Comparison of Anterior Segment Measurements Obtained by Aladdin Optical Biometer and Sirius Corneal Topography
To assess the agreement of anterior segment parameter measurements derived from Aladdin optical biometer using optical low coherence interferometer and Sirius corneal topography using combined Scheimpflug-Placido disk. Data obtained using the Aladdin and Sirius systems from 110 eyes of 59 subjects w...
Saved in:
Published in: | Turk oftalmoloji gazetesi 2016-12, Vol.46 (6), p.259-263 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To assess the agreement of anterior segment parameter measurements derived from Aladdin optical biometer using optical low coherence interferometer and Sirius corneal topography using combined Scheimpflug-Placido disk.
Data obtained using the Aladdin and Sirius systems from 110 eyes of 59 subjects who had no health problems other than refractive errors were retrospectively evaluated. Anterior chamber depth (ACD), flat (K1) and steep (K2) keratometry readings, and white-to-white distance (WTW) measurements taken with both devices were noted.
The mean age of the patients was 47.31±18.57 years (range, 25 to 79 years). Mean ACD was 3.35±0.4 mm using Aladdin and 3.42±0.44 mm using Sirius. Mean difference in ACD was 0.075 mm greater with Sirius than Aladdin (p |
---|---|