Reversible Sorptive Preconcentration of Noble Metals Followed by FI-ICP-MS Determination

In this paper, we propose the combined procedure of noble metal (NM) determination, including fire assay, acid digestion, and reversible dynamic sorptive preconcentration, followed by flow-injection ICP-MS. Reversible preconcentration of all NMs was carried out using micro-column packed new PVBC-VP...

Full description

Saved in:
Bibliographic Details
Published in:Molecules (Basel, Switzerland) Switzerland), 2022-10, Vol.27 (19), p.6746
Main Authors: Maksimova, Yulia A, Dubenskiy, Alexander S, Pavlova, Lyudmila A, Shigapov, Ilya V, Korshunov, Dmitry M, Seregina, Irina F, Davankov, Vadim A, Bolshov, Mikhail A
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In this paper, we propose the combined procedure of noble metal (NM) determination, including fire assay, acid digestion, and reversible dynamic sorptive preconcentration, followed by flow-injection ICP-MS. Reversible preconcentration of all NMs was carried out using micro-column packed new PVBC-VP sorbent and elution with a mixture of thiourea, potassium thiocyanate, and HCl, which recovers Pd, Ir, Pt, and Au by 95% and Ru, and Rh by 90%. The proposed procedure was approved using certified reference materials.