Loading…

Ideal Illumination for Smartphone-based Trabeculectomy Bleb Photography

Ophthalmology has seen numerous novel uses for smartphones over the years including fundus photography, telemedicine, and operative videography. However, anterior segment photography for assessing and documenting trabeculectomy bleb morphology using a smartphone has not been explored in detail. With...

Full description

Saved in:
Bibliographic Details
Published in:Journal of ophthalmic & vision research 2021-07, Vol.16 (3), p.357-366
Main Authors: Kalra, Gagan, Ichhpujani, Parul, Thakur, Sahil, Sharma, Urvashi
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c420t-7fe7114c1236565e1447681151d807453abb639b02cc7b36fc1663577393c2003
cites cdi_FETCH-LOGICAL-c420t-7fe7114c1236565e1447681151d807453abb639b02cc7b36fc1663577393c2003
container_end_page 366
container_issue 3
container_start_page 357
container_title Journal of ophthalmic & vision research
container_volume 16
creator Kalra, Gagan
Ichhpujani, Parul
Thakur, Sahil
Sharma, Urvashi
description Ophthalmology has seen numerous novel uses for smartphones over the years including fundus photography, telemedicine, and operative videography. However, anterior segment photography for assessing and documenting trabeculectomy bleb morphology using a smartphone has not been explored in detail. With the current study, we aim to characterize ideal illumination for the anterior segment smartphone photography in trabeculectomy patients. Thirty status post-trabeculectomy patients were enrolled in this study. Native camera application and FiLMiC pro camera application were used on iPhone X to compare bleb images using yellow and white pen-torches as illumination source. Measured bleb area was compared using ImageJ software from the two apps in different illumination settings by charting boxplots and using one-way ANOVA test using R software to establish consistency. Bland-Altman interoperability for repeatability of bleb-area measurements was analyzed by plotting Bland-Altman plots. Signal-to-noise ratio was calculated using ImageJ for native camera images using slit-lamp camera images as reference. Subjective rating of these images was then performed by two experienced ophthalmologists and kappa coefficient was calculated for inter-operator repeatability. Statistical analysis was performed. The measured bleb area from images taken from both apps showed no significant difference, thereby establishing consistency, and Bland-Altman analysis indicated good repeatability and reproducibility. It was noted that SNR was lower for images shot in close illumination as compared to the ones shot in intermediate and distant illumination. Cohen's kappa coefficient was 0.7 for images with distant illumination using white light and 0.65 for images clicked with illumination at an intermediate distance using yellow light, suggesting substantial agreement between the observers. Smartphone photography is a reliable tool for morphological assessment trabeculectomy blebs. Optimal illumination helps achieve results free from digital noise and better delineation of specific morphological features. Intermediate illumination and distant illumination provides much better results in terms of high SNR while avoiding overexposure and clipping of highlight information in the images.
doi_str_mv 10.18502/jovr.v16i3.9432
format article
fullrecord <record><control><sourceid>proquest_doaj_</sourceid><recordid>TN_cdi_doaj_primary_oai_doaj_org_article_fabe23063d9f409ab3df8647c8d8b92e</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><doaj_id>oai_doaj_org_article_fabe23063d9f409ab3df8647c8d8b92e</doaj_id><sourcerecordid>2561917700</sourcerecordid><originalsourceid>FETCH-LOGICAL-c420t-7fe7114c1236565e1447681151d807453abb639b02cc7b36fc1663577393c2003</originalsourceid><addsrcrecordid>eNpVkU1r3DAQhkVpaL56zyn42Is3kkZfvhTa0CQLgQSSQG9CkuVdL7K1leyF_fd1djchOWmQZh49w4vQBcEzojimV6u4SbMNES3MKgb0CzqhGKsSKP379VBTTPAxOs15hbEQgqlv6BgYVEwJdoJu57U3oZiHMHZtb4Y29kUTU_HUmTSsl7H3pTXZ18VzMta7MXg3xG5b_A7eFo_LOMRFMuvl9hwdNSZk__1wnqGXmz_P13fl_cPt_PrXfekYxUMpGy8JYY5QEFxwTxiTQhHCSa2wZByMtQIqi6lz0oJoHBECuJRQgZvWgTM033PraFZ6ndrJc6ujafXuIqaFnsRbF7xuJmEKWEBdNQxXxkLdTDtLp2plK-on1s89az3aztfO90My4RP080vfLvUibrQCriSnE-DHAZDiv9HnQXdtdj4E0_s4Zk25IBWRcueN960uxZyTb96_IVjvstSvWepdlvo1y2nk8qPe-8BbePAf3wKb4A</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2561917700</pqid></control><display><type>article</type><title>Ideal