Assessment of Image Quality and Lesion Detectability With Digital PET/CT System

The aim of this study was to assess image quality and lesion detectability acquired with a digital Positron Emission Tomography/Computed Tomography (PET/CT) Siemens Biograph Vision 600 system. Consecutive patients who underwent a FDG PET/CT during the first week of use of a digital PET/CT (Siemens B...

Full description

Saved in:
Bibliographic Details
Published in:Frontiers in medicine 2021-02, Vol.8, p.629096-629096
Main Authors: Delcroix, Olivier, Bourhis, David, Keromnes, Nathalie, Robin, Philippe, Le Roux, Pierre-Yves, Abgral, Ronan, Salaun, Pierre-Yves, Querellou, Solène
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c462t-960844a5bace9b66c2a6a73906f26cd65d84326a707c6d664962048dc07c3d293
cites cdi_FETCH-LOGICAL-c462t-960844a5bace9b66c2a6a73906f26cd65d84326a707c6d664962048dc07c3d293
container_end_page 629096
container_issue
container_start_page 629096
container_title Frontiers in medicine
container_volume 8
creator Delcroix, Olivier
Bourhis, David
Keromnes, Nathalie
Robin, Philippe
Le Roux, Pierre-Yves
Abgral, Ronan
Salaun, Pierre-Yves
Querellou, Solène
description The aim of this study was to assess image quality and lesion detectability acquired with a digital Positron Emission Tomography/Computed Tomography (PET/CT) Siemens Biograph Vision 600 system. Consecutive patients who underwent a FDG PET/CT during the first week of use of a digital PET/CT (Siemens Biograph Vision 600) at the nuclear medicine department of the university hospital of Brest were analyzed. PET were realized using list mode acquisition. For all patients, 4 datasets were reconstructed. We determined, according to phantom measurements, an equivalent time acquisition/reconstruction parameters pair of the digital PET/CT corresponding to an analog PET/CT image quality ("analog-like") as reference dataset. We compared the reference dataset with 3 others digital PET/CT reconstruction parameters, allowing a decrease of emission duration: 60, 90, and 120 s per bed position. Three nuclear medicine physicians evaluated independently, for each dataset, overall image quality [Maximal Intensity Projection (MIP), noise, sharpness] using a 4-point scale. Physicians assessed also lesion detection capability by reporting new visible lesions on each digital datasets with their confidence level in comparison with analog-like dataset. Ninety-eight patients were analyzed. Image quality of MIP (IQ ), sharpness (IQ ), and noise (IQ ) of all digital datasets (60, 90, and 120 s) were better than those evaluated with analog-like reconstruction. Moreover, digital PET/CT system improved IQ , IQ , and IQ whatever the BMI. Lesion detection capability and confidence level were higher for 60, 90, 120 s per bed position, respectively, than for analog-like images. Our study demonstrated an improvement of image quality and lesion detectability with a digital PET/CT system.
doi_str_mv 10.3389/fmed.2021.629096
format article
fullrecord <record><control><sourceid>proquest_doaj_</sourceid><recordid>TN_cdi_doaj_primary_oai_doaj_org_article_9eecbc6457494c138090777326711759</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><doaj_id>oai_doaj_org_article_9eecbc6457494c138090777326711759</doaj_id><sourcerecordid>2500386151</sourcerecordid><originalsourceid>FETCH-LOGICAL-c462t-960844a5bace9b66c2a6a73906f26cd65d84326a707c6d664962048dc07c3d293</originalsourceid><addsrcrecordid>eNpVkUtPGzEQgK2KChBw51T5yCVh_Njx-oKEAoVIkaBqqvZmeb3eYLS7hrVTKf-eDQEEJ9vz-Maaj5BTBlMhSn3edL6ecuBsilyDxm_kkHONk7Io_-19uh-Qk5QeAYAJXkgm9smBEKgFMDwkd5cp-ZQ632caGzrv7MrTX2vbhryhtq_pwqcQe3rls3fZVuE18TfkB3oVViHblt5fL89nS_p7k7Lvjsn3xrbJn7ydR-TPz-vl7HayuLuZzy4XEyeR54lGKKW0RWWd1xWi4xatEhqw4ehqLOpSCj6GQDmsEaVGDrKs3fgWNdfiiMx33DraR_M0hM4OGxNtMK-BOKyMHXJwrTfae1c5lIWSWjomStCglBrxijFVbFkXO9bTuhp36sZdDLb9Av2a6cODWcX_RmmhFIMRcPYGGOLz2qdsupCcb1vb-7hOhhcAokRWsLEUdqVuiCkNvvkYw8BstZqtVrPVanZax5Yfn7_30fAuUbwA7tmcoA</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2500386151</pqid></control><display><type>article</type><title>Assessment