The qualitative determination of oseltamivir phosphate in Tamiflu® capsule by cyclic voltammetry
A gold electrode was applied in the voltammetric determination of oseltamivir phosphate standard in 0.05 M NaHCO3. Oseltamivir phosphate as a standard and as a component of Tamiflu? capsule exhibited the identical cyclic voltammogram. The peaks originated from excipients in capsule do not appear und...
Saved in:
Published in: | Hemijska industrija 2011, Vol.65 (1), p.87-91 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A gold electrode was applied in the voltammetric determination of oseltamivir
phosphate standard in 0.05 M NaHCO3. Oseltamivir phosphate as a standard and
as a component of Tamiflu? capsule exhibited the identical cyclic
voltammogram. The peaks originated from excipients in capsule do not appear
under the applied electrochemical conditions. The electrochemical method for
the qualitative determination of oseltamivir phosphate in Tamiflu? capsule by
cyclic voltammetry was developed. The presence of oseltamivir phosphate as
standard and as a content of Tamiflu? capsule in electrolyte as well as their
concentrations were simultaneously checked by HPLC. The lack of the
current/concentration dependency was established. The not pretreated glassy
carbon electrode cannot be used for the determination of oseltamivir
phosphate under identical experimental conditions presented for gold
electrode.