Mathematical Modeling of Hemolysis in Pulsatile Blood Pumps: Theory and Design
The paper presents the results of calculation of the normalized index of hemolysis using three different approaches based on mathematical modeling of a pulsatile circulatory support system with a stroke volume of 30 [cm.sup.3]. The suggested approaches are shown to be useful for estimating the hemol...
Saved in:
Published in: | Biomedical engineering 2017-07, Vol.51 (2), p.77 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The paper presents the results of calculation of the normalized index of hemolysis using three different approaches based on mathematical modeling of a pulsatile circulatory support system with a stroke volume of 30 [cm.sup.3]. The suggested approaches are shown to be useful for estimating the hemolysis level in pulsatile circulatory support systems at various stages of the design process. |
---|---|
ISSN: | 0006-3398 1573-8256 |