Illumination for Smartphone-based Trabeculectomy Bleb Photography</title><source>PubMed Central</source><creator>Kalra, Gagan ; Ichhpujani, Parul ; Thakur, Sahil ; Sharma, Urvashi</creator><creatorcontrib>Kalra, Gagan ; Ichhpujani, Parul ; Thakur, Sahil ; Sharma, Urvashi</creatorcontrib><description>Ophthalmology has seen numerous novel uses for smartphones over the years including fundus photography, telemedicine, and operative videography. However, anterior segment photography for assessing and documenting trabeculectomy bleb morphology using a smartphone has not been explored in detail. With the current study, we aim to characterize ideal illumination for the anterior segment smartphone photography in trabeculectomy patients. Thirty status post-trabeculectomy patients were enrolled in this study. Native camera application and FiLMiC pro camera application were used on iPhone X to compare bleb images using yellow and white pen-torches as illumination source. Measured bleb area was compared using ImageJ software from the two apps in different illumination settings by charting boxplots and using one-way ANOVA test using R software to establish consistency. Bland-Altman interoperability for repeatability of bleb-area measurements was analyzed by plotting Bland-Altman plots. Signal-to-noise ratio was calculated using ImageJ for native camera images using slit-lamp camera images as reference. Subjective rating of these images was then performed by two experienced ophthalmologists and kappa coefficient was calculated for inter-operator repeatability. Statistical analysis was performed. The measured bleb area from images taken from both apps showed no significant difference, thereby establishing consistency, and Bland-Altman analysis indicated good repeatability and reproducibility. It was noted that SNR was lower for images shot in close illumination as compared to the ones shot in intermediate and distant illumination. Cohen's kappa coefficient was 0.7 for images with distant illumination using white light and 0.65 for images clicked with illumination at an intermediate distance using yellow light, suggesting substantial agreement between the observers. Smartphone photography is a reliable tool for morphological assessment trabeculectomy blebs. Optimal illumination helps achieve results free from digital noise and better delineation of specific morphological features. Intermediate illumination and distant illumination provides much better results in terms of high SNR while avoiding overexposure and clipping of highlight information in the images.</description><identifier>ISSN: 2008-2010</identifier><identifier>EISSN: 2008-322X</identifier><identifier>DOI: 10.18502/jovr.v16i3.9432</identifier><identifier>PMID: 34394864</identifier><language>eng</language><publisher>United Arab Emirates: PUBLISHED BY KNOWLEDGE E</publisher><subject>bleb ; Original ; smartphone photography ; teleophthalmology ; trabeculectomy</subject><ispartof>Journal of ophthalmic &amp; vision research, 2021-07, Vol.16 (3), p.357-366</ispartof><rights>Copyright © 2021 Kalra et al.</rights><rights>Copyright © 2021 Kalra et al. 2021</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c420t-7fe7114c1236565e1447681151d807453abb639b02cc7b36fc1663577393c2003</citedby><cites>FETCH-LOGICAL-c420t-7fe7114c1236565e1447681151d807453abb639b02cc7b36fc1663577393c2003</cites><orcidid>0000-0001-6256-4002 ; 0000-0002-3367-3047</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC8358752/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC8358752/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,885,27923,27924,53790,53792</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/34394864$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Kalra, Gagan</creatorcontrib><creatorcontrib>Ichhpujani, Parul</creatorcontrib><creatorcontrib>Thakur, Sahil</creatorcontrib><creatorcontrib>Sharma, Urvashi</creatorcontrib><title>Ideal Illumination for Smartphone-based Trabeculectomy Bleb Photography</title><title>Journal of ophthalmic &amp; vision research</title><addtitle>J Ophthalmic Vis Res</addtitle><description>Ophthalmology has seen numerous novel uses for smartphones over the years including