Turbulent Systolic Flow Downstream of a Bioprosthetic Aortic Valve: Velocity Spectra, Wall Shear Stresses, and Turbulent Dissipation Rates

Every year, a quarter million patients receive prosthetic heart valves in aortic valve replacement therapy. Prosthetic heart valves are known to lead to turbulent blood flow. This turbulent flow field may have adverse effects on blood itself, on the aortic wall and on the valve performance. A detail...

Full description

Saved in:
Bibliographic Details
Published in:Frontiers in physiology 2020-09, Vol.11, p.577188-577188
Main Authors: Becsek, Barna, Pietrasanta, Leonardo, Obrist, Dominik
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c442t-16fd675e412b312c20f951244a03aef196f4b2b34e5ea07fd35243d6d0dd220d3
cites cdi_FETCH-LOGICAL-c442t-16fd675e412b312c20f951244a03aef196f4b2b34e5ea07fd35243d6d0dd220d3
container_end_page 577188
container_issue
container_start_page 577188
container_title Frontiers in physiology
container_volume 11
creator Becsek, Barna
Pietrasanta, Leonardo
Obrist, Dominik
description Every year, a quarter million patients receive prosthetic heart valves in aortic valve replacement therapy. Prosthetic heart valves are known to lead to turbulent blood flow. This turbulent flow field may have adverse effects on blood itself, on the aortic wall and on the valve performance. A detailed characterization of the turbulent flow downstream of a valve could yield better understanding of its effect on shear-induced thrombocyte activation, unphysiological wall shear stresses and hemodynamic valve performance. Therefore, computational simulations of the flow past a bioprosthetic heart valve were performed. The computational results were validated against experimental measurements of the turbulent flow field with tomographic particle image velocimetry. The turbulent flow was analyzed for disturbance amplitudes, dissipation rates and shear stress distributions. It was found that approximately 26% of the hydrodynamic resistance of the valve was due to turbulent dissipation and that this dissipation mainly took place in a region about one valve diameter downstream of the valve orifice. Farther downstream, the turbulent fluctuations became weaker which was also reflected in the turbulent velocity spectra of the flow field. Viscous shear stresses were found to be in the range of the critical level for blood platelet activation. The turbulent flow led to elevated shear stress levels along the wall of the ascending aorta with strongly fluctuating and chaotic wall shear stress patterns. Further, we identified leaflet fluttering at 40 Hz which was connected to repeated shedding of vortex rings that appeared to feed the turbulent flow downstream of the valve.
doi_str_mv 10.3389/fphys.2020.577188
format article
fullrecord <record><control><sourceid>proquest_doaj_</sourceid><recordid>TN_cdi_doaj_primary_oai_doaj_org_article_fb2183f6909b4db0bb51ceedf732cb89</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><doaj_id>oai_doaj_org_article_fb2183f6909b4db0bb51ceedf732cb89</doaj_id><sourcerecordid>2455834846</sourcerecordid><originalsourceid>FETCH-LOGICAL-c442t-16fd675e412b312c20f951244a03aef196f4b2b34e5ea07fd35243d6d0dd220d3</originalsourceid><addsrcrecordid>eNpVkstOVEEQhk-MRgjyAO566YIZ-3ouLkwQRElITBxEd52-VDNNek4funsg8wo8NT0MUalNVar-fJVK_U3znuA5Y_3w0U3LTZ5TTPFcdB3p-1fNPmlbPsOc_nn9X73XHOZ8g2vwKsbkbbPHGCEdGfh-83C5TnodYCxoscklBm_QWYj36DTej7kkUCsUHVLoi49TirksoVTJcUzbdKXCHXxCVxCi8WWDFhOYktQR-q1CQIslqIQWFZIz5COkRov-rTv1OftJFR9H9FMVyO-aN06FDIfP-aD5dfb18uT77OLHt_OT44uZ4ZyWGWmdbTsBnFDNCDUUu0EQyrnCTIEjQ-u4riMOAhTunGWCcmZbi62lFFt20JzvuDaqGzklv1JpI6Py8qkR07VU2-sCSKcp6ZlrBzxobjXWWhADYF3HqNH9UFmfd6xprVdgTT0sqfAC-nIy-qW8jneyEwJ3raiAD8-AFG_XkItc-WwgBDVCXGdJuRA94z1vq5TspKb-ISdwf9cQLLeWkE-WkFtLyJ0l2CNoq6yP</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2455834846</pqid></control><display><type>article</type><title>Turbulent