Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus
A portable analytical chemistry analyzer was used to make field assessments of wild harlequin ducks (Histrionicus histrionicus) in association with telemetry studies of winter survival in Prince William Sound, Alaska. We compared serum chemistry results obtained on-site with results from a tradition...
Saved in:
Published in: | Veterinary Medicine International 2010, Vol.2010 (2010), p.105-109 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
cited_by | cdi_FETCH-LOGICAL-a5156-e3830589923be980558b5d0b33f55cd1b7e5daf9795388e8dfbea079c7a63da33 |
---|---|
cites | cdi_FETCH-LOGICAL-a5156-e3830589923be980558b5d0b33f55cd1b7e5daf9795388e8dfbea079c7a63da33 |
container_end_page | 109 |
container_issue | 2010 |
container_start_page | 105 |
container_title | Veterinary Medicine International |
container_volume | 2010 |
creator | Stoskopf, Michael K. Mulcahy, Daniel M. Esler, Daniel |
description | A portable analytical chemistry analyzer was used to make field assessments of wild harlequin ducks (Histrionicus histrionicus) in association with telemetry studies of winter survival in Prince William Sound, Alaska. We compared serum chemistry results obtained on-site with results from a traditional laboratory. Particular attention was paid to serum glucose and potassium concentrations as potential indicators of high-risk surgical candidates based on evaluation of the field data. The median differential for glucose values (N=82) between methods was 0.6 mmol/L (quartiles 0.3 and 0.9 mmol/L) with the median value higher when assayed on site. Analysis of potassium on site returned a median of 2.7 mmol/L (N=88 ; quartiles 2.4 and 3.0 mmol/L). Serum potassium values were too low for quantitation by the traditional laboratory. Changes in several serum chemistry values following a three-day storm during the study support the value of on site evaluation of serum potassium to identify presurgical patients with increased anesthetic risk. |
doi_str_mv | 10.4061/2010/418596 |
format | article |
fullrecord | <record><control><sourceid>gale_doaj_</sourceid><recordid>TN_cdi_doaj_primary_oai_doaj_org_article_3df0596964cb4ee2a283fdaa12e6f60b</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A262885808</galeid><airiti_id>P20150803003_201012_201704250005_201704250005_105_109</airiti_id><doaj_id>oai_doaj_org_article_3df0596964cb4ee2a283fdaa12e6f60b</doaj_id><sourcerecordid>A262885808</sourcerecordid><originalsourceid>FETCH-LOGICAL-a5156-e3830589923be980558b5d0b33f55cd1b7e5daf9795388e8dfbea079c7a63da33</originalsourceid><addsrcrecordid>eNqFks9v0zAUxyMEYlPZiTPIEgckoMw_4sS5IFVlo5MmMQk4Wy_Jc-uSxJudDJUDfztOM6r1RCTLcd4nHz9b3yR5yejHlGbsnFNGz1OmZJE9SU45Tfmc0lQ9Hd8LOleMiZPkLIQtjY_gBRfF8-QkcqnMlThN_lzcQzNAb11HnCFAbpzvoWyQLIbetdBjTb6hH1qy3GBrQ-93ZNFBs_uNnhjnyaXFpiaLEDCEFrt-tKzAN3g32I58Hqqf4QNZjT_GLWw1BLJ5tHiRPDPQBDx7mGfJj8uL78vV_Prrl6vl4noOkslsjkIJKlUR2y-xUFRKVcqalkIYKaualTnKGkyRF1Iohao2JQLNiyqHTNQgxCy5mry1g62-9bYFv9MOrN5_cH6twfe2alCL2tB4m0WWVmWKyIErYWoAxjEzWdxzlnyaXLdD2WJdxUN7aI6kx5XObvTa3WuuMsoKFQVvHwTe3Q0Yeh0vtsKmgQ7dEHQuREqVlHkk30zkGmJntjMuCquR1guecaWkoqPv_URV3oXg0Rx6YVSPKdFjSvSUkki_ftz-gf2XiQi8m4CN7Wr4Zf9jezXBGBE0cIDTnGYyjfXrqQ7W297qrRt8jE_QN9EiqaIixnJvZHyc8phgGaMqjxdsPwrxF67Q41c</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>733408557</pqid></control><display><type>article</type><title>Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus</title><source>Open Access: Wiley-Blackwell Open Access Journals</source><source>PubMed Central(OpenAccess)</source><creator>Stoskopf, Michael K. ; Mulcahy, Daniel M. ; Esler, Daniel</creator><contributor>Benazzi, Cinzia</contributor><creatorcontrib>Stoskopf, Michael K. ; Mulcahy, Daniel M. ; Esler, Daniel ; Benazzi, Cinzia</creatorcontrib><description>A portable analytical chemistry analyzer was used to make field assessments of wild harlequin ducks (Histrionicus histrionicus) in association with telemetry studies of winter survival in Prince William Sound, Alaska. We compared serum chemistry results obtained on-site with results from a traditional laboratory. Particular attention was paid to serum glucose and potassium concentrations as potential indicators of high-risk surgical candidates based on evaluation of the field data. The median differential for glucose values (N=82) between methods was 0.6 mmol/L (quartiles 0.3 and 0.9 mmol/L) with the median value higher when assayed on site. Analysis of potassium on site returned a median of 2.7 mmol/L (N=88 ; quartiles 2.4 and 3.0 mmol/L). Serum potassium values were too low for quantitation by the traditional laboratory. Changes in several serum chemistry values following a three-day storm during the study support the value of on site evaluation of serum potassium to identify presurgical patients with increased anesthetic risk.