed Central
subjects Performance characterization
small monitor unit
small segment
step-and-shoot intensity-modulated radiotherapy
title Performance characterization of siemens primus linear accelerator under small monitor unit and small segments for the implementation of step-and-shoot intensitymodulated radiotherapy
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-07T10%3A48%3A38IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_doaj_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Performance%20characterization%20of%20siemens%20primus%20linear%20accelerator%20under%20small%20monitor%20unit%20and%20small%20segments%20for%20the%20implementation%20of%20step-and-shoot%20intensitymodulated%20radiotherapy&rft.jtitle=Journal%20of%20medical%20physics&rft.au=Reena,%20P&rft.date=2006-10-01&rft.volume=31&rft.issue=4&rft.spage=269&rft.epage=274&rft.pages=269-274&rft.issn=0971-6203&rft_id=info:doi/10.4103/0971-6203.29197&rft_dat=%3Cgale_doaj_%3EA156027468%3C/gale_doaj_%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c355t-b7b3ca3f15d5ee9747dc4b78c06c4c9873992556a7ab68f2e1103ad4f81b33843%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rft_galeid=A156027468&rfr_iscdi=true