Systolic Flow Downstream of a Bioprosthetic Aortic Valve: Velocity Spectra, Wall Shear Stresses, and Turbulent Dissipation Rates</title><source>PubMed Central</source><creator>Becsek, Barna ; Pietrasanta, Leonardo ; Obrist, Dominik</creator><creatorcontrib>Becsek, Barna ; Pietrasanta, Leonardo ; Obrist, Dominik</creatorcontrib><description>Every year, a quarter million patients receive prosthetic heart valves in aortic valve replacement therapy. Prosthetic heart valves are known to lead to turbulent blood flow. This turbulent flow field may have adverse effects on blood itself, on the aortic wall and on the valve performance. A detailed characterization of the turbulent flow downstream of a valve could yield better understanding of its effect on shear-induced thrombocyte activation, unphysiological wall shear stresses and hemodynamic valve performance. Therefore, computational simulations of the flow past a bioprosthetic heart valve were performed. The computational results were validated against experimental measurements of the turbulent flow field with tomographic particle image velocimetry. The turbulent flow was analyzed for disturbance amplitudes, dissipation rates and shear stress distributions. It was found that approximately 26% of the hydrodynamic resistance of the valve was due to turbulent dissipation and that this dissipation mainly took place in a region about one valve diameter downstream of the valve orifice. Farther downstream, the turbulent fluctuations became weaker which was also reflected in the turbulent velocity spectra of the flow field. Viscous shear stresses were found to be in the range of the critical level for blood platelet activation. The turbulent flow led to elevated shear stress levels along the wall of the ascending aorta with strongly fluctuating and chaotic wall shear stress patterns. Further, we identified leaflet fluttering at 40 Hz which was connected to repeated shedding of vortex rings that appeared to feed the turbulent flow downstream of the valve.</description><identifier>ISSN: 1664-042X</identifier><identifier>EISSN: 1664-042X</identifier><identifier>DOI: 10.3389/fphys.2020.577188</identifier><identifier>PMID: 33117194</identifier><language>eng</language><publisher>Frontiers Media S.A</publisher><subject>bioprosthetic heart valve ; fluid-structure interaction ; laminar-turbulent transition ; leaflet fluttering ; Physiology ; thrombocyte activation ; wall shear stress</subject><ispartof>Frontiers in physiology, 2020-09, Vol.11, p.577188-577188</ispartof><rights>Copyright © 2020 Becsek, Pietrasanta and Obrist. 2020 Becsek, Pietrasanta and Obrist</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c442t-16fd675e412b312c20f951244a03aef196f4b2b34e5ea07fd35243d6d0dd220d3</citedby><cites>FETCH-LOGICAL-c442t-16fd675e412b312c20f951244a03aef196f4b2b34e5ea07fd35243d6d0dd220d3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC7550765/pdf/$$EPDF$$P50$$Gpubmedcentral$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC7550765/$$EHTML$$P50$$Gpubmedcentral$$Hfree_for_read</linktohtml><link.rule.ids>230,314,727,780,784,885,27924,27925,53791,53793</link.rule.ids></links><search><creatorcontrib>Becsek, Barna</creatorcontrib><creatorcontrib>Pietrasanta, Leonardo</creatorcontrib><creatorcontrib>Obrist, Dominik</creatorcontrib><title>Turbulent Systolic Flow Downstream of a Bioprosthetic Aortic Valve: Velocity Spectra, Wall Shear Stresses, and Turbulent Dissipation Rates</title><title>Frontiers in physiology</title><description>Every year, a quarter million patients receive prosthetic heart valves in aortic valve replacement therapy. Prosthetic heart valves are known to lead to