The effect of breast shielding outside the field of view on breast entrance surface dose in axial X-ray examinations: a phantom study
The purpose of this study was to evaluate the effect of outside-field-of-view (FOV) lead shielding on the entrance surface dose (ESD) of the breast on an anthropomorphic X-ray phantom for a variety of axial skeleton X-ray examinations. Using an anthropomorphic phantom and radiation dosimeter, the ES...
Saved in:
Published in: | Diagnostic and interventional radiology (Ankara, Turkey) Turkey), 2023-05, Vol.29 (3), p.555-560 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | |
---|---|
cites | cdi_FETCH-LOGICAL-c481t-bf2249d54f4dde8b7305a377580cf5a5d8be660244f7c9ae6abd8b2cac383d13 |
container_end_page | 560 |
container_issue | 3 |
container_start_page | 555 |
container_title | Diagnostic and interventional radiology (Ankara, Turkey) |
container_volume | 29 |
creator | Hurley, Lauren Alashban, Yazeed Albeshan, Salman England, Andrew McEntee, Mark F |
description | The purpose of this study was to evaluate the effect of outside-field-of-view (FOV) lead shielding on the entrance surface dose (ESD) of the breast on an anthropomorphic X-ray phantom for a variety of axial skeleton X-ray examinations.
Using an anthropomorphic phantom and radiation dosimeter, the ESD of the breast was measured with and without outside-FOV shielding in anterior-posterior (AP) abdomen, AP cervical spine, occipitomental 30° (OM30) facial bones, AP lumbar spine, and lateral lumbar spine radiography. The effect of several exposure parameters, including a low milliampere-seconds technique, grid use, automatic exposure control use, wraparound lead (WAL) use, trolley use, and X-ray table use, on the ESD of the breast with and without outside-FOV shielding was investigated. The mean ESD (μSv) and standard deviation for each radiographic protocol were calculated. A one-tailed Student's t-test was carried out to evaluate whether ESD to the breast was reduced with the use of outside-FOV shielding.
A total of 920 breast ESD measurements were recorded across the different protocol parameters. The largest decrease in mean ESD of the breast with outside-FOV shielding was 0.002 μSv (
= 0.084), recorded in the AP abdomen on the table with a grid, OM30 on the table with a grid, OM30 standard protocol on the trolley, and OM30 on the trolley with WAL protocols. This decrease was found to be statistically non-significant.
This study found no significant decrease in the ESD of the breast with the use of outside-FOV shielding for the AP abdomen, AP cervical spine, OM30 facial bones, AP lumbar spine, or lateral lumbar spine radiography across a range of protocols. |
doi_str_mv | 10.4274/dir.2023.232126 |
format | article |
fullrecord | <record><control><sourceid>gale_doaj_</sourceid><recordid>TN_cdi_doaj_primary_oai_doaj_org_article_38929d52762941ad8a8001ae4b07a9fd</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A752011494</galeid><doaj_id>oai_doaj_org_article_38929d52762941ad8a8001ae4b07a9fd</doaj_id><sourcerecordid>A752011494</sourcerecordid><originalsourceid>FETCH-LOGICAL-c481t-bf2249d54f4dde8b7305a377580cf5a5d8be660244f7c9ae6abd8b2cac383d13</originalsourceid><addsrcrecordid>eNptkk1vEzEQhlcIREvhzA1Z4sIlqb_2iwuqqgKVKnHJgZs1a48TV7t2sHfb5gfwv-ttmopIyIexxs-8nrHfovjI6FLyWp4bF5eccrHkgjNevSpOmaDlQlSMvz7sG16eFO9SuqW0LFsm3xYnoma8FZSdFn9XGyRoLeqRBEu6iJBGkjYOe-P8moRpTM4gGTNm5-RM3Tm8J8EfaPRjBK-RpClayNGEhMR5Ag8OevJ7EWFH8AEG52F0waevBMh2A34MA0njZHbvizcW-oQfnuNZsfp-tbr8ubj59eP68uJmoWXDxkVnOZetKaWVxmDT1Xk-EHVdNlTbEkrTdFhVlEtpa90CVtDlFNegRSMME2fF9V7WBLhV2-gGiDsVwKmnRIhrBXF0ukclmpbnm3hd8VYyMA00lDJA2dEaWmuy1re91nbqBjT66RH6I9HjE-82ah3uFKNV3Va0ygpfnhVi-DNhGtXgksa-B49hSoo3tJn_q-YZ_bxH15B7c96GLKlnXF3UJaeMyVZmavkfKi-Dg9PBo3U5f1Rwvi_QMaQU0b60z6ia_aWyv9TsL7X3V6749O_UL_zBUOIRrZfMfg</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2808591472</pqid></control><display><type>article</type><title>The effect of breast shielding outside the field of view on breast entrance surface dose in axial X-ray examinations: a phantom study</title><source>PubMed Central</source><source>Directory of Open Access Journals</source><creator>Hurley, Lauren ; Alashban, Yazeed ; Albeshan, Salman ; England, Andrew ; McEntee, Mark F</creator><creatorcontrib>Hurley, Lauren ; Alashban, Yazeed ; Albeshan, Salman ; England, Andrew ; McEntee, Mark F</creatorcontrib><description>The purpose of this study was to evaluate the effect of outside-field-of-view (FOV) lead shielding on the entrance surface dose (ESD) of the breast on an anthropomorphic X-ray phantom for a variety of axial skeleton X-ray examinations.
