Dosimetric comparison between microSelectron iridium-192 and flexi cobalt-60 sources in high-dose-rate brachytherapy using Geant4 Monte Carlo code

Manufacturing of miniaturized high activity iridium-192 ( Ir) sources have been made a market preference in modern brachytherapy. Smaller dimensions of the sources are flexible for smaller diameter of the applicators, and it is also suitable for interstitial implants. Presently, cobalt-60 ( Co) sour...

Full description

Saved in:
Bibliographic Details
Published in:Journal of contemporary brachytherapy 2023-01, Vol.15 (2), p.141-147
Main Authors: Gebremariam, Tsige Yhidego, Geraily, Ghazale, Jassim, Hussam Hameed, Gholami, Somayeh
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Manufacturing of miniaturized high activity iridium-192 ( Ir) sources have been made a market preference in modern brachytherapy. Smaller dimensions of the sources are flexible for smaller diameter of the applicators, and it is also suitable for interstitial implants. Presently, cobalt-60 ( Co) sources have been commercialized as an alternative to Ir sources for high-dose-rate (HDR) brachytherapy, since Co source have an advantage of longer half-life comparing with Ir source. One of them is the HDR Co Flexisource manufactured by Elekta. The purpose of this study was to compare the TG-43 dosimetric parameters of HDR flexi Co and HDR microSelectron Ir sources. Monte Carlo simulation code of Geant4 (v.11.0) was applied. Following the recommendations of AAPM TG-43 formalism report, Monte Carlo code of HDR flexi Co and HDR microSelectron Ir was validated by calculating radial dose function, anisotropy function, and dose-rate constants in a water phantom. Finally, results of both radionuclide sources were compared. The calculated dose-rate constants per unit air-kerma strength in water medium were 1.108 cGy h U for HDR microSelectron Ir, and 1.097 cGy h U for HDR flexi Co source, with the percentage uncertainty of 1.1% and 0.2%, respectively. The values of radial dose function for distances above 22 cm for HDR flexi Co source were higher than that of the other source. The anisotropic values sharply increased to the longitudinal sides of HDR flexi Co source, and the rise was comparatively sharper to that of the other source. The primary photons from the lower-energy HDR microSelectron Ir source have a limited range and are partially attenuated when considering the results of radial and anisotropic dose distribution functions. This implies that a HDR flexi Co radionuclide could be used to treat tumors beyond the source compared with a HDR microSelectron Ir source, despite the fact that Ir has a lower exit dose than HDR flexi Co radionuclide source.
ISSN:1689-832X
2081-2841
DOI:10.5114/jcb.2023.126238