Loading…
Determination of trace elements in refined gold samples by inductively coupled plasma atomic emission spectrometry
This paper presents a method for determination the trace contents of silver, copper, iron, palladium, zinc and platinum in refined gold samples. Simultaneous inductively coupled plasma atomic emission spectrometer with radial torch position and cross flow nebulizer was used for determination. In ord...
Saved in:
Published in: | Journal of the Serbian Chemical Society 2013, Vol.78 (4), p.565-577 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | This paper presents a method for determination the trace contents of silver,
copper, iron, palladium, zinc and platinum in refined gold samples.
Simultaneous inductively coupled plasma atomic emission spectrometer with
radial torch position and cross flow nebulizer was used for determination. In
order to compare the different calibration strategies, two sets of
calibration standards were prepared. The first set was based on matrix
matched calibration standards and the second was prepared without the
addition of matrix material. Detection limits for matrix matching
calibrations were higher for some elements than those without matrix
matching. In addition, the internal standardization method was applied and
experiments indicated that indium was the best option as internal standard.
The obtained results for gold sample by matrix matching and matrix free
calibrations were compared with the obtained results by standard addition
method. The accuracy of the methods was tested performing recovery test.
Recoveries for spiked sample were in the range of 90-115 %. The accuracy of
the methods was also tested by analysis of certified reference material of
high pure goldAuGHP1. The best results were achieved by matrix free
calibration and standard addition method using indium as internal standard at
wavelength of 230 nm. |
---|---|
ISSN: | 0352-5139 1820-7421 |
DOI: | 10.2298/JSC120505135S |