Using an anthropomorphic phantom and radiation dosimeter, the ESD of the breast was measured with and without outside-FOV shielding in anterior-posterior (AP) abdomen, AP cervical spine, occipitomental 30° (OM30) facial bones, AP lumbar spine, and lateral lumbar spine radiography. The effect of several exposure parameters, including a low milliampere-seconds technique, grid use, automatic exposure control use, wraparound lead (WAL) use, trolley use, and X-ray table use, on the ESD of the breast with and without outside-FOV shielding was investigated. The mean ESD (μSv) and standard deviation for each radiographic protocol were calculated. A one-tailed Student's t-test was carried out to evaluate whether ESD to the breast was reduced with the use of outside-FOV shielding.
A total of 920 breast ESD measurements were recorded across the different protocol parameters. The largest decrease in mean ESD of the breast with outside-FOV shielding was 0.002 μSv (
= 0.084), recorded in the AP abdomen on the table with a grid, OM30 on the table with a grid, OM30 standard protocol on the trolley, and OM30 on the trolley with WAL protocols. This decrease was found to be statistically non-significant.
This study found no significant decrease in the ESD of the breast with the use of outside-FOV shielding for the AP abdomen, AP cervical spine, OM30 facial bones, AP lumbar spine, or lateral lumbar spine radiography across a range of protocols.</description><identifier>ISSN: 1305-3825</identifier><identifier>EISSN: 1305-3612</identifier><identifier>DOI: 10.4274/dir.2023.232126</identifier><identifier>PMID: 37129301</identifier><language>eng</language><publisher>Turkey: Galenos Yayinevi Tic. Ltd</publisher><subject>anthropomorphic phantom ; Breast ; entrance surface dose ; field of view ; Health aspects ; Humans ; Lumbar Vertebrae ; Methods ; Phantoms, Imaging ; Physiological aspects ; Radiation Dosage ; Radiation dosimetry ; Radiography ; Radiology Physics - Original ; shielding ; Thorax ; wraparound lead ; X-Rays</subject><ispartof>Diagnostic and interventional radiology (Ankara, Turkey), 2023-05, Vol.29 (3), p.555-560</ispartof><rights>COPYRIGHT 2023 Galenos Yayinevi Tic. Ltd.</rights><rights>Copyright 2023 by Turkish Society of Radiology | Diagnostic and Interventional Radiology, published by Galenos Publishing House. 2023</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c481t-bf2249d54f4dde8b7305a377580cf5a5d8be660244f7c9ae6abd8b2cac383d13</cites><orcidid>0000-0003-0857-524X ; 0000-0001-7045-2190 ; 0000-0002-8612-1721</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC10679606/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC10679606/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,864,885,2102,27924,27925,53791,53793</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/37129301$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Hurley, Lauren</creatorcontrib><creatorcontrib>Alashban, Yazeed</creatorcontrib><creatorcontrib>Albeshan, Salman</creatorcontrib><creatorcontrib>England, Andrew</creatorcontrib><creatorcontrib>McEntee, Mark F</creatorcontrib><title>The effect of breast shielding outside the field of view on breast entrance surface dose in axial X-ray examinations: a phantom study</title><title>Diagnostic and interventional radiology (Ankara, Turkey)</title><addtitle>Diagn Interv Radiol</addtitle><description>The purpose of this study was to evaluate the effect of outside-field-of-view (FOV) lead shielding on the entrance surface dose (ESD) of the breast on an anthropomorphic X-ray phantom for a variety of axial skeleton X-ray examinations.
Using an anthropomorphic phantom and radiation dosimeter, the ESD of the breast was measured with and without outside-FOV shielding in anterior-posterior (AP) abdomen, AP cervical spine, occipitomental 30° (OM30) facial bones, AP lumbar spine, and lateral lumbar spine radiography. The effect of several exposure parameters, including a low milliampere-seconds technique, grid use, automatic exposure control use, wraparound lead (WAL) use, trolley use, and X-ray table use, on the ESD of the breast with and without outside-FOV shielding was investigated. The mean ESD (μSv) and standard deviation for each radiographic protocol were calculated. A one-tailed Student's t-test was carried out to evaluate whether ESD to the breast was reduced with the use of outside-FOV shielding.
