Loading…
Analysis of the Biochemical Reaction Status by Real-Time Monitoring Molecular Diffusion Behaviors Using a Transistor Biosensor Integrated with a Microfluidic Channel
Traditional methods of monitoring biochemical reactions measure certain detectable reagents or products while assuming that the undetectable species follow the stoichiometry of the reactions. Here, based upon the metal-oxide thin-film transistor (TFT) biosensor, we develop a real-time molecular diff...
Saved in:
Published in: | ACS omega 2021-05, Vol.6 (18), p.11911-11917 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Traditional methods of monitoring biochemical reactions measure certain detectable reagents or products while assuming that the undetectable species follow the stoichiometry of the reactions. Here, based upon the metal-oxide thin-film transistor (TFT) biosensor, we develop a real-time molecular diffusion model to benchmark the concentration of the reagents and products. Using the nicotinamide adenine dinucleotide (NADH)–oxaloacetic acid with the enzyme of malate dehydrogenase as an example, mixtures of different reagent concentrations were characterized to extract the ratio of remaining concentrations between NAD+ and NADH. We can thus obtain the apparent equilibrium constant of the reaction, (8.06 ± 0.61) × 104. Because the whole analysis was conducted using a TFT sensor fabricated using a semiconductor process, our approach has the advantages of exploring biochemical reaction kinetics in a massively parallel manner. |
---|---|
ISSN: | 2470-1343 2470-1343 |
DOI: | 10.1021/acsomega.1c00222 |