Loading…
Preliminary results of a quick, simple method of detecting antimony in water samples
Preliminary results of development of a direct and fast method of determination of antimony in samples of tap water using GFAAS are presented. The found levels of antimony were lower than permitted for human consumption. A mixture of Pd and Mg(NO 3 ) 2 (concentrations in the injected solution: 8.6 μ...
Saved in:
Published in: | Central European Journal of Chemistry 2008-12, Vol.6 (4), p.520-525 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Preliminary results of development of a direct and fast method of determination of antimony in samples of tap water using GFAAS are presented. The found levels of antimony were lower than permitted for human consumption. A mixture of Pd and Mg(NO
3
)
2
(concentrations in the injected solution: 8.6 μg mL
−1
and 5.8 μg mL
−1
respectively) was used as the chemical modifier. The pyrolysis and atomization temperatures were 1000 and 1700°C, respectively and the mean analytical recovery 98.2%. |
---|---|
ISSN: | 1895-1066 2391-5420 1644-3624 2391-5420 |
DOI: | 10.2478/s11532-008-0071-y |