Development and Testing of Breadboard Electrochemical Organic Content Analyzer

A Breadboard Electrochemical Organic Content (EOC) Analyzer was designed and developed for on-line monitoring of organic solute concentrations in the ozonated effluent of the Water Processing Element Pilot Plant in the Medical Unit, Self-Contained, Transportable (MUST) hospital. Techniques were inco...

Full description

Saved in:
Bibliographic Details
Main Authors: Davenport,R J, Wynveen,R A
Format: Report
Language:English
Subjects:
Online Access:Request full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A Breadboard Electrochemical Organic Content (EOC) Analyzer was designed and developed for on-line monitoring of organic solute concentrations in the ozonated effluent of the Water Processing Element Pilot Plant in the Medical Unit, Self-Contained, Transportable (MUST) hospital. Techniques were incorporated to eliminate various interferences. The test program included tests of the analyzer's response to organics and tests of the effectiveness of the interference eliminating components. The analyzer successfully completed 30 days of continuous, on-line operation without component failures. Results, conclusions and recommendations are presented. (Author)