A total of 920 breast ESD measurements were recorded across the different protocol parameters. The largest decrease in mean ESD of the breast with outside-FOV shielding was 0.002 μSv (
= 0.084), recorded in the AP abdomen on the table with a grid, OM30 on the table with a grid, OM30 standard protocol on the trolley, and OM30 on the trolley with WAL protocols. This decrease was found to be statistically non-significant.
This study found no significant decrease in the ESD of the breast with the use of outside-FOV shielding for the AP abdomen, AP cervical spine, OM30 facial bones, AP lumbar spine, or lateral lumbar spine radiography across a range of protocols.</description><subject>anthropomorphic phantom</subject><subject>Breast</subject><subject>entrance surface dose</subject><subject>field of view</subject><subject>Health aspects</subject><subject>Humans</subject><subject>Lumbar Vertebrae</subject><subject>Methods</subject><subject>Phantoms, Imaging</subject><subject>Physiological aspects</subject><subject>Radiation Dosage</subject><subject>Radiation dosimetry</subject><subject>Radiography</subject><subject>Radiology Physics - Original</subject><subject>shielding</subject><subject>Thorax</subject><subject>wraparound lead</subject><subject>X-Rays</subject><issn>1305-3825</issn><issn>1305-3612</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNptkk1vEzEQhlcIREvhzA1Z4sIlqb_2iwuqqgKVKnHJgZs1a48TV7t2sHfb5gfwv-ttmopIyIexxs-8nrHfovjI6FLyWp4bF5eccrHkgjNevSpOmaDlQlSMvz7sG16eFO9SuqW0LFsm3xYnoma8FZSdFn9XGyRoLeqRBEu6iJBGkjYOe-P8moRpTM4gGTNm5-RM3Tm8J8EfaPRjBK-RpClayNGEhMR5Ag8OevJ7EWFH8AEG52F0waevBMh2A34MA0njZHbvizcW-oQfnuNZsfp-tbr8ubj59eP68uJmoWXDxkVnOZetKaWVxmDT1Xk-EHVdNlTbEkrTdFhVlEtpa90CVtDlFNegRSMME2fF9V7WBLhV2-gGiDsVwKmnRIhrBXF0ukclmpbnm3hd8VYyMA00lDJA2dEaWmuy1re91nbqBjT66RH6I9HjE-82ah3uFKNV3Va0ygpfnhVi-DNhGtXgksa-B49hSoo3tJn_q-YZ_bxH15B7c96GLKlnXF3UJaeMyVZmavkfKi-Dg9PBo3U5f1Rwvi_QMaQU0b60z6ia_aWyv9TsL7X3V6749O_UL_zBUOIRrZfMfg</recordid><startdate>20230501</startdate><enddate>20230501</enddate><creator>Hurley, Lauren</creator><creator>Alashban, Yazeed</creator><creator>Albeshan, Salman</creator><creator>England, Andrew</creator><creator>McEntee, Mark F</creator><general>Galenos Yayinevi Tic. Ltd</general><general>Galenos Publishing</general><general>Galenos Publishing House</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><scope>DOA</scope><orcidid>https://orcid.org/0000-0003-0857-524X</orcidid><orcidid>https://orcid.org/0000-0001-7045-2190</orcidid><orcidid>https://orcid.org/0000-0002-8612-1721</orcidid></search><sort><creationdate>20230501</creationdate><title>The effect of breast shielding outside the field of view on breast entrance surface dose in axial X-ray examinations: a phantom study</title><author>Hurley, Lauren ; Alashban, Yazeed ; Albeshan, Salman ; England, Andrew ; McEntee, Mark F</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c481t-bf2249d54f4dde8b7305a377580cf5a5d8be660244f7c9ae6abd8b2cac383d13</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><topic>anthropomorphic phantom</topic><topic>Breast</topic><topic>entrance surface dose</topic><topic>field of view</topic><topic>Health aspects</topic><topic>Humans</topic><topic>Lumbar Vertebrae</topic><topic>Methods</topic><topic>Phantoms, Imaging</topic><topic>Physiological aspects</topic><topic>Radiation Dosage</topic><topic>Radiation dosimetry</topic><topic>Radiography</topic><topic>Radiology Physics - Original</topic><topic>shielding</topic><topic>Thorax</topic><topic>wraparound lead</topic><topic>X-Rays</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Hurley, Lauren</creatorcontrib><creatorcontrib>Alashban, Yazeed</creatorcontrib><creatorcontrib>Albeshan, Salman</creatorcontrib><creatorcontrib>England, Andrew</creatorcontrib><creatorcontrib>McEntee, Mark F</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><collection>Directory of Open Access Journals</collection><jtitle>Diagnostic and interventional radiology (Ankara, Turkey)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Hurley, Lauren</au><au>Alashban, Yazeed</au><au>Albeshan, Salman</au><au>England, Andrew</au><au>McEntee, Mark F</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>The effect of breast shielding outside the field of view on breast entrance surface dose in axial X-ray examinations: a phantom study</atitle><jtitle>Diagnostic and interventional radiology (Ankara, Turkey)</jtitle><addtitle>Diagn Interv Radiol</addtitle><date>2023-05-01</date><risdate>2023</risdate><volume>29</volume><issue>3</issue><spage>555</spage><epage>560</epage><pages>555-560</pages><issn>1305-3825</issn><eissn>1305-3612</eissn><abstract>The purpose of this study was to evaluate the effect of outside-field-of-view (FOV) lead shielding on the entrance surface dose (ESD) of the breast on an anthropomorphic X-ray phantom for a variety of axial skeleton X-ray examinations.