turbulent blood flow. This turbulent flow field may have adverse effects on blood itself, on the aortic wall and on the valve performance. A detailed characterization of the turbulent flow downstream of a valve could yield better understanding of its effect on shear-induced thrombocyte activation, unphysiological wall shear stresses and hemodynamic valve performance. Therefore, computational simulations of the flow past a bioprosthetic heart valve were performed. The computational results were validated against experimental measurements of the turbulent flow field with tomographic particle image velocimetry. The turbulent flow was analyzed for disturbance amplitudes, dissipation rates and shear stress distributions. It was found that approximately 26% of the hydrodynamic resistance of the valve was due to turbulent dissipation and that this dissipation mainly took place in a region about one valve diameter downstream of the valve orifice. Farther downstream, the turbulent fluctuations became weaker which was also reflected in the turbulent velocity spectra of the flow field. Viscous shear stresses were found to be in the range of the critical level for blood platelet activation. The turbulent flow led to elevated shear stress levels along the wall of the ascending aorta with strongly fluctuating and chaotic wall shear stress patterns. Further, we identified leaflet fluttering at 40 Hz which was connected to repeated shedding of vortex rings that appeared to feed the turbulent flow downstream of the valve.</description><subject>bioprosthetic heart valve</subject><subject>fluid-structure interaction</subject><subject>laminar-turbulent transition</subject><subject>leaflet fluttering</subject><subject>Physiology</subject><subject>thrombocyte activation</subject><subject>wall shear stress</subject><issn>1664-042X</issn><issn>1664-042X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><sourceid>DOA</sourceid><recordid>eNpVkstOVEEQhk-MRgjyAO566YIZ-3ouLkwQRElITBxEd52-VDNNek4funsg8wo8NT0MUalNVar-fJVK_U3znuA5Y_3w0U3LTZ5TTPFcdB3p-1fNPmlbPsOc_nn9X73XHOZ8g2vwKsbkbbPHGCEdGfh-83C5TnodYCxoscklBm_QWYj36DTej7kkUCsUHVLoi49TirksoVTJcUzbdKXCHXxCVxCi8WWDFhOYktQR-q1CQIslqIQWFZIz5COkRov-rTv1OftJFR9H9FMVyO-aN06FDIfP-aD5dfb18uT77OLHt_OT44uZ4ZyWGWmdbTsBnFDNCDUUu0EQyrnCTIEjQ-u4riMOAhTunGWCcmZbi62lFFt20JzvuDaqGzklv1JpI6Py8qkR07VU2-sCSKcp6ZlrBzxobjXWWhADYF3HqNH9UFmfd6xprVdgTT0sqfAC-nIy-qW8jneyEwJ3raiAD8-AFG_XkItc-WwgBDVCXGdJuRA94z1vq5TspKb-ISdwf9cQLLeWkE-WkFtLyJ0l2CNoq6yP</recordid><startdate>20200929</startdate><enddate>20200929</enddate><creator>Becsek, Barna</creator><creator>Pietrasanta, Leonardo</creator><creator>Obrist, Dominik</creator><general>Frontiers Media S.A</general><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>5PM</scope><scope>DOA</scope></search><sort><creationdate>20200929</creationdate><title>Turbulent Systolic Flow Downstream of a Bioprosthetic Aortic Valve: Velocity Spectra, Wall Shear Stresses, and Turbulent Dissipation Rates</title><author>Becsek, Barna ; Pietrasanta, Leonardo ; Obrist, Dominik</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c442t-16fd675e412b312c20f951244a03aef196f4b2b34e5ea07fd35243d6d0dd220d3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>bioprosthetic heart valve</topic><topic>fluid-structure interaction</topic><topic>laminar-turbulent transition</topic><topic>leaflet fluttering</topic><topic>Physiology</topic><topic>thrombocyte activation</topic><topic>wall shear stress</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Becsek, Barna</creatorcontrib><creatorcontrib>Pietrasanta, Leonardo</creatorcontrib><creatorcontrib>Obrist, Dominik</creatorcontrib><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><collection>Directory of Open Access Journals</collection><jtitle>Frontiers in physiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Becsek, Barna</au><au>Pietrasanta, Leonardo</au><au>Obrist, Dominik</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Turbulent Systolic Flow Downstream of a Bioprosthetic Aortic Valve: Velocity Spectra, Wall Shear Stresses, and Turbulent Dissipation Rates</atitle><jtitle>Frontiers in physiology</jtitle><date>2020-09-29</date><risdate>2020</risdate><volume>11</volume><spage>577188</spage><epage>577188</epage><pages>577188-577188</pages><issn>1664-042X</issn><eissn>1664-042X</eissn><abstract>Every year, a quarter million patients receive prosthetic heart valves in aortic valve replacement therapy. Prosthetic heart valves are known to lead to turbulent blood flow. This turbulent flow field may have adverse effects on blood itself, on the aortic wall and on the valve performance. A detailed characterization of the turbulent flow downstream of a valve could yield better understanding of its effect on shear-induced thrombocyte activation, unphysiological wall shear stresses and hemodynamic valve performance. Therefore, computational simulations of the flow past a bioprosthetic heart valve were performed. The computational results were validated against experimental measurements of the turbulent flow field with tomographic particle image velocimetry. The turbulent flow was analyzed for disturbance amplitudes, dissipation rates and shear stress distributions. It was found that approximately 26% of the hydrodynamic resistance of the valve was due to turbulent dissipation and that this dissipation mainly took place in a region about one valve diameter downstream of the valve orifice. Farther downstream, the turbulent fluctuations became weaker which was also reflected in the turbulent velocity spectra of the flow field. Viscous shear stresses were found to be in the range of the critical level for blood platelet activation. The turbulent flow led to elevated shear stress levels along the wall of the ascending aorta with strongly fluctuating and chaotic wall shear stress patterns. Further, we identified leaflet fluttering at 40 Hz which was connected to repeated shedding of vortex rings that appeared to feed the turbulent flow downstream of the valve.</abstract><pub>Frontiers Media S.A</pub><pmid>33117194</pmid><doi>10.3389/fphys.2020.577188</doi><tpages>1</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 1664-042X
ispartof Frontiers in physiology, 2020-09, Vol.11, p.577188-577188
issn 1664-042X
1664-042X
language eng
recordid cdi_doaj_primary_oai_doaj_org_article_fb2183f6909b4db0bb51ceedf732cb89
source PubMed Central
subjects bioprosthetic heart valve
fluid-structure interaction
laminar-turbulent transition
leaflet fluttering
Physiology
thrombocyte activation
wall shear stress
title Turbulent Systolic Flow Downstream of a Bioprosthetic Aortic Valve: Velocity Spectra, Wall Shear Stresses, and Turbulent Dissipation Rates
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-06T19%3A05%3A49IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_doaj_&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Turbulent%20Systolic%20Flow%20Downstream%20of%20a%20Bioprosthetic%20Aortic%20Valve:%20Velocity%20Spectra,%20Wall%20Shear%20Stresses,%20and%20Turbulent%20Dissipation%20Rates&rft.jtitle=Frontiers%20in%20physiology&rft.au=Becsek,%20Barna&rft.date=2020-09-29&rft.volume=11&rft.spage=577188&rft.epage=577188&rft.pages=577188-577188&rft.issn=1664-042X&rft.eissn=1664-042X&rft_id=info:doi/10.3389/fphys.2020.577188&rft_dat=%3Cproquest_doaj_%3E2455834846%3C/proquest_doaj_%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c442t-16fd675e412b312c20f951244a03aef196f4b2b34e5ea07fd35243d6d0dd220d3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2455834846&rft_id=info:pmid/33117194&rfr_iscdi=true