</description><identifier>ISSN: 2090-8113</identifier><identifier>ISSN: 2042-0048</identifier><identifier>EISSN: 2042-0048</identifier><identifier>DOI: 10.4061/2010/418596</identifier><identifier>PMID: 20445783</identifier><language>eng</language><publisher>Cairo, Egypt: Hindawi Limiteds</publisher><subject>Anesthetics ; Dextrose ; Glucose ; Physiological aspects ; Potassium acetate ; Risk factors</subject><ispartof>Veterinary Medicine International, 2010, Vol.2010 (2010), p.105-109</ispartof><rights>Copyright © 2010</rights><rights>COPYRIGHT 2010 John Wiley & Sons, Inc.</rights><rights>Copyright © 2010 Michael K. Stoskopf et al. 2010</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-a5156-e3830589923be980558b5d0b33f55cd1b7e5daf9795388e8dfbea079c7a63da33</citedby><cites>FETCH-LOGICAL-a5156-e3830589923be980558b5d0b33f55cd1b7e5daf9795388e8dfbea079c7a63da33</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC2860198/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC2860198/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,885,4024,27923,27924,27925,53791,53793</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/20445783$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><contributor>Benazzi, Cinzia</contributor><creatorcontrib>Stoskopf, Michael K.</creatorcontrib><creatorcontrib>Mulcahy, Daniel M.</creatorcontrib><creatorcontrib>Esler, Daniel</creatorcontrib><title>Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus</title><title>Veterinary Medicine International</title><addtitle>Vet Med Int</addtitle><description>A portable analytical chemistry analyzer was used to make field assessments of wild harlequin ducks (Histrionicus histrionicus) in association with telemetry studies of winter survival in Prince William Sound, Alaska. We compared serum chemistry results obtained on-site with results from a traditional laboratory. Particular attention was paid to serum glucose and potassium concentrations as potential indicators of high-risk surgical candidates based on evaluation of the field data. The median differential for glucose values (N=82) between methods was 0.6 mmol/L (quartiles 0.3 and 0.9 mmol/L) with the median value higher when assayed on site. Analysis of potassium on site returned a median of 2.7 mmol/L (N=88 ; quartiles 2.4 and 3.0 mmol/L). Serum potassium values were too low for quantitation by the traditional laboratory. Changes in several serum chemistry values following a three-day storm during the study support the value of on site evaluation of serum potassium to identify presurgical patients with increased anesthetic risk.</description><subject>Anesthetics</subject><subject>Dextrose</subject><subject>Glucose</subject><subject>Physiological aspects</subject><subject>Potassium acetate</subject><subject>Risk factors</subject><issn>2090-8113</issn><issn>2042-0048</issn><issn>2042-0048</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2010</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNqFks9v0zAUxyMEYlPZiTPIEgckoMw_4sS5IFVlo5MmMQk4Wy_Jc-uSxJudDJUDfztOM6r1RCTLcd4nHz9b3yR5yejHlGbsnFNGz1OmZJE9SU45Tfmc0lQ9Hd8LOleMiZPkLIQtjY_gBRfF8-QkcqnMlThN_lzcQzNAb11HnCFAbpzvoWyQLIbetdBjTb6hH1qy3GBrQ-93ZNFBs_uNnhjnyaXFpiaLEDCEFrt-tKzAN3g32I58Hqqf4QNZjT_GLWw1BLJ5tHiRPDPQBDx7mGfJj8uL78vV_Prrl6vl4noOkslsjkIJKlUR2y-xUFRKVcqalkIYKaualTnKGkyRF1Iohao2JQLNiyqHTNQgxCy5mry1g62-9bYFv9MOrN5_cH6twfe2alCL2tB4m0WWVmWKyIErYWoAxjEzWdxzlnyaXLdD2WJdxUN7aI6kx5XObvTa3WuuMsoKFQVvHwTe3Q0Yeh0vtsKmgQ7dEHQuREqVlHkk30zkGmJntjMuCquR1guecaWkoqPv_URV3oXg0Rx6YVSPKdFjSvSUkki_ftz-gf2XiQi8m4CN7Wr4Zf9jezXBGBE0cIDTnGYyjfXrqQ7W297qrRt8jE_QN9EiqaIixnJvZHyc8phgGaMqjxdsPwrxF67Q41c</recordid><startdate>2010</startdate><enddate>2010</enddate><creator>Stoskopf, Michael K.</creator><creator>Mulcahy, Daniel M.