fundus photography, telemedicine, and operative videography. However, anterior segment photography for assessing and documenting trabeculectomy bleb morphology using a smartphone has not been explored in detail. With the current study, we aim to characterize ideal illumination for the anterior segment smartphone photography in trabeculectomy patients. Thirty status post-trabeculectomy patients were enrolled in this study. Native camera application and FiLMiC pro camera application were used on iPhone X to compare bleb images using yellow and white pen-torches as illumination source. Measured bleb area was compared using ImageJ software from the two apps in different illumination settings by charting boxplots and using one-way ANOVA test using R software to establish consistency. Bland-Altman interoperability for repeatability of bleb-area measurements was analyzed by plotting Bland-Altman plots. Signal-to-noise ratio was calculated using ImageJ for native camera images using slit-lamp camera images as reference. Subjective rating of these images was then performed by two experienced ophthalmologists and kappa coefficient was calculated for inter-operator repeatability. Statistical analysis was performed. The measured bleb area from images taken from both apps showed no significant difference, thereby establishing consistency, and Bland-Altman analysis indicated good repeatability and reproducibility. It was noted that SNR was lower for images shot in close illumination as compared to the ones shot in intermediate and distant illumination. Cohen's kappa coefficient was 0.7 for images with distant illumination using white light and 0.65 for images clicked with illumination at an intermediate distance using yellow light, suggesting substantial agreement between the observers. Smartphone photography is a reliable tool for morphological assessment trabeculectomy blebs. Optimal illumination helps achieve results free from digital noise and better delineation of specific morphological features. Intermediate illumination and distant illumination provides much better results in terms of high SNR while avoiding overexposure and clipping of highlight information in the images.</description><subject>bleb</subject><subject>Original</subject><subject>smartphone photography</subject><subject>teleophthalmology</subject><subject>trabeculectomy</subject><issn>2008-2010</issn><issn>2008-322X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNpVkU1r3DAQhkVpaL56zyn42Is3kkZfvhTa0CQLgQSSQG9CkuVdL7K1leyF_fd1djchOWmQZh49w4vQBcEzojimV6u4SbMNES3MKgb0CzqhGKsSKP379VBTTPAxOs15hbEQgqlv6BgYVEwJdoJu57U3oZiHMHZtb4Y29kUTU_HUmTSsl7H3pTXZ18VzMta7MXg3xG5b_A7eFo_LOMRFMuvl9hwdNSZk__1wnqGXmz_P13fl_cPt_PrXfekYxUMpGy8JYY5QEFxwTxiTQhHCSa2wZByMtQIqi6lz0oJoHBECuJRQgZvWgTM033PraFZ6ndrJc6ujafXuIqaFnsRbF7xuJmEKWEBdNQxXxkLdTDtLp2plK-on1s89az3aztfO90My4RP080vfLvUibrQCriSnE-DHAZDiv9HnQXdtdj4E0_s4Zk25IBWRcueN960uxZyTb96_IVjvstSvWepdlvo1y2nk8qPe-8BbePAf3wKb4A</recordid><startdate>20210701</startdate><enddate>20210701</enddate><creator>Kalra, Gagan</creator><creator>Ichhpujani, Parul</creator><creator>Thakur, Sahil</creator><creator>Sharma, Urvashi</creator><general>PUBLISHED BY KNOWLEDGE E</general><general>Knowledge E</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><scope>DOA</scope><orcidid>https://orcid.org/0000-0001-6256-4002</orcidid><orcidid>https://orcid.org/0000-0002-3367-3047</orcidid></search><sort><creationdate>20210701</creationdate><title>Ideal Illumination for Smartphone-based Trabeculectomy Bleb Photography</title><author>Kalra, Gagan ; Ichhpujani, Parul ; Thakur, Sahil ; Sharma, Urvashi</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c420t-7fe7114c1236565e1447681151d807453abb639b02cc7b36fc1663577393c2003</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>bleb</topic><topic>Original</topic><topic>smartphone photography</topic><topic>teleophthalmology</topic><topic>trabeculectomy</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Kalra, Gagan</creatorcontrib><creatorcontrib>Ichhpujani, Parul</creatorcontrib><creatorcontrib>Thakur, Sahil</creatorcontrib><creatorcontrib>Sharma, Urvashi</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><collection>DOAJ Directory of Open Access Journals</collection><jtitle>Journal of ophthalmic &amp; vision