Quantitative accuracy of .sup.177Lu SPECT imaging for molecular radiotherapy

The purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with .sup....

Full description

Saved in:
Bibliographic Details
Published in:PloS one 2017-08, Vol.12 (8), p.e0182888
Main Authors: Mezzenga, Emilio, D'Errico, Vincenzo, D'Arienzo, Marco, Strigari, Lidia, Panagiota, Koutla, Matteucci, Federica, Severi, Stefano, Paganelli, Giovanni, Fenwick, Andrew, Bianchini, David, Marcocci, Francesco, Sarnelli, Anna
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page
container_issue 8
container_start_page e0182888
container_title PloS one
container_volume 12
creator Mezzenga, Emilio
D'Errico, Vincenzo
D'Arienzo, Marco
Strigari, Lidia
Panagiota, Koutla
Matteucci, Federica
Severi, Stefano
Paganelli, Giovanni
Fenwick, Andrew
Bianchini, David
Marcocci, Francesco
Sarnelli, Anna
description The purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with .sup.177 Lu. The accuracy of .sup.177 Lu activity quantification was assessed both in small and in large objects. Two different reference geometries, namely a cylindrical homogeneous phantom and a Jaszczak 16 ml sphere surrounded by cold water, were used to determine the gamma camera calibration factor of a commercial SPECT/CT system. Moreover, the noise level and the concentration recovery coefficient were evaluated as a function of the number of 3D-OSEM updates by using the SPECT/CT images of the reference geometry phantoms and those of a cold Jaszczak phantom with three hot spheres (16ml, 8ml and 4ml), respectively. The optimal choice of the number of 3D-OSEM updates was based on a compromise between the noise level achievable in the reconstructed SPECT images and the concentration recovery coefficients. The quantitative accuracy achievable was finally validated on a test phantom, where a spherical insert composed of two concentric spheres was used to simulate a lesion in a warm background. Our data confirm and extend previous observations. Using the calibration factor obtained with the cylindrical homogeneous phantom and the Jaszczak 16 ml sphere, the recovered activity in the test phantom was underestimated by -16.4% and -24.8%, respectively. Our work has led us to conclude that gamma camera calibration performed with large homogeneous phantom outperforms calibration executed with the Jaszczak 16ml sphere. Furthermore, the results obtained support the assumption that approximately 50 OSEM updates represent a good trade-off to reach convergence in small volumes, meanwhile minimizing the noise level.
doi_str_mv 10.1371/journal.pone.0182888
format article
fullrecord <record><control><sourceid>gale</sourceid><recordid>TN_cdi_gale_infotracmisc_A500574567</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A500574567</galeid><sourcerecordid>A500574567</sourcerecordid><originalsourceid>FETCH-LOGICAL-g1667-b6fa8c371d8ce24908cc4909ce9bcefdf160c2afb5ab6eb84df65cec5d138c5c3</originalsourceid><addsrcrecordid>eNqNkE1Lw0AQhoMoWKv_wMOCIHhIzCbNZnMsxY9CoGqr1zKZ7KZbkmzJ7or99y7ooQUPMjAzDM87vDNBcE3jiKY5vd9qN_TQRjvdiyimPOGcnwQjWqRJyJI4PT3oz4MLY7ZxnKWcsVFQvjrorbJg1acggOgGwD3RkkTG7SKa56Ujy5eH2YqoDhrVN0TqgXS6FehaGMgAtdJ2IwbY7S-DMwmtEVe_dRy8Pz6sZs9huXiaz6Zl2FDG8rBiEjh64zVHkUyKmCP6XKAoKhSylpTFmICsMqiYqPiklixDgVlNU44ZpuPg5mdvA61Yq15q6113yuB6mvnT8knGck9Ff1A-atEp9K-Sys-PBHdHAs9Y8WUbcMas58u3_7OLj2P29oDdCGjtxujWWaV7cwh-A-10iz0</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Quantitative accuracy of .sup.177Lu SPECT imaging for molecular radiotherapy</title><source>PubMed (Medline)</source><source>Publicly Available Content Database</source><creator>Mezzenga, Emilio ; D'Errico, Vincenzo ; D'Arienzo, Marco ; Strigari, Lidia ; Panagiota, Koutla ; Matteucci, Federica ; Severi, Stefano ; Paganelli, Giovanni ; Fenwick, Andrew ; Bianchini, David ; Marcocci, Francesco ; Sarnelli, Anna</creator><creatorcontrib>Mezzenga, Emilio ; D'Errico, Vincenzo ; D'Arienzo, Marco ; Strigari, Lidia ; Panagiota, Koutla ; Matteucci, Federica ; Severi, Stefano ; Paganelli, Giovanni ; Fenwick, Andrew ; Bianchini, David ; Marcocci, Francesco ; Sarnelli, Anna</creatorcontrib><description>The purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with .sup.177 Lu. The accuracy of .sup.177 Lu activity quantification was assessed both in small and in large objects. Two different reference geometries, namely a cylindrical homogeneous phantom and a Jaszczak 16 ml sphere surrounded by cold water, were used to determine the gamma camera calibration factor of a commercial SPECT/CT system. Moreover, the noise level and the concentration recovery coefficient were evaluated as a function of the number of 3D-OSEM updates by using the SPECT/CT images of the reference geometry phantoms and those of a cold Jaszczak phantom with three hot spheres (16ml, 8ml and 4ml), respectively. The optimal choice of the number of 3D-OSEM updates was based on a compromise between the noise level achievable in the reconstructed SPECT images and the concentration recovery coefficients. The quantitative accuracy achievable was finally validated on a test phantom, where a spherical insert composed of two concentric spheres was used to simulate a lesion in a warm background. Our data confirm and extend previous observations. Using the calibration factor obtained with the cylindrical homogeneous phantom and the Jaszczak 16 ml sphere, the recovered activity in the test phantom was underestimated by -16.4% and -24.8%, respectively. Our work has led us to conclude that gamma camera calibration performed with large homogeneous phantom outperforms calibration executed with the Jaszczak 16ml sphere. Furthermore, the results obtained support the assumption that approximately 50 OSEM updates represent a good trade-off to reach convergence in small volumes, meanwhile minimizing the noise level.</description><identifier>ISSN: 1932-6203</identifier><identifier>EISSN: 1932-6203</identifier><identifier>DOI: 10.1371/journal.pone.0182888</identifier><language>eng</language><publisher>Public Library of Science</publisher><subject>Gamma cameras ; Radiotherapy ; Single photon emission computed tomography</subject><ispartof>PloS one, 2017-08, Vol.12 (8), p.e0182888</ispartof><rights>COPYRIGHT 2017 Public Library of Science</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids></links><search><creatorcontrib>Mezzenga, Emilio</creatorcontrib><creatorcontrib>D'Errico, Vincenzo</creatorcontrib><creatorcontrib>D'Arienzo, Marco</creatorcontrib><creatorcontrib>Strigari, Lidia</creatorcontrib><creatorcontrib>Panagiota, Koutla</creatorcontrib><creatorcontrib>Matteucci, Federica</creatorcontrib><creatorcontrib>Severi, Stefano</creatorcontrib><creatorcontrib>Paganelli, Giovanni</creatorcontrib><creatorcontrib>Fenwick, Andrew</creatorcontrib><creatorcontrib>Bianchini, David</creatorcontrib><creatorcontrib>Marcocci, Francesco</creatorcontrib><creatorcontrib>Sarnelli, Anna</creatorcontrib><title>Quantitative accuracy of .sup.177Lu SPECT imaging for molecular radiotherapy</title><title>PloS one</title><description>The purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with .sup.177 Lu. The accuracy of .sup.177 Lu activity quantification was assessed both in small and in large objects. Two different reference geometries, namely a cylindrical homogeneous phantom and a Jaszczak 16 ml sphere surrounded by cold water, were used to determine the gamma camera calibration factor of a commercial SPECT/CT system. Moreover, the noise level and the concentration recovery coefficient were evaluated as a function of the number of 3D-OSEM updates by using the SPECT/CT images of the reference geometry phantoms