Evaluation of manual resuscitators used in ICUs in Brazil
To evaluate the performance of manual resuscitators (MRs) used in Brazil in accordance with international standards. Using a respiratory system simulator, four volunteer physiotherapists employed eight MRs (five produced in Brazil and three produced abroad), which were tested for inspiratory and exp...
Saved in:
Published in: | Jornal brasileiro de pneumologia 2013-09, Vol.39 (5), p.595-603 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | eng ; por |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To evaluate the performance of manual resuscitators (MRs) used in Brazil in accordance with international standards.
Using a respiratory system simulator, four volunteer physiotherapists employed eight MRs (five produced in Brazil and three produced abroad), which were tested for inspiratory and expiratory resistance of the patient valve; functioning of the pressure-limiting valve; and tidal volume (VT) generated when the one-handed and two-handed techniques were used. The tests were performed and analyzed in accordance with the American Society for Testing and Materials (ASTM) F920-93 criteria.
Expiratory resistance was greater than 6 cmH2O . L-1 . s-1 in only one MR. The pressure-limiting valve, a feature of five of the MRs, opened at low pressures (< 17 cmH2O), and the maximal pressure was 32.0-55.9 cmH2O. Mean VT varied greatly among the MRs tested. The mean VT values generated with the one-handed technique were lower than the 600 mL recommended by the ASTM. In the situations studied, mean VT was generally lower from the Brazilian-made MRs that had a pressure-limiting valve.
The resistances imposed by the patient valve met the ASTM criteria in all but one of the MRs tested. The pressure-limiting valves of the Brazilian-made MRs usually opened at low pressures, providing lower VT values in the situations studied, especially when the one-handed technique was used, suggesting that both hands should be used and that the pressure-limiting valve should be closed whenever possible. |
---|---|