Reliability, repeatability, and accordance between three different corneal diagnostic imaging devices for evaluating the ocular surface
PurposeTo evaluate repeatability, reproducibility, and accordance between ocular surface measurements within three different imaging devices.MethodsWe performed an observational study on 66 healthy eyes. Tear meniscus height, non-invasive tear break-up time (NITBUT) and meibography were measured usi...
Saved in:
Published in: | Frontiers in medicine 2022-07, Vol.9, p.893688-893688 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | PurposeTo evaluate repeatability, reproducibility, and accordance between ocular surface measurements within three different imaging devices.MethodsWe performed an observational study on 66 healthy eyes. Tear meniscus height, non-invasive tear break-up time (NITBUT) and meibography were measured using three corneal imaging devices: Keratograph 5M (Oculus, Wetzlar, Germany), Antares (Lumenis, Sidney, Australia), and LacryDiag (Quantel Medical, Cournon d'Auvergne, France). One-way ANOVAs with post hoc analyses were used to calculate accordance between the tear meniscus and NITBUT. Reproducibility was assessed through coefficients of variation and repeatability with intraclass correlation coefficients (ICC). Reliability of meibography classification was analyzed by calculating Fleiss' Kappa Index and presented in Venn diagrams.ResultsCoefficients of variation were high and differed greatly depending on the device and measurement. ICCs showed moderate reliability of NITBUT and tear meniscus height measurements. We observed discordance between measurements of tear meniscus height between the three devices, F2, 195 = 15.24, p < 0.01. Measurements performed with Antares were higher; 0.365 ± 0.0851, than those with Keratograph 5M and LacryDiag; 0.293 ± 0.0790 and 0.306 ± 0.0731. NITBUT also showed discordance between devices, F2, 111 = 13.152, p < 0.01. Measurements performed with LacryDiag were lower (10.4 ± 1.82) compared to those of Keratograph 5M (12.6 ± 4.01) and Antares (12.6 ± 4.21). Fleiss' Kappa showed a value of -0.00487 for upper lid and 0.128 for inferior lid Meibography classification, suggesting discrete to poor agreement between measurements.ConclusionDepending on the device used and parameter analyzed, measurements varied between each other, showing a difference in image processing. |
---|---|
ISSN: | 2296-858X 2296-858X |
DOI: | 10.3389/fmed.2022.893688 |