</creator><creator>Esler, Daniel</creator><general>Hindawi Limiteds</general><general>Hindawi Puplishing Corporation</general><general>SAGE-Hindawi Access to Research</general><general>John Wiley & Sons, Inc</general><general>Hindawi Limited</general><scope>188</scope><scope>ADJCN</scope><scope>AHFXO</scope><scope>RHU</scope><scope>RHW</scope><scope>RHX</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><scope>DOA</scope></search><sort><creationdate>2010</creationdate><title>Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus</title><author>Stoskopf, Michael K. ; Mulcahy, Daniel M. ; Esler, Daniel</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-a5156-e3830589923be980558b5d0b33f55cd1b7e5daf9795388e8dfbea079c7a63da33</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2010</creationdate><topic>Anesthetics</topic><topic>Dextrose</topic><topic>Glucose</topic><topic>Physiological aspects</topic><topic>Potassium acetate</topic><topic>Risk factors</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Stoskopf, Michael K.</creatorcontrib><creatorcontrib>Mulcahy, Daniel M.</creatorcontrib><creatorcontrib>Esler, Daniel</creatorcontrib><collection>Airiti Library</collection><collection>الدوريات العلمية والإحصائية - e-Marefa Academic and Statistical Periodicals</collection><collection>معرفة - المحتوى العربي الأكاديمي المتكامل - e-Marefa Academic Complete</collection><collection>Hindawi Publishing Complete</collection><collection>Hindawi Publishing Subscription Journals</collection><collection>Hindawi Publishing Open Access</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><collection>Directory of Open Access Journals(OpenAccess)</collection><jtitle>Veterinary Medicine International</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Stoskopf, Michael K.</au><au>Mulcahy, Daniel M.</au><au>Esler, Daniel</au><au>Benazzi, Cinzia</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus</atitle><jtitle>Veterinary Medicine International</jtitle><addtitle>Vet Med Int</addtitle><date>2010</date><risdate>2010</risdate><volume>2010</volume><issue>2010</issue><spage>105</spage><epage>109</epage><pages>105-109</pages><issn>2090-8113</issn><issn>2042-0048</issn><eissn>2042-0048</eissn><abstract>A portable analytical chemistry analyzer was used to make field assessments of wild harlequin ducks (Histrionicus histrionicus) in association with telemetry studies of winter survival in Prince William Sound, Alaska. We compared serum chemistry results obtained on-site with results from a traditional laboratory. Particular attention was paid to serum glucose and potassium concentrations as potential indicators of high-risk surgical candidates based on evaluation of the field data. The median differential for glucose values (N=82) between methods was 0.6 mmol/L (quartiles 0.3 and 0.9 mmol/L) with the median value higher when assayed on site. Analysis of potassium on site returned a median of 2.7 mmol/L (N=88 ; quartiles 2.4 and 3.0 mmol/L). Serum potassium values were too low for quantitation by the traditional laboratory. Changes in several serum chemistry values following a three-day storm during the study support the value of on site evaluation of serum potassium to identify presurgical patients with increased anesthetic risk.</abstract><cop>Cairo, Egypt</cop><pub>Hindawi Limiteds</pub><pmid>20445783</pmid><doi>10.4061/2010/418596</doi><tpages>5</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 2090-8113 |
ispartof | Veterinary Medicine International, 2010, Vol.2010 (2010), p.105-109 |
issn | 2090-8113 2042-0048 2042-0048 |
language | eng |
recordid | cdi_doaj_primary_oai_doaj_org_article_3df0596964cb4ee2a283fdaa12e6f60b |
source | Open Access: Wiley-Blackwell Open Access Journals; PubMed Central(OpenAccess) |
subjects | Anesthetics Dextrose Glucose Physiological aspects Potassium acetate Risk factors |
title | Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus |
url | http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-29T04%3A26%3A02IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_doaj_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Evaluation%20of%20a%20Portable%20Automated%20Serum%20Chemistry%20Analyzer%20for%20Field%20Assessment%20of%20Harlequin%20Ducks,%20Histrionicus%20histrionicus&rft.jtitle=Veterinary%20Medicine%20International&rft.au=Stoskopf,%20Michael%20K.&rft.date=2010&rft.volume=2010&rft.issue=2010&rft.spage=105&rft.epage=109&rft.pages=105-109&rft.issn=2090-8113&rft.eissn=2042-0048&rft_id=info:doi/10.4061/2010/418596&rft_dat=%3Cgale_doaj_%3EA262885808%3C/gale_doaj_%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-a5156-e3830589923be980558b5d0b33f55cd1b7e5daf9795388e8dfbea079c7a63da33%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=733408557&rft_id=info:pmid/20445783&rft_galeid=A262885808&rft_airiti_id=P20150803003_201012_201704250005_201704250005_105_109&rfr_iscdi=true |