Using an anthropomorphic phantom and radiation dosimeter, the ESD of the breast was measured with and without outside-FOV shielding in anterior-posterior (AP) abdomen, AP cervical spine, occipitomental 30° (OM30) facial bones, AP lumbar spine, and lateral lumbar spine radiography. The effect of several exposure parameters, including a low milliampere-seconds technique, grid use, automatic exposure control use, wraparound lead (WAL) use, trolley use, and X-ray table use, on the ESD of the breast with and without outside-FOV shielding was investigated. The mean ESD (μSv) and standard deviation for each radiographic protocol were calculated. A one-tailed Student's t-test was carried out to evaluate whether ESD to the breast was reduced with the use of outside-FOV shielding.
A total of 920 breast ESD measurements were recorded across the different protocol parameters. The largest decrease in mean ESD of the breast with outside-FOV shielding was 0.002 μSv (
= 0.084), recorded in the AP abdomen on the table with a grid, OM30 on the table with a grid, OM30 standard protocol on the trolley, and OM30 on the trolley with WAL protocols. This decrease was found to be statistically non-significant.
This study found no significant decrease in the ESD of the breast with the use of outside-FOV shielding for the AP abdomen, AP cervical spine, OM30 facial bones, AP lumbar spine, or lateral lumbar spine radiography across a range of protocols.</abstract><cop>Turkey</cop><pub>Galenos Yayinevi Tic. Ltd</pub><pmid>37129301</pmid><doi>10.4274/dir.2023.232126</doi><tpages>6</tpages><orcidid>https://orcid.org/0000-0003-0857-524X</orcidid><orcidid>https://orcid.org/0000-0001-7045-2190</orcidid><orcidid>https://orcid.org/0000-0002-8612-1721</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1305-3825 |
ispartof | Diagnostic and interventional radiology (Ankara, Turkey), 2023-05, Vol.29 (3), p.555-560 |
issn | 1305-3825 1305-3612 |
language | eng |
recordid | cdi_doaj_primary_oai_doaj_org_article_38929d52762941ad8a8001ae4b07a9fd |
source | PubMed Central; Directory of Open Access Journals |
subjects | anthropomorphic phantom Breast entrance surface dose field of view Health aspects Humans Lumbar Vertebrae Methods Phantoms, Imaging Physiological aspects Radiation Dosage Radiation dosimetry Radiography Radiology Physics - Original shielding Thorax wraparound lead X-Rays |
title | The effect of breast shielding outside the field of view on breast entrance surface dose in axial X-ray examinations: a phantom study |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-26T22%3A01%3A36IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_doaj_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=The%20effect%20of%20breast%20shielding%20outside%20the%20field%20of%20view%20on%20breast%20entrance%20surface%20dose%20in%20axial%20X-ray%20examinations:%20a%20phantom%20study&rft.jtitle=Diagnostic%20and%20interventional%20radiology%20(Ankara,%20Turkey)&rft.au=Hurley,%20Lauren&rft.date=2023-05-01&rft.volume=29&rft.issue=3&rft.spage=555&rft.epage=560&rft.pages=555-560&rft.issn=1305-3825&rft.eissn=1305-3612&rft_id=info:doi/10.4274/dir.2023.232126&rft_dat=%3Cgale_doaj_%3EA752011494%3C/gale_doaj_%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c481t-bf2249d54f4dde8b7305a377580cf5a5d8be660244f7c9ae6abd8b2cac383d13%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2808591472&rft_id=info:pmid/37129301&rft_galeid=A752011494&rfr_iscdi=true |