Loading…

Reliability of Hand-Held Transcranial Doppler with M-mode Ultrasound in Middle Cerebral Artery Measurement

To determine the intra- and interrater agreement of mean flow velocity (MFV) and pulsatility index (PI) measurement in middle cerebral arteries, assessed by transcranial Doppler (TCD) with M-mode. Masked experienced neurosonologists performed TCD with M-mode using handheld probe in healthy adult vol...

Full description

Saved in:
Bibliographic Details
Published in:Journal of medical ultrasound 2017-06, Vol.25 (2), p.76-81
Main Authors: Muñoz Venturelli, Paula, Brunser, Alejandro M., Gaete, Javier, Illanes, Sergio, López, Javiera, Olavarría, Verónica V., Reccius, Andrés, Brinck, Pablo, González, Francisca, Cavada, Gabriel, Lavados, Pablo M.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:To determine the intra- and interrater agreement of mean flow velocity (MFV) and pulsatility index (PI) measurement in middle cerebral arteries, assessed by transcranial Doppler (TCD) with M-mode. Masked experienced neurosonologists performed TCD with M-mode using handheld probe in healthy adult volunteers. The Bland–Altman method for concordance and intraclass correlation coefficient were used. Seventy-seven healthy volunteers and seven raters participated (3 on regular TCD shift and 4 off-shift). The intrarater absolute mean difference between measurements was 5.5 cm/s [95% confidence interval (CI), 4.7–6.3] for MVF and 0.073 (95% CI, 0.063–0.083) for PI. The difference between MFV measurements was significantly higher in off-shift raters (p=0.015). The interrater absolute mean difference between measurements was 6.5 cm/s (95% CI, 5.5–7.5) for MVF and 0.065 (95% CI, 0.059–0.071) for PI. No influence was found for the middle cerebral artery side, volunteer's sex, or age, and there was no significant difference between raters. The intraclass correlation coefficient was 82.2% (95% CI 77.8–85.6) and 72.9% (95% CI 67.4–77.6) for MFV and PI, respectively. There exists good intra- and interrater agreement in MFV and PI measurements using M-mode TCD. These results support the use of this noninvasive tool and are important for clinical and investigational purposes.
ISSN:0929-6441
2212-1552
DOI:10.1016/j.jmu.2016.12.001