Rapid Campimetry in glaucoma – correspondence with standard perimetry and OCT
The Rapid Campimetry (RC), a kinetic visual field test proved to reliably detect visual field defects within the central 10° degrees, the most crucial part for visual acuity and quality of life, affected even at very early stages of glaucoma, within a short measurement epoch, ~ 1 min. This study aim...
Saved in:
Published in: | Scientific reports 2024-10, Vol.14 (1), p.25400-14, Article 25400 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The Rapid Campimetry (RC), a kinetic visual field test proved to reliably detect visual field defects within the central 10° degrees, the most crucial part for visual acuity and quality of life, affected even at very early stages of glaucoma, within a short measurement epoch, ~ 1 min. This study aims to further investigate RC correspondence with standard tests in glaucoma, namely standard automated perimetry (SAP) and optical coherence tomography (OCT) within 10° of visual field (VF). For this purpose, we included 41 participants, [21 glaucoma (GLA, mean age: 65.9 ± 12.4; 12 preperimetric eyes and 11 with VF defects) and 20 healthy controls (HC; mean age: 65.0 ± 10.3); 20 eyes]. At first, we compared the rate of detection/exclusion of VF defects in RC vs. SAP. Then, for those with VF defects (11 eyes), we investigated the 68-pointwise correspondence of 10 − 2 layout of RC and SAP. For functional and structural (SF) correspondence, pointwise correspondence of RC, SAP vs. OCT
macula
[ macular ganglion cell thickness (GCL)] was also performed. Further, we compared sector-based correspondence of RC, SAP vs. OCT
macula
accounting for GCL displacement at the fovea as well as sector-based correspondence with OCT
disc
[peripapillary retinal nerve fiber layer thickness (pRNFL)]. Agreement estimates were reported along with Cohen Kappa (
) statistic. For overall performance, RC and SAP showed 100% agreement (
) for the exclusion of VF defects (HC and preperimetric GLA) and for detection of VF defects [11 eyes of 9 GLA, (
]. Further, RC outperformed SAP in detection of arcuate scotomas, 7 vs. 5, respectively. Pointwise correspondence of VF defects (11 eyes), RC-SAP agreement reached 90% accuracy (
). For SF correspondence, RC [SAP] showed 62% [69%] pointwise agreement with OCT
macula
, (
). For macular sector-based correspondence, SF correspondence improved and reached 83% [83%] agreement, (
). For OCT
disc
sector-based analysis, SF correspondence was highest, 100% [100%] agreement, (
). Rapid Campimetry gave reproducible results in comparison to SAP and OCT with high potential as a screening VF method given its short testing duration, ~ 1 min in screening mode, and compatibility with telemedicine technologies upon future optimization and automation. |
---|---|