Loading…

Electrical field measurements and simulations of the H7 and D-B80 coils: Non-equivalence of the TMS coils for obsessive compulsive disorder

•Recently FDA approved coils for OCD treatment show substantial differences in their distributions of electric fields.•The H7 coil stimulates significantly broader and deeper brain volume compared to the D-B80.•Specific regions known to be implicated with OCD are stimulated by the H7, but not by the...

Full description

Saved in:
Bibliographic Details
Published in:Brain stimulation 2021-11, Vol.14 (6), p.1525-1527
Main Authors: Tzirini, Marietta, Roth, Yiftach, Harmelech, Tal, Zibman, Samuel, Pell, Gaby S., Kimiskidis, Vasilios, Tendler, Aron, Zangen, Abraham, Samaras, Theodoros
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:•Recently FDA approved coils for OCD treatment show substantial differences in their distributions of electric fields.•The H7 coil stimulates significantly broader and deeper brain volume compared to the D-B80.•Specific regions known to be implicated with OCD are stimulated by the H7, but not by the D-B80 coil.