of Image Quality and Lesion Detectability With Digital PET/CT System</title><source>PMC (PubMed Central)</source><creator>Delcroix, Olivier ; Bourhis, David ; Keromnes, Nathalie ; Robin, Philippe ; Le Roux, Pierre-Yves ; Abgral, Ronan ; Salaun, Pierre-Yves ; Querellou, Solène</creator><creatorcontrib>Delcroix, Olivier ; Bourhis, David ; Keromnes, Nathalie ; Robin, Philippe ; Le Roux, Pierre-Yves ; Abgral, Ronan ; Salaun, Pierre-Yves ; Querellou, Solène</creatorcontrib><description>The aim of this study was to assess image quality and lesion detectability acquired with a digital Positron Emission Tomography/Computed Tomography (PET/CT) Siemens Biograph Vision 600 system. Consecutive patients who underwent a FDG PET/CT during the first week of use of a digital PET/CT (Siemens Biograph Vision 600) at the nuclear medicine department of the university hospital of Brest were analyzed. PET were realized using list mode acquisition. For all patients, 4 datasets were reconstructed. We determined, according to phantom measurements, an equivalent time acquisition/reconstruction parameters pair of the digital PET/CT corresponding to an analog PET/CT image quality ("analog-like") as reference dataset. We compared the reference dataset with 3 others digital PET/CT reconstruction parameters, allowing a decrease of emission duration: 60, 90, and 120 s per bed position. Three nuclear medicine physicians evaluated independently, for each dataset, overall image quality [Maximal Intensity Projection (MIP), noise, sharpness] using a 4-point scale. Physicians assessed also lesion detection capability by reporting new visible lesions on each digital datasets with their confidence level in comparison with analog-like dataset. Ninety-eight patients were analyzed. Image quality of MIP (IQ ), sharpness (IQ ), and noise (IQ ) of all digital datasets (60, 90, and 120 s) were better than those evaluated with analog-like reconstruction. Moreover, digital PET/CT system improved IQ , IQ , and IQ whatever the BMI. Lesion detection capability and confidence level were higher for 60, 90, 120 s per bed position, respectively, than for analog-like images. Our study demonstrated an improvement of image quality and lesion detectability with a digital PET/CT system.</description><identifier>ISSN: 2296-858X</identifier><identifier>EISSN: 2296-858X</identifier><identifier>DOI: 10.3389/fmed.2021.629096</identifier><identifier>PMID: 33693016</identifier><language>eng</language><publisher>Switzerland: Frontiers Media S.A</publisher><subject>analog detectors ; clinical evaluation ; image quality ; Medicine ; PET/CT ; SiPM</subject><ispartof>Frontiers in medicine, 2021-02, Vol.8, p.629096-629096</ispartof><rights>Copyright © 2021 Delcroix, Bourhis, Keromnes, Robin, Le Roux, Abgral, Salaun and Querellou.</rights><rights>Copyright © 2021 Delcroix, Bourhis, Keromnes, Robin, Le Roux, Abgral, Salaun and Querellou. 2021 Delcroix, Bourhis, Keromnes, Robin, Le Roux, Abgral, Salaun and Querellou</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c462t-960844a5bace9b66c2a6a73906f26cd65d84326a707c6d664962048dc07c3d293</citedby><cites>FETCH-LOGICAL-c462t-960844a5bace9b66c2a6a73906f26cd65d84326a707c6d664962048dc07c3d293</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC7937710/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC7937710/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,723,776,780,881,27898,27899,53763,53765</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/33693016$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Delcroix, Olivier</creatorcontrib><creatorcontrib>Bourhis, David</creatorcontrib><creatorcontrib>Keromnes, Nathalie</creatorcontrib><creatorcontrib>Robin, Philippe</creatorcontrib><creatorcontrib>Le Roux, Pierre-Yves</creatorcontrib><creatorcontrib>Abgral, Ronan</creatorcontrib><creatorcontrib>Salaun, Pierre-Yves</creatorcontrib><creatorcontrib>Querellou, Solène</creatorcontrib><title>Assessment of Image Quality and Lesion Detectability With Digital PET/CT