Loading…

Ultrasonic Transducer Peak-to-Peak Optical Measurement

Possible optical setups for measurement of the peak-to-peak value of an ultrasonic transducer are described in this work. The Michelson interferometer with the calibrated nanopositioner in reference path and laser Doppler vibrometer were used for the basic measurement of vibration displacement. Lang...

Full description

Saved in:
Bibliographic Details
Published in:Advances in electrical and electronic engineering 2012-06, Vol.10 (2), p.125-129
Main Authors: Skarvada, Pavel, Tofel, Pavel, Tomanek, Pavel
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Possible optical setups for measurement of the peak-to-peak value of an ultrasonic transducer are described in this work. The Michelson interferometer with the calibrated nanopositioner in reference path and laser Doppler vibrometer were used for the basic measurement of vibration displacement. Langevin type of ultrasonic transducer is used for the purposes of Electro-Ultrasonic Nonlinear Spectroscopy (EUNS). Parameters of produced mechanical vibration have to been well known for EUNS. Moreover, a monitoring of mechanical vibration frequency shift with a mass load and sample-transducer coupling is important for EUNS measurement.
ISSN:1336-1376
1804-3119
DOI:10.15598/aeee.v10i2.618