Loading…

Intraocular Pressure Measurement Using Ocular Response Analyzer, Dynamic Contour Tonometer, and Scheimpflug Analyzer Corvis ST

Purpose. To compare intraocular pressure (IOP) measurements with Goldmann applanation tonometry (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST (CST) in healthy subjects. Methods. In a prospective, observational study, IOP measurements with GAT (GAT-IOPc), ORA (...

Full description

Saved in:
Bibliographic Details
Published in:Journal of ophthalmology 2019, Vol.2019 (2019), p.1-9
Main Authors: Pillunat, Lutz E., Raiskup, Frederik, Spoerl, Eberhard, Herber, Robert, Ramm, Lisa, Terai, Naim
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Purpose. To compare intraocular pressure (IOP) measurements with Goldmann applanation tonometry (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST (CST) in healthy subjects. Methods. In a prospective, observational study, IOP measurements with GAT (GAT-IOPc), ORA (IOPcc), DCT (DCT-IOP), and CST (bIOP) were performed and analyzed in 94 healthy subjects. Results. Mean age of the participants was 45.6 ± 17.2 years (range 18 to 81 years). Mean GAT-IOPc was 12.9 ± 2.4 mmHg, mean DCT-IOP was 16.1 ± 2.6 mmHg, and mean IOPcc was 15.6 ± 3.3 mmHg. DCT-IOP and IOPcc were significantly higher than GAT-IOPc (P
ISSN:2090-004X
2090-0058