and those of a cold Jaszczak phantom with three hot spheres (16ml, 8ml and 4ml), respectively. The optimal choice of the number of 3D-OSEM updates was based on a compromise between the noise level achievable in the reconstructed SPECT images and the concentration recovery coefficients. The quantitative accuracy achievable was finally validated on a test phantom, where a spherical insert composed of two concentric spheres was used to simulate a lesion in a warm background. Our data confirm and extend previous observations. Using the calibration factor obtained with the cylindrical homogeneous phantom and the Jaszczak 16 ml sphere, the recovered activity in the test phantom was underestimated by -16.4% and -24.8%, respectively. Our work has led us to conclude that gamma camera calibration performed with large homogeneous phantom outperforms calibration executed with the Jaszczak 16ml sphere. Furthermore, the results obtained support the assumption that approximately 50 OSEM updates represent a good trade-off to reach convergence in small volumes, meanwhile minimizing the noise level.</description><subject>Gamma cameras</subject><subject>Radiotherapy</subject><subject>Single photon emission computed tomography</subject><issn>1932-6203</issn><issn>1932-6203</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2017</creationdate><recordtype>article</recordtype><recordid>eNqNkE1Lw0AQhoMoWKv_wMOCIHhIzCbNZnMsxY9CoGqr1zKZ7KZbkmzJ7or99y7ooQUPMjAzDM87vDNBcE3jiKY5vd9qN_TQRjvdiyimPOGcnwQjWqRJyJI4PT3oz4MLY7ZxnKWcsVFQvjrorbJg1acggOgGwD3RkkTG7SKa56Ujy5eH2YqoDhrVN0TqgXS6FehaGMgAtdJ2IwbY7S-DMwmtEVe_dRy8Pz6sZs9huXiaz6Zl2FDG8rBiEjh64zVHkUyKmCP6XKAoKhSylpTFmICsMqiYqPiklixDgVlNU44ZpuPg5mdvA61Yq15q6113yuB6mvnT8knGck9Ff1A-atEp9K-Sys-PBHdHAs9Y8WUbcMas58u3_7OLj2P29oDdCGjtxujWWaV7cwh-A-10iz0</recordid><startdate>20170814</startdate><enddate>20170814</enddate><creator>Mezzenga, Emilio</creator><creator>D'Errico, Vincenzo</creator><creator>D'Arienzo, Marco</creator><creator>Strigari, Lidia</creator><creator>Panagiota, Koutla</creator><creator>Matteucci, Federica</creator><creator>Severi, Stefano</creator><creator>Paganelli, Giovanni</creator><creator>Fenwick, Andrew</creator><creator>Bianchini, David</creator><creator>Marcocci, Francesco</creator><creator>Sarnelli, Anna</creator><general>Public Library of Science</general><scope>IOV</scope><scope>ISR</scope></search><sort><creationdate>20170814</creationdate><title>Quantitative accuracy of .sup.177Lu SPECT imaging for molecular radiotherapy</title><author>Mezzenga, Emilio ; D'Errico, Vincenzo ; D'Arienzo, Marco ; Strigari, Lidia ; Panagiota, Koutla ; Matteucci, Federica ; Severi, Stefano ; Paganelli, Giovanni ; Fenwick, Andrew ; Bianchini, David ; Marcocci, Francesco ; Sarnelli, Anna</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-g1667-b6fa8c371d8ce24908cc4909ce9bcefdf160c2afb5ab6eb84df65cec5d138c5c3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2017</creationdate><topic>Gamma cameras</topic><topic>Radiotherapy</topic><topic>Single photon emission computed tomography</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Mezzenga, Emilio</creatorcontrib><creatorcontrib>D'Errico, Vincenzo</creatorcontrib><creatorcontrib>D'Arienzo, Marco</creatorcontrib><creatorcontrib>Strigari, Lidia</creatorcontrib><creatorcontrib>Panagiota, Koutla</creatorcontrib><creatorcontrib>Matteucci, Federica</creatorcontrib><creatorcontrib>Severi, Stefano</creatorcontrib><creatorcontrib>Paganelli, Giovanni</creatorcontrib><creatorcontrib>Fenwick, Andrew</creatorcontrib><creatorcontrib>Bianchini, David</creatorcontrib><creatorcontrib>Marcocci, Francesco</creatorcontrib><creatorcontrib>Sarnelli, Anna</creatorcontrib><collection>Opposing Viewpoints In Context</collection><collection>Gale In Context: Science</collection><jtitle>PloS