Comparison of central corneal thickness measurements with three different optical devices
Purpose: The purpose of this study was to compare and evaluate the agreement of central corneal thickness (CCT) values obtained with three different devices working according to optical principle in healthy eyes. Methods: 60 eyes of 60 individuals (30 men and 30 women) were enrolled in this study. C...
Saved in:
Published in: | Therapeutic advances in ophthalmology 2021-01, Vol.13, p.2515841421995633-2515841421995633 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Purpose:
The purpose of this study was to compare and evaluate the agreement of central corneal thickness (CCT) values obtained with three different devices working according to optical principle in healthy eyes.
Methods:
60 eyes of 60 individuals (30 men and 30 women) were enrolled in this study. CCT measurements performed with Scheimpflug–Placido topographer (Sirius), spectral-domain optical coherence tomography (RTVue) with an anterior segment module, and optical biometer (AL-Scan) were compared. Bland–Altman analysis was used to demonstrate agreement between methods.
Results:
The mean age was 30.07 ± 7.313 years (range, 18–47 years). The mean CCT values obtained by RTVue, Sirius, and AL-Scan were 518.25 ± 36.38 µm, 526.08 ± 36.33 µm, and 513.50 ± 39.09 µm, respectively. The mean differences in CCT were 7.83 ± 14.15 µm between Sirius and RTVue, 12.58 ± 11.87 µm between Sirius and AL-Scan, and 4.75 ± 4.50 µm between RTVue and AL-Scan. The mean CCT was statistically different among the three groups (p |
---|---|