Loading…
Indoor calibration method for UV index meters with a solar simulator and a reference spectroradiometer
We present a calibration method of UV index meters under a solar simulator based on a spectroradiometer as reference. The reference spectroradiometer is a double-grating-based type that can measure the erythemal effective irradiance and the global UV index in a wavelength range from 250 nm to 450 nm...
Saved in:
Published in: | International journal of metrology and quality engineering 2016, Vol.7 (1), p.101 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We present a calibration method of UV index meters under a solar simulator based on a spectroradiometer as reference. The reference spectroradiometer is a double-grating-based type that can measure the erythemal effective irradiance and the global UV index in a wavelength range from 250 nm to 450 nm. After calibration of its wavelength scale and spectral irradiance scale by using the KRISS standard lamp, we applied the spectroradiometer to measure the reference value for calibrating broadband UV index meters under a solar simulator. The uncertainty of the spectroradiometer for measuring the reference value for the global UV index is evaluated to be 5% as a relative expanded uncertainty (k = 2). |
---|---|
ISSN: | 2107-6839 2107-6847 |
DOI: | 10.1051/ijmqe/2015031 |