Geometric accuracy in patient positioning for stereotactic radiotherapy of intracranial tumors
This study determines and compares the geometric setup errors between stereoscopic x-ray and kilo-voltage cone beam CT (CBCT) in phantom tests on a linear accelerator (linac) for image-guided (IG) stereotactic radiotherapy of intracranial tumors. Additionally, dose-volume metrics in the target volum...
Saved in:
Published in: | Physics and imaging in radiation oncology 2023-07, Vol.27, p.100461-100461, Article 100461 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | This study determines and compares the geometric setup errors between stereoscopic x-ray and kilo-voltage cone beam CT (CBCT) in phantom tests on a linear accelerator (linac) for image-guided (IG) stereotactic radiotherapy of intracranial tumors. Additionally, dose-volume metrics in the target volumes of the setup errors of CBCT were evaluated.
A Winston-Lutz- and an anthropomorphic phantom were used. The mean deviation and root mean square error (RMSE) of CBCT and stereoscopic x-ray were compared. Dose-volume metrics of the planning target volume (PTV) and gross target volume (GTV) for CBCT were calculated.
The RMSEs in the tests with the Winston-Lutz-Phantom were 0.3 mm, 1.1 mm and 0.3 mm for CBCT and 0.1 mm, 0,1 mm and |
---|---|