research</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Kalra, Gagan</au><au>Ichhpujani, Parul</au><au>Thakur, Sahil</au><au>Sharma, Urvashi</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Ideal Illumination for Smartphone-based Trabeculectomy Bleb Photography</atitle><jtitle>Journal of ophthalmic &amp; vision research</jtitle><addtitle>J Ophthalmic Vis Res</addtitle><date>2021-07-01</date><risdate>2021</risdate><volume>16</volume><issue>3</issue><spage>357</spage><epage>366</epage><pages>357-366</pages><issn>2008-2010</issn><eissn>2008-322X</eissn><abstract>Ophthalmology has seen numerous novel uses for smartphones over the years including fundus photography, telemedicine, and operative videography. However, anterior segment photography for assessing and documenting trabeculectomy bleb morphology using a smartphone has not been explored in detail. With the current study, we aim to characterize ideal illumination for the anterior segment smartphone photography in trabeculectomy patients. Thirty status post-trabeculectomy patients were enrolled in this study. Native camera application and FiLMiC pro camera application were used on iPhone X to compare bleb images using yellow and white pen-torches as illumination source. Measured bleb area was compared using ImageJ software from the two apps in different illumination settings by charting boxplots and using one-way ANOVA test using R software to establish consistency. Bland-Altman interoperability for repeatability of bleb-area measurements was analyzed by plotting Bland-Altman plots. Signal-to-noise ratio was calculated using ImageJ for native camera images using slit-lamp camera images as reference. Subjective rating of these images was then performed by two experienced ophthalmologists and kappa coefficient was calculated for inter-operator repeatability. Statistical analysis was performed. The measured bleb area from images taken from both apps showed no significant difference, thereby establishing consistency, and Bland-Altman analysis indicated good repeatability and reproducibility. It was noted that SNR was lower for images shot in close illumination as compared to the ones shot in intermediate and distant illumination. Cohen's kappa coefficient was 0.7 for images with distant illumination using white light and 0.65 for images clicked with illumination at an intermediate distance using yellow light, suggesting substantial agreement between the observers. Smartphone photography is a reliable tool for morphological assessment trabeculectomy blebs. Optimal illumination helps achieve results free from digital noise and better delineation of specific morphological features. Intermediate illumination and distant illumination provides much better results in terms of high SNR while avoiding overexposure and clipping of highlight information in the images.</abstract><cop>United Arab Emirates</cop><pub>PUBLISHED BY KNOWLEDGE E</pub><pmid>34394864</pmid><doi>10.18502/jovr.v16i3.9432</doi><tpages>10</tpages><orcidid>https://orcid.org/0000-0001-6256-4002</orcidid><orcidid>https://orcid.org/0000-0002-3367-3047</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2008-2010
ispartof Journal of ophthalmic & vision research, 2021-07, Vol.16 (3), p.357-366
issn 2008-2010
2008-322X
language eng
recordid cdi_doaj_primary_oai_doaj_org_article_fabe23063d9f409ab3df8647c8d8b92e
source PubMed Central
subjects bleb
Original
smartphone photography
teleophthalmology
trabeculectomy
title Ideal Illumination for Smartphone-based Trabeculectomy Bleb Photography
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-08T23%3A49%3A36IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_doaj_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Ideal%20Illumination%20for%20Smartphone-based%20Trabeculectomy%20Bleb%20Photography&rft.jtitle=Journal%20of%20ophthalmic%20&%20vision%20research&rft.au=Kalra,%20Gagan&rft.date=2021-07-01&rft.volume=16&rft.issue=3&rft.spage=357&rft.epage=366&rft.pages=357-366&rft.issn=2008-2010&rft.eissn=2008-322X&rft_id=info:doi/10.18502/jovr.v16i3.9432&rft_dat=%3Cproquest_doaj_%3E2561917700%3C/proquest_doaj_%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c420t-7fe7114c1236565e1447681151d807453abb639b02cc7b36fc1663577393c2003%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2561917700&rft_id=info:pmid/34394864&rfr_iscdi=true