Loading…

Developing a new Monte Carlo algorithm as an alternative tool to simulate virtual source model on an Elekta Versa HD Linac

The aim of this study is to develop in-house Monte Carlo (MC) algorithm as an alternative tool to simulate all mechanisms of interactions for dose calculation in radiation therapy. Additionally, evaluating the MC-SCAN algorithm with beam modeling of Elekta VersaHD Linac was aimed. Mohan 6 MV photon...

Full description

Saved in:
Bibliographic Details
Published in:Journal of radiation research and applied sciences 2022-09, Vol.15 (3), p.299-308
Main Authors: Can, Sümeyra, Şahi̇ner, Eren, Karaçetin, Didem, Meriç, Niyazi
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The aim of this study is to develop in-house Monte Carlo (MC) algorithm as an alternative tool to simulate all mechanisms of interactions for dose calculation in radiation therapy. Additionally, evaluating the MC-SCAN algorithm with beam modeling of Elekta VersaHD Linac was aimed. Mohan 6 MV photon spectrum was used and the source was modeled as a point source. The absorbed dose is calculated via modeling all interaction mechanisms to gain energy deposition. TPR20/10, PDD, photon profile of different field sizes as well as point dose was calculated after that, a comparison of all the achieved results with experimental measurements went through. The difference between calculated mean energy and theoretical one was obtained
ISSN:1687-8507
1687-8507
DOI:10.1016/j.jrras.2022.07.009