System</title><title>Frontiers in medicine</title><addtitle>Front Med (Lausanne)</addtitle><description>The aim of this study was to assess image quality and lesion detectability acquired with a digital Positron Emission Tomography/Computed Tomography (PET/CT) Siemens Biograph Vision 600 system. Consecutive patients who underwent a FDG PET/CT during the first week of use of a digital PET/CT (Siemens Biograph Vision 600) at the nuclear medicine department of the university hospital of Brest were analyzed. PET were realized using list mode acquisition. For all patients, 4 datasets were reconstructed. We determined, according to phantom measurements, an equivalent time acquisition/reconstruction parameters pair of the digital PET/CT corresponding to an analog PET/CT image quality ("analog-like") as reference dataset. We compared the reference dataset with 3 others digital PET/CT reconstruction parameters, allowing a decrease of emission duration: 60, 90, and 120 s per bed position. Three nuclear medicine physicians evaluated independently, for each dataset, overall image quality [Maximal Intensity Projection (MIP), noise, sharpness] using a 4-point scale. Physicians assessed also lesion detection capability by reporting new visible lesions on each digital datasets with their confidence level in comparison with analog-like dataset. Ninety-eight patients were analyzed. Image quality of MIP (IQ ), sharpness (IQ ), and noise (IQ ) of all digital datasets (60, 90, and 120 s) were better than those evaluated with analog-like reconstruction. Moreover, digital PET/CT system improved IQ , IQ , and IQ whatever the BMI. Lesion detection capability and confidence level were higher for 60, 90, 120 s per bed position, respectively, than for analog-like images. Our study demonstrated an improvement of image quality and lesion detectability with a digital PET/CT system.</description><subject>analog detectors</subject><subject>clinical evaluation</subject><subject>image quality</subject><subject>Medicine</subject><subject>PET/CT</subject><subject>SiPM</subject><issn>2296-858X</issn><issn>2296-858X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNpVkUtPGzEQgK2KChBw51T5yCVh_Njx-oKEAoVIkaBqqvZmeb3eYLS7hrVTKf-eDQEEJ9vz-Maaj5BTBlMhSn3edL6ecuBsilyDxm_kkHONk7Io_-19uh-Qk5QeAYAJXkgm9smBEKgFMDwkd5cp-ZQ632caGzrv7MrTX2vbhryhtq_pwqcQe3rls3fZVuE18TfkB3oVViHblt5fL89nS_p7k7Lvjsn3xrbJn7ydR-TPz-vl7HayuLuZzy4XEyeR54lGKKW0RWWd1xWi4xatEhqw4ehqLOpSCj6GQDmsEaVGDrKs3fgWNdfiiMx33DraR_M0hM4OGxNtMK-BOKyMHXJwrTfae1c5lIWSWjomStCglBrxijFVbFkXO9bTuhp36sZdDLb9Av2a6cODWcX_RmmhFIMRcPYGGOLz2qdsupCcb1vb-7hOhhcAokRWsLEUdqVuiCkNvvkYw8BstZqtVrPVanZax5Yfn7_30fAuUbwA7tmcoA</recordid><startdate>20210222</startdate><enddate>20210222</enddate><creator>Delcroix, Olivier</creator><creator>Bourhis, David</creator><creator>Keromnes, Nathalie</creator><creator>Robin, Philippe</creator><creator>Le Roux, Pierre-Yves</creator><creator>Abgral, Ronan</creator><creator>Salaun, Pierre-Yves</creator><creator>Querellou, Solène</creator><general>Frontiers Media S.A</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><scope>DOA</scope></search><sort><creationdate>20210222</creationdate><title>Assessment of Image Quality and Lesion Detectability With Digital PET/CT System</title><author>Delcroix, Olivier ; Bourhis, David ; Keromnes, Nathalie ; Robin, Philippe ; Le Roux, Pierre-Yves ; Abgral, Ronan ; Salaun, Pierre-Yves ; Querellou, Solène</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c462t-960844a5bace9b66c2a6a73906f26cd65d84326a707c6d664962048dc07c3d293</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>analog detectors</topic><topic>clinical evaluation</topic><topic>image quality</topic><topic>Medicine</topic><topic>PET/CT</topic><topic>SiPM</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Delcroix, Olivier</creatorcontrib><creatorcontrib>Bourhis, David</creatorcontrib><creatorcontrib>Keromnes, Nathalie</creatorcontrib><creatorcontrib>Robin, Philippe</creatorcontrib><creatorcontrib>Le Roux, Pierre-Yves</creatorcontrib><creatorcontrib>Abgral, Ronan</creatorcontrib><creatorcontrib>Salaun, Pierre-Yves</creatorcontrib><creatorcontrib>Querellou, Solène</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><collection>Directory of Open Access Journals (DOAJ)</collection><jtitle>Frontiers in medicine</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Delcroix, Olivier</au><au>Bourhis, David</au><au>Keromnes, Nathalie</au><au>Robin, Philippe</au><au>Le Roux, Pierre-Yves</au><au>Abgral, Ronan</au><au>Salaun, Pierre-Yves</au><au>Querellou, Solène</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Assessment of Image Quality and Lesion Detectability With Digital PET/CT System</atitle><jtitle>Frontiers in medicine</jtitle><addtitle>Front Med (Lausanne)</addtitle><date>2021-02-22</date><risdate>2021</risdate><volume>8</volume><spage>629096</spage><epage>629096</epage><pages>629096-629096</pages><issn>2296-858X</issn><eissn>2296-858X</eissn><abstract>The aim of this study was to assess image quality and lesion detectability acquired with a digital Positron Emission Tomography/Computed Tomography (PET/CT) Siemens Biograph Vision 600 system. Consecutive patients who underwent a FDG PET/CT during the first week of use of a digital PET/CT (Siemens Biograph Vision 600) at the nuclear medicine department of the university hospital of Brest were analyzed. PET were realized using list mode acquisition. For all patients, 4 datasets were reconstructed. We determined, according to phantom measurements, an equivalent time acquisition/reconstruction parameters pair of the digital PET/CT corresponding to an analog PET/CT image quality ("analog-like") as reference dataset. We compared the reference dataset with 3 others digital PET/CT reconstruction parameters, allowing a decrease of emission duration: 60, 90, and 120 s per bed position. Three nuclear medicine physicians evaluated independently, for each dataset, overall image quality [Maximal Intensity Projection (MIP), noise, sharpness] using a 4-point scale. Physicians assessed also lesion detection capability by reporting new visible lesions on each digital datasets with their confidence level in comparison with analog-like dataset. Ninety-eight patients were analyzed. Image quality of MIP (IQ ), sharpness (IQ ), and noise (IQ ) of all digital datasets (60, 90, and 120 s) were better than those evaluated with analog-like reconstruction. Moreover, digital PET/CT system improved IQ , IQ , and IQ whatever the BMI. Lesion detection capability and confidence level were higher for 60, 90, 120 s per bed position, respectively, than for analog-like images. Our study demonstrated an improvement of image quality and lesion detectability with a digital PET/CT system.</abstract><cop>Switzerland</cop><pub>Frontiers Media S.A</pub><pmid>33693016</pmid><doi>10.3389/fmed.2021.629096</doi><tpages>1</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2296-858X
ispartof Frontiers in medicine, 2021-02, Vol.8, p.629096-629096
issn 2296-858X
2296-858X
language eng
recordid cdi_doaj_primary_oai_doaj_org_article_9eecbc6457494c138090777326711759
source PMC (PubMed Central)
subjects analog detectors
clinical evaluation
image quality
Medicine
PET/CT
SiPM
title Assessment of Image Quality and Lesion Detectability With Digital PET/CT System
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-25T16%3A22%3A33IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_doaj_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Assessment%20of%20Image%20Quality%20and%20Lesion%20Detectability%20With%20Digital%20PET/CT%20System&rft.jtitle=Frontiers%20in%20medicine&rft.au=Delcroix,%20Olivier&rft.date=2021-02-22&rft.volume=8&rft.spage=629096&rft.epage=629096&rft.pages=629096-629096&rft.issn=2296-858X&rft.eissn=2296-858X&rft_id=info:doi/10.3389/fmed.2021.629096&rft_dat=%3Cproquest_doaj_%3E2500386151%3C/proquest_doaj_%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c462t-960844a5bace9b66c2a6a73906f26cd65d84326a707c6d664962048dc07c3d293%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2500386151&rft_id=info:pmid/33693016&rfr_iscdi=true