Loading…

Billowing Of Endologix Powerlink Stent Mimicking Endoleaks

Endoleaks remains one of the primary concerns of endovascular aortic aneurysm repair (EVAR) and is routinely followed with CT angiography (CTA). However, certain imaging findings can mimic endoleaks. A 65-year-old woman who had endovascular aortic repair (EVAR) of an abdominal aortic aneurysm with E...

Full description

Saved in:
Bibliographic Details
Published in:Research in cardiovascular medicine 2016-04, Vol.5 (2), p.e32448-1
Main Authors: Wu, Alex, Karuppasamy, Karunakaravel, Wang, Weiping
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Endoleaks remains one of the primary concerns of endovascular aortic aneurysm repair (EVAR) and is routinely followed with CT angiography (CTA). However, certain imaging findings can mimic endoleaks. A 65-year-old woman who had endovascular aortic repair (EVAR) of an abdominal aortic aneurysm with Endologix Powerlink system developed marked new circumferential cauliflower-like bulging of contrast-filled sacs at mid-stent-graft with enlargement of the excluded aneurysm at 3-year follow-up. Considering the unique construct of the Powerlink stents, this is thought to represent aneurysmal degeneration of the outer fabric material from the metal struts and may potentially pressurize the excluded sac with risk for rupture.
ISSN:2251-9572
2251-9580
DOI:10.5812/cardiovascmed.32448