Loading…
Hard X‐ray imaging and tomography at the Biomedical Imaging and Therapy beamlines of Canadian Light Source
The Biomedical Imaging and Therapy facility of the Canadian Light Source comprises two beamlines, which together cover a wide X‐ray energy range from 13 keV up to 140 keV. The beamlines were designed with a focus on synchrotron applications in preclinical imaging and veterinary science as well as mi...
Saved in:
Published in: | Journal of synchrotron radiation 2024-09, Vol.31 (5), p.1346-1357 |
---|---|
Main Authors: | , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The Biomedical Imaging and Therapy facility of the Canadian Light Source comprises two beamlines, which together cover a wide X‐ray energy range from 13 keV up to 140 keV. The beamlines were designed with a focus on synchrotron applications in preclinical imaging and veterinary science as well as microbeam radiation therapy. While these remain a major part of the activities of both beamlines, a number of recent upgrades have enhanced the versatility and performance of the beamlines, particularly for high‐resolution microtomography experiments. As a result, the user community has been quickly expanding to include researchers in advanced materials, batteries, fuel cells, agriculture, and environmental studies. This article summarizes the beam properties, describes the endstations together with the detector pool, and presents several application cases of the various X‐ray imaging techniques available to users.
The current state of the Biomedical Imaging and Therapy beamlines of the Canadian Light Source is described and new capabilities are presented. |
---|---|
ISSN: | 1600-5775 0909-0495 1600-5775 |
DOI: | 10.1107/S1600577524005241 |