one</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Mezzenga, Emilio</au><au>D'Errico, Vincenzo</au><au>D'Arienzo, Marco</au><au>Strigari, Lidia</au><au>Panagiota, Koutla</au><au>Matteucci, Federica</au><au>Severi, Stefano</au><au>Paganelli, Giovanni</au><au>Fenwick, Andrew</au><au>Bianchini, David</au><au>Marcocci, Francesco</au><au>Sarnelli, Anna</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Quantitative accuracy of .sup.177Lu SPECT imaging for molecular radiotherapy</atitle><jtitle>PloS one</jtitle><date>2017-08-14</date><risdate>2017</risdate><volume>12</volume><issue>8</issue><spage>e0182888</spage><pages>e0182888-</pages><issn>1932-6203</issn><eissn>1932-6203</eissn><abstract>The purpose of this study is to investigate the optimal reference geometry for gamma camera calibration. Yet another question of interest was to assess the influence of the number of 3D Ordered Subsets Expectation Maximization (3D-OSEM) updates on activity quantification for SPECT imaging with .sup.177 Lu. The accuracy of .sup.177 Lu activity quantification was assessed both in small and in large objects. Two different reference geometries, namely a cylindrical homogeneous phantom and a Jaszczak 16 ml sphere surrounded by cold water, were used to determine the gamma camera calibration factor of a commercial SPECT/CT system. Moreover, the noise level and the concentration recovery coefficient were evaluated as a function of the number of 3D-OSEM updates by using the SPECT/CT images of the reference geometry phantoms and those of a cold Jaszczak phantom with three hot spheres (16ml, 8ml and 4ml), respectively. The optimal choice of the number of 3D-OSEM updates was based on a compromise between the noise level achievable in the reconstructed SPECT images and the concentration recovery coefficients. The quantitative accuracy achievable was finally validated on a test phantom, where a spherical insert composed of two concentric spheres was used to simulate a lesion in a warm background. Our data confirm and extend previous observations. Using the calibration factor obtained with the cylindrical homogeneous phantom and the Jaszczak 16 ml sphere, the recovered activity in the test phantom was underestimated by -16.4% and -24.8%, respectively. Our work has led us to conclude that gamma camera calibration performed with large homogeneous phantom outperforms calibration executed with the Jaszczak 16ml sphere. Furthermore, the results obtained support the assumption that approximately 50 OSEM updates represent a good trade-off to reach convergence in small volumes, meanwhile minimizing the noise level.</abstract><pub>Public Library of Science</pub><doi>10.1371/journal.pone.0182888</doi><tpages>e0182888</tpages></addata></record>
fulltext fulltext
identifier ISSN: 1932-6203
ispartof PloS one, 2017-08, Vol.12 (8), p.e0182888
issn 1932-6203
1932-6203
language eng
recordid cdi_gale_infotracmisc_A500574567
source PubMed (Medline); Publicly Available Content Database
subjects Gamma cameras
Radiotherapy
Single photon emission computed tomography
title Quantitative accuracy of .sup.177Lu SPECT imaging for molecular radiotherapy
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-31T04%3A33%3A56IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Quantitative%20accuracy%20of%20.sup.177Lu%20SPECT%20imaging%20for%20molecular%20radiotherapy&rft.jtitle=PloS%20one&rft.au=Mezzenga,%20Emilio&rft.date=2017-08-14&rft.volume=12&rft.issue=8&rft.spage=e0182888&rft.pages=e0182888-&rft.issn=1932-6203&rft.eissn=1932-6203&rft_id=info:doi/10.1371/journal.pone.0182888&rft_dat=%3Cgale%3EA500574567%3C/gale%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-g1667-b6fa8c371d8ce24908cc4909ce9bcefdf160c2afb5ab6eb84df65cec5d138c5c3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rft_galeid=A500574567&rfr_iscdi=true