Comparison of Icare Pro Tonometry and Icare One Tonometry Measurements in Healthy Eyes

To compare intraocular pressure (IOP) measurements obtained with the Icare Pro tonometer used in clinical practice and the Icare One self-tonometer. Fifty-two eyes of 52 healthy, right-handed individuals with no prior intraocular surgery or ocular trauma, structural ocular pathology, or systemic dis...

Full description

Saved in:
Bibliographic Details
Published in:Turk oftalmoloji gazetesi 2019-06, Vol.49 (3), p.130-133
Main Authors: Mayalı, Hüseyin, Sarıgül, Çağlar, Kurt, Emin, Kayıkçıoğlu, Özcan Rasim, İlker, Süleyman Sami
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:To compare intraocular pressure (IOP) measurements obtained with the Icare Pro tonometer used in clinical practice and the Icare One self-tonometer. Fifty-two eyes of 52 healthy, right-handed individuals with no prior intraocular surgery or ocular trauma, structural ocular pathology, or systemic disease were evaluated. IOP was first measured using the Icare Pro tonometer. The participants were then told how to use the Icare One tonometer and asked to measure their own IOP. The results were analyzed statistically using SPSS v.24. Of the 52 healthy participants, 16 (30.7%) were male and 36 (69.3%) were female. Their mean age was 31.6±6.3 (23-47) years. Mean IOP measured with the Icare Pro was 17.10±6.2 (11.5-25.2) mmHg, and the mean self-measured IOP with Icare One was 14.01±3.4 (7-24) mmHg. When the two methods were compared using Levene’s t-test, there was a significant mean difference of -3.08±0.6 (95% confidence interval: -4.39 -1.78; p
ISSN:1300-0659
2149-8695
2149-8709
2147-2661
DOI:10.4274/tjo.galenos.2018.06787