Loading…
Developing a new Monte Carlo algorithm as an alternative tool to simulate virtual source model on an Elekta Versa HD Linac
The aim of this study is to develop in-house Monte Carlo (MC) algorithm as an alternative tool to simulate all mechanisms of interactions for dose calculation in radiation therapy. Additionally, evaluating the MC-SCAN algorithm with beam modeling of Elekta VersaHD Linac was aimed. Mohan 6 MV photon...
Saved in:
Published in: | Journal of radiation research and applied sciences 2022-09, Vol.15 (3), p.299-308 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The aim of this study is to develop in-house Monte Carlo (MC) algorithm as an alternative tool to simulate all mechanisms of interactions for dose calculation in radiation therapy. Additionally, evaluating the MC-SCAN algorithm with beam modeling of Elekta VersaHD Linac was aimed. Mohan 6 MV photon spectrum was used and the source was modeled as a point source. The absorbed dose is calculated via modeling all interaction mechanisms to gain energy deposition. TPR20/10, PDD, photon profile of different field sizes as well as point dose was calculated after that, a comparison of all the achieved results with experimental measurements went through. The difference between calculated mean energy and theoretical one was obtained |
---|---|
ISSN: | 1687-8507 1687-8507 |