Robust Adaptive Control of Hypnosis During Anesthesia
A closed-loop controller for hypnosis was designed and validated on humans at our laboratory. The controller aims at regulating the Bispectral Index (BIS) - a surrogate measure of hypnosis derived from the electroencephalogram of the patient - with the volatile anesthetic isoflurane administered wit...
Saved in:
Main Authors: | , , , |
---|---|
Format: | Report |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A closed-loop controller for hypnosis was designed and validated on humans at our laboratory. The controller aims at regulating the Bispectral Index (BIS) - a surrogate measure of hypnosis derived from the electroencephalogram of the patient - with the volatile anesthetic isoflurane administered with a closed-circuit breathing system. The control algorithm consists of a cascaded Internal Model Controller (IMC) where the master loop aims at regulating BIS. The slave loop tracks endtidal concentration references provided by the master controller. In this paper, a new tuning method is presented.
Papers from 23rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, October 25-28, 2001, held in Istanbul, Turkey. See also ADM001351 for entire